Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.2478/acph-2013-0012

Quantitation of nitrofurantoin in human plasma by liquid chromatography tandem mass spectrometry

DINESH S. PATEL ; Chemistry Department, Pramukh Swami Science and H.D. Patel Arts College, Sarva Vidyalaya Campus, Kadi-382 715, Gujarat, India; Bio-Analytical Laboratory, Cliantha Research Ltd., Bodakdev, Ahmedabad-380054, Gujarat, India
NAVEEN SHARMA ; Bio-Analytical Laboratory, Cliantha Research Ltd., Bodakdev, Ahmedabad-380054, Gujarat, India
MUKESH C. PATEL ; 1 Chemistry Department, Pramukh Swami Science and H.D. Patel Arts College, Sarva Vidyalaya Campus, Kadi-382 715, Gujarat, India
BHAVIN N. PATEL ; Bio-Analytical Laboratory, Cliantha Research Ltd., Bodakdev, Ahmedabad-380054, Gujarat, India; Department of Chemistry, School of Sciences, Gujarat University, Navrangpura, Ahmedabad-380009, Gujarat, India
PRANAV S. SHRIVASTAV ; Department of Chemistry, School of Sciences, Gujarat University, Navrangpura, Ahmedabad-380009, Gujarat, India
MALLIKA SANYAL ; Department of Chemistry, St. Xavier’s College, Navrangpura, Ahmedabad-380009, Gujarat, India


Puni tekst: engleski pdf 348 Kb

str. 141-158

preuzimanja: 2.469

citiraj


Sažetak

A reliable, selective and sensitive LC-MS/MS assay has been proposed for the determination of nitrofurantoin in human plasma. The analyte and nitrofurazone were extracted from 100 µL of human plasma via SPE on Strata-X 33 µm extraction cartridges. Chromatography was done on a BDS Hypersil C18 (100 mm × 4.6 mm, 5 µm) column under isocratic conditions. Quantitation was done using the multiple reaction monitoring (MRM) mode for deprotonated precursor to product ion transitions of nitrofurantoin (m/z 237.0 → 151.8) and nitrofurazone (m/z 197.0 →123.9). The limit of detection and the lowest limit of quantitation of the method were 0.25 ng mL–1 and 5.00 ng mL–1, respectively, with a linear dynamic range of 5.00–1500 ng mL–1 for nitrofurantoin. The intra-batch and inter-batch precision (RSD, %) was ≤ 5.8 %, while the mean extraction recovery was > 92 %. The method was successfully applied to a bioequivalence study of a 100 mg nitrofurantoin capsule formulation in 36 healthy subjects.

Ključne riječi

nitrofurantoin; LC-MS/MS; solid phase extraction; human plasma; bioequivalence; incurred sample reanalysis

Hrčak ID:

97694

URI

https://hrcak.srce.hr/97694

Datum izdavanja:

30.6.2013.

Posjeta: 3.199 *