The Effect of Optical Conditioning of Preparations with Scan Spray on Preparation Form

Karl M Lehmann ; Zavod za stomatološku protetiku, Medicinski fakultet Sveučilišta u Mainzu, Njemačka
Mehdi Sattari Azar ; Zavod za stomatološku protetiku, Medicinski fakultet Sveučilišta u Mainzu, Njemačka
Peer W Kämmerer ; Zavod za oralnu i maksilofacijalnu kirurgiju, Medicinski fakultet Sveučilište u Mainzu, Njemačka
Stefan Wentaschek ; Zavod za stomatološku protetiku, Medicinski fakultet Sveučilišta u Mainzu, Njemačka
Ernst NF Hell ; Zavod za stomatološku protetiku, Medicinski fakultet Sveučilišta u Mainzu, Njemačka
Herbert Scheller ; Zavod za stomatološku protetiku, Medicinski fakultet Sveučilišta u Mainzu, Njemačka


Puni tekst: hrvatski pdf 338 Kb

str. 86-92

preuzimanja: 705

citiraj

Puni tekst: engleski pdf 338 Kb

str. 86-92

preuzimanja: 599

citiraj


Sažetak

Objective: CAD/CAM systems require scan spray for chairside sampling of tooth surfaces. However, this leads to alterations of the sprayed preparation and may affect marginal and internal adaption
of the restoration. The aim of this study was to define the effect of scan spray on preparation form by evaluating the morphologies and the thicknesses of three different scan sprays. Materials and Methods: The surface morphology of the coatings of different scan sprays (A, B, C) was evaluated by use of a confocal laser scanning microscope. The thicknesses of spray coatings were measured on a glass plate as well as under simulated clinical conditions using a profilometer and a surface digitalisation device. The data were analysed by a 1-way ANOVA (α = .05). Results: The confocal laser scanning microscopy showed rough and irregular formed coating surfaces of all sprays, whereas an increasing numbers of small spray particles were seen from spray A over spray B to spray C. Mean coating thicknesses on the glass plate were: spray A 25.3 Μm, spray B 18.9 Μm and spray C 19.2 Μm. Significant differences were detected between spray A and B (P = .017). Under simulated clinical conditions median coating thicknesses were: spray A 15.5 Μm, spray B 15.0 Μm and spray C 13.3 Μm. A statistically significant difference could be detected only between spray A and C (P = .033). Conclusions: The coatings of the used scan spray systems showed different surface morphologies and exhibited different thicknesses. Consequently, the application of scan sprays affects the precision of the final CAD/CAM restorations. However, the scan sprays under investigation showed acceptable coating thicknesses for clinical use.

Ključne riječi

Computer-Aided Design; Computer simulation; Microscopy, Confocal; cad/cam; dental casting technique; dental impression technique