Changes of the retentional force ceka intraoral anchors caused by the repeated removals

Miodrag M. Nastić ; Klinika za stomatološku protetiku, Stomatološkog fakulteta Univerziteta u Beogradu, Beograd, Srbija


Puni tekst: srpski pdf 7.989 Kb

str. 127-136

preuzimanja: 829

citiraj


Sažetak

A retainer should provide force capable of securing the position of the partial denture an still not damaging the periodontium of the abutment teeth. It’s almost generally accepted that the retentional force should not be greater than 10 N nor weaker than 5 N. This study was designed to measure the forces required to disconnect parts of CEKA intracoronal anchors initialy and repeatedly up to 50000 cycles. The results are compared with suggested values. It was found that the initial force exeeded 10 N. A fter repeated removals retentional force decreased, but never under values of 5 N.

Ključne riječi

CEKA intracorolal anchor; retainer; abutment teeth