https://doi.org/10.11613/BM.2015.014

Paraprotein interference with turbidimetric gentamicin assay

Goce Dimeski orcid id orcid.org/0000-0002-0228-0742 ; Department of Chemical Pathology, Pathology Queensland, Princess Alexandra Hospital, Brisbane, Australia
Kendra Bassett ; Department of Chemical Pathology, Pathology Queensland, Princess Alexandra Hospital, Brisbane, Australia
Nigel Brown ; Department of Chemical Pathology, Pathology Queensland, Princess Alexandra Hospital, Brisbane, Australia


Puni tekst: engleski pdf 368 Kb

str. 117-124

preuzimanja: 479

citiraj


Sažetak


Introduction: Gentamicin due to its low level of resistance and rapid bactericidal activity is commonly used to treat gram-negative bacteria. However, due to its toxic effects it needs to be monitored. To date, no interference has been reported with gentamicin assays.
Materials and methods: A patient with leg cellulitis and sepsis received a single dose of gentamicin and a sample was sent for gentamicin analysis. The sample showed high blank absorbance readings on Beckman DxC800 and DC800 analysers with various dilutions. A second sample was received and analysed on a Roche Cobas system to obtain a result. A third sample was received 107 hours later with the same results and this sample was then analysed neat and post ethanol precipitation on all the turbidimetric assays available on the DxC800 analyser.
Results: The high blank absorbance was observed upon addition of the reactive reagents due to protein precipitation. Although not obvious from the patient protein results, it was shown the presence of high IgM paraprotein, 18.9 g/L (reference range 0.4-2.3 g/L) was the cause of precipitation, giving high blank readings. Of all the other turbidimetric assays, only vancomicin and valproate showed similar high blank absorbance readings. To be able to provide more rapid results it was shown ethanol could be used as a precipitant of proteins in both calibrators and patient samples with acceptable recovery.
Conclusion: IgM paraprotein was identified as the cause of interference with the gentamicin, vancomicin and valproate assays. Protein interference in these assays can be overcome by precipitation with ethanol.

Ključne riječi

turbidimetry; IgM; paraprotein; interference