Effectiveness of citrate buffer-fluoride mixture in Terumo tubes as an inhibitor of in vitro glycolysis

Graziella Bonetti ; Clinical Chemistry Laboratory, Spedali Civili of Brescia, Brescia, Italy
Mariarosa Carta ; Clinical Chemistry and Haematology Laboratory, St. Bortolo Hospital, Vicenza, Italy
Martina Montagnana ; Clinical Biochemistry Section, University of Verona, Verona, Italy
Claudia Lo Cascio ; Clinical Laboratory, AOUI Verona, Verona, Italy
Anna Rita Bonfigli ; Scientific Direction, Italian National Research Center on Aging (INRCA), Ancona, Italy
Andrea Mosca ; Department of Physiopathology and Transplantation, University of Milano, Milano, Italy
Roberto Testa ; Experimental Models in Clinical Pathology, INRCA-IRCCS National Institute, Ancona, Italy


Puni tekst: engleski pdf 243 Kb

str. 68-76

preuzimanja: 572

citiraj


Sažetak

Introduction: Glycolysis affects glucose determination in vitro. The placement of sample tubes in ice-water slurry with plasma separation within 30 minutes is recommended, or alternatively the use of a glycolysis inhibitor. The aim of our two-steps study was to evaluate which Terumo tube is best for glucose determination in routine clinical setting.
Materials and methods: In the first study, blood from 100 volunteers was collected into lithium heparin (LH), NaF/Na heparin (FH) and NaF/citrate buffer/Na2EDTA (FC-Mixture) tubes. LH sample was treated as recommended and considered as reference, while FH and FC-Mixture samples were aliquoted, maintained at room temperature (RT) for 1, 2 and 4 hours; centrifuged and plasma analysed in triplicate. In the second study, samples from 375 volunteers were collected in LH, FH and FC-Mixture tubes and held at RT before centrifugation from 10 to 340 minutes, depending on each laboratory practice. Samples were analysed in one analytical run.
Results: In the first study, FH glucose concentrations were 5.15 ± 0.66 mmol/L, 5.05 ± 0.65 mmol/L and 5.00 ± 0.65 mmol/L (P < 0.001) in tubes stored at RT for 1, 2 and 4 hours, respectively. Mean biases in all time points exceeded the analytical goal for desirable bias based on biological variation criteria. FC-Mixture glucose concentrations were 5.48 ± 0.65 mmol/L, 5.46 ± 0.6 mmol/L and 5.46 ± 0.64 mmol/L in tubes stored at RT for 1, 2 and 4 hours, respectively. Mean biases for FC-Mixture glucose in all time points reached optimal analytical goals. In the second study, the biases for LH and FH glucose compared to reference FC-Mixture glucose exceeded the preset analytical goals, regardless of the blood collection to centrifugation time interval.
Conclusions: FC-mixture tubes glucose concentrations were preserved up to 4h storage at RT. We confirmed that NaF alone does not allow immediate glycolysis inhibition in real life pre-centrifugation storage conditions (up to 340 minutes). FC-Mixture should be used exclusively for glucose determination in laboratories unable to implement the recommended blood samples’ treatment.

Ključne riječi

glucose; pre-analytical phase; sodium fluoride; citrate acidification; stability