Measurement of HbA1c and HbA2 by Capillarys 2 Flex Piercing HbA1c programme for simultaneous management of diabetes and screening for thalassemia

Peifeng Ke ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Jiawei Liu ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China; Second Clinical Medical College, Guangzhou University of Chinese Medicine, Guangzhou, China
Yan Chao ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Xiaobin Wu ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Yujuan Xiong ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Li Lin ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Zemin Wan ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Xinzhong Wu ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Jianhua Xu ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Junhua Zhuang ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China
Xianzhang Huang ; Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China


Puni tekst: engleski pdf 929 Kb

str. 515-521

preuzimanja: 483

citiraj


Sažetak

Introduction: Thalassemia could interfere with some assays for haemoglobin A1c (HbA1c) measurement, therefore, it is useful to be able to screen
for thalassemia while measuring HbA1c. We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA1c programme to simultaneously measure HbA1c and
screen for thalassemia.
Materials and methods: Samples from 498 normal controls and 175 thalassemia patients were analysed by Capillarys 2FP HbA1c programme
(Sebia, France). For method comparison, HbA1c was quantified by Premier Hb9210 (Trinity Biotech, Ireland) in 98 thalassaemia patients samples. For
verification, HbA1c from eight thalassaemia patients was confirmed by IFCC reference method.
Results: Among 98 thalassaemia samples, Capillarys 2FP did not provide an HbA1c result in three samples with HbH due to the overlapping of
HbBart’s with HbA1c fraction; for the remaining 95 thalassaemia samples, Bland-Altman plot showed 0.00 ± 0.35% absolute bias between two
systems, and a significant positive bias above 7% was observed only in two HbH samples. The HbA1c values obtained by Capillarys 2FP were consistent
with the IFCC targets (relative bias below ± 6%) in all of the eight samples tested by both methods. For screening samples with alpha (α-)
thalassaemia silent/trait or beta (β-) thalassemia trait, the optimal HbA2 cut-off values were ≤ 2.2% and > 2.8%, respectively.
Conclusions: Our results demonstrated the Capillarys 2FP HbA1c system could report an accurate HbA1c value in thalassemia silent/trait, and HbA2
value (≤ 2.2% for α-thalassaemia silent/trait and > 2.8% for β-thalassemia trait) and abnormal bands (HbH and/or HbBart’s for HbH disease, HbF
for β-thalassemia) may provide valuable information for screening.

Ključne riječi

haemoglobin A1c; thalassemia; capillary electrophoresis (CE); haemoglobin A2