hrcak mascot   Srce   HID

Kratko priopćenje

Surface Charging of Zinc Oxide During XPS Examination

M. Mintas ; Institut fur Radiochemie, Kernforschungszentrum Karlsruhe, 7500 Karlsruhe, Postfach 3640, Germany
W. G. Filby ; Institut fur Radiochemie, Kernforschungszentrum Karlsruhe, 7500 Karlsruhe, Postfach 3640, Germany

Puni tekst: engleski, pdf (2 MB) str. 403-406 preuzimanja: 82* citiraj
APA 6th Edition
Mintas, M. i Filby, W.G. (1981). Surface Charging of Zinc Oxide During XPS Examination. Croatica Chemica Acta, 54 (4), 403-406. Preuzeto s https://hrcak.srce.hr/194302
MLA 8th Edition
Mintas, M. i W. G. Filby. "Surface Charging of Zinc Oxide During XPS Examination." Croatica Chemica Acta, vol. 54, br. 4, 1981, str. 403-406. https://hrcak.srce.hr/194302. Citirano 20.01.2020.
Chicago 17th Edition
Mintas, M. i W. G. Filby. "Surface Charging of Zinc Oxide During XPS Examination." Croatica Chemica Acta 54, br. 4 (1981): 403-406. https://hrcak.srce.hr/194302
Harvard
Mintas, M., i Filby, W.G. (1981). 'Surface Charging of Zinc Oxide During XPS Examination', Croatica Chemica Acta, 54(4), str. 403-406. Preuzeto s: https://hrcak.srce.hr/194302 (Datum pristupa: 20.01.2020.)
Vancouver
Mintas M, Filby WG. Surface Charging of Zinc Oxide During XPS Examination. Croatica Chemica Acta [Internet]. 1981 [pristupljeno 20.01.2020.];54(4):403-406. Dostupno na: https://hrcak.srce.hr/194302
IEEE
M. Mintas i W.G. Filby, "Surface Charging of Zinc Oxide During XPS Examination", Croatica Chemica Acta, vol.54, br. 4, str. 403-406, 1981. [Online]. Dostupno na: https://hrcak.srce.hr/194302. [Citirano: 20.01.2020.]

Sažetak
Surface charging of zinc oxide during XPS examination has
been studied. The magnitude of the charge is 0.4-0.6 eV at room
temperature and decreases at higher temperature to almost zero
(nominal 0.1 eV at 300 °C) . Reproducible BE values can only be
achieved after correction for this charging. Possible mechanisms are
briefly discussed.