Skoči na glavni sadržaj

Sažetak sa skupa

https://doi.org/10.15836/ccar2019.57

The magnitude and clinical implications of the interobserver variability of echocardiographic left ventricular ejection fraction measurement

Karlo Golubić ; University Hospital Centre „Sestre milosrdnice“, Zagreb, Croatia
Ivana Tomašić ; University Hospital Centre „Sestre milosrdnice“, Zagreb, Croatia
Ivo Darko Gabrić ; University Hospital Centre „Sestre milosrdnice“, Zagreb, Croatia


Puni tekst: engleski pdf 204 Kb

str. 57-57

preuzimanja: 286

citiraj


Sažetak

The ejection fraction (EF) is the most commonly used echocardiographic measure of the systolic left ventricular (LV) function. Echocardiography has however several limitations in measuring LV volumes and EF since the formulas for quantifications are based on geometrical assumptions, the EF parameters (end-systolic and end-diastolic volumes) are preload dependent and adequate projections are not always possible to obtain. Previous studies have also shown correlation between different
echocardiographic techniques of obtaining EF and even correlation between echocardiography and other methods (MRI, SPECT, CT) but the data on interobserver and intraobserver variability remains scarce.1 In this study we examined the interobserver limits of agreement and the therapeutic implications of the uncertainty obtained by the result.
We examined 22 sequential patients referred to our echocardiography
laboratory for LV function assessment. All echocardiography studies were performed by two independent cardiologists blinded to the measurements of the other observer. 2D methods including the biplane Simpson method and “eyeballing” were conducted, while automatic and semiautomatic 2D and 3D methods were not allowed. Bland-Altman analysis was performed to determine the systematic bias and limits of
agreement (LOA) of EF between different observers (Figure 1).
Next the LOA were used to obtain new “confidence intervals” on a previously gathered set of echocardiograms of 171 heart failure patients. We recorded the number of patients that now had different treatment recommendations according to the current ESC guidelines on heart failure and device therapy i.e. the “new” EF was <35% or >40%. McNemar’s test was used to determine the statistical significance of the difference with a two-tailed P=0.05 taken as a threshold. The bias was calculated to be -2.3 with LOA (7.8 to -12.5) which is in agreement with
similar studies. The number of patients that received “new” treatment recommendations was 99 for the lower LOA and 65 for upper LOA, P<0.001 for both groups. We therefore conclude that the interobserver variability of the EF measurement can lead to significantly different treatment strategies in patients with heart failure. Echocardiographic
methods introducing less variability (like automatic 3D techniques)
should be preferred in determining the EF.

Ključne riječi

echocardiography; left ventricular ejection fraction; interobserver variability; heart failure

Hrčak ID:

220579

URI

https://hrcak.srce.hr/220579

Datum izdavanja:

31.5.2019.

Posjeta: 744 *