Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.11613/BM.2020.020709

Patient-based quality control for glucometers: using the moving sum of positive patient results and moving average

Chun Yee Lim ; Engineering Cluster, Singapore Institute of Technology, Singapore
Tony Badrick ; RCPA Quality Assurance Programs, St Leonards, Sydney, Australia; International Federation of Clinical Chemistry and Laboratory Medicine Committee on Analytical Quality
Tze Ping Loh ; Department of Laboratory Medicine, National University Hospital, Singapore


Puni tekst: engleski pdf 536 Kb

str. 296-306

preuzimanja: 175

citiraj


Sažetak

Introduction: The capability of glucometer internal quality control (QC) in detecting varying magnitude of systematic error (bias), and the potential
use of moving sum of positive results (MovSum) and moving average (MA) techniques as potential alternatives were evaluated.
Materials and methods: The probability of error detection using routine QC and manufacturer’s control limits were investigated using historical
data. Moving sum of positive results and MA algorithms were developed and optimized before being evaluated through numerical simulation for
false positive rate and probability of error detection.
Results: When the manufacturer’s default control limits (that are multiple times higher than the running standard deviation (SD) of the glucometer)
was used, they had 0-75% probability of detecting small errors up to 0.8 mmol/L. However, the error detection capability improved to 20-100%
when the running SD of the glucometer was used. At a binarization threshold of 6.2 mmol/L and block sizes of 200 to 400, MovSum has a 100%
probability of detecting a bias that is greater than 0.5 mmol/L. Compared to MovSum, the MA technique had lower probability of bias detection,
especially for smaller bias magnitudes; MA also had higher false positive rates.
Conclusions: The MovSum technique is suited for detecting small, but clinically significant biases. Point of care QC should follow conventional
practice by setting the control limits according to the running mean and SD to allow proper error detection. The glucometer manufacturers have an
active role to play in liberalizing QC settings and also enhancing the middleware to facility patient-based QC practices.

Ključne riječi

laboratory management; quality control; moving average; moving sum of positive patient results; patient-based quality control

Hrčak ID:

238912

URI

https://hrcak.srce.hr/238912

Datum izdavanja:

15.6.2020.

Posjeta: 488 *