Calibration of the Wiechert seismographs relative to a reference seismometer

Authors

  • Marijan Herak Geophysical Institute, Department of Geophysics, Faculty of Science and Mathematics, University of Zagreb, Zagreb, Croatia
  • Ivo Allegretti Croatian Seismological Survey, Department of Geophysics, Faculty of Science and Mathematics, University of Zagreb, Zagreb, Croatia
  • Davorka Herak Geophysical Institute, Department of Geophysics, Faculty of Science and Mathematics, University of Zagreb, Zagreb, Croatia
  • S. J. Duda Institute of Geophysics, Hamburg University, Hamburg, Germany

Keywords:

Seismograph calibration, historical instrumentation, Wiechert seismograph

Abstract

The horizontal 1000 kg Wiechert seismograph has been calibrated on the basis of a comparison of its records to the ones that were obtained by the reference seismometer. The resulting magnification curve shows notable deviations from the theoretical one. Its most prominent feature is a sudden magnification increase observed for frequencies above 5 Hz, which is attributed to frame oscillations excited by the seismic waves. Also, the observed average response curve exhibits larger magnification for frequencies below the resonance frequency. The results indicate a non-linear behavior of the Wiechert instrument related to the degree of impulsiveness of the excitation.