Skoči na glavni sadržaj

Prethodno priopćenje

A comparative study of the corrosion of wire used in urological treatment under sterilization

W. Walke ; Faculty of Mechanical Engineering, The Silesian University of Technology, Gliwice, Poland
J. Przondziono ; Faculty of Materials Science and Metallurgy, The Silesian University of Technology, Katowice, Poland


Puni tekst: engleski pdf 250 Kb

str. 237-240

preuzimanja: 633

citiraj


Sažetak

The purpose of the tests was to determine whether and how sterilization process of samples made of AISI 316L stainless steel with different strain impacts their corrosion resistance. Tests were made on steel samples that had been electrochemically polished in order to assure proper surface roughness. In order to evaluate the influence of sterilization on physical and chemical properties of steel surface, tests of corrosion resistance were made by means of potentiodynamical method. The tests were made in alternative solution simulating human urine. Recorded anodic polarization curves created the ground for determination of typical parameters describing pitting corrosion resistance, that enabled to evaluate steel wire corrosion behavior under sterilization conditions.

Ključne riječi

AISI 316L stainless steel; corrosion resistance; wire; steam sterilization; urology

Hrčak ID:

74859

URI

https://hrcak.srce.hr/74859

Datum izdavanja:

2.4.2012.

Podaci na drugim jezicima: hrvatski

Posjeta: 1.421 *