Skoči na glavni sadržaj

Izvorni znanstveni članak

Calibration of the vertical measuring system of stylus instrument Perthometer S8P

Vedran Mudronja ; Faculty of Mechanical Engineering and Naval Architecture, Ivana Lučića 1, 10000 Zagreb, Croatia
Gorana Baršić ; Faculty of Mechanical Engineering and Naval Architecture, Ivana Lučića 1, 10000 Zagreb, Croatia
Biserka Runje orcid id orcid.org/0000-0002-7786-0864 ; Faculty of Mechanical Engineering and Naval Architecture, Ivana Lučića 1, 10000 Zagreb, Croatia


Puni tekst: hrvatski pdf 629 Kb

str. 141-145

preuzimanja: 891

citiraj

Puni tekst: engleski pdf 629 Kb

str. 141-145

preuzimanja: 522

citiraj


Sažetak

Measurement result without defined metrological traceability does not represent valid metrological information. National laboratory for length (HMI/FSB-LPMD), as constitutive part of Croatian Metrology Institute (HMI), has a responsibility to assure traceability using national and first line standards for length, roughness and angle. In HMI/FSB-LPMD traceability for roughness parameters is assured using calibrated National roughness standards. In this paper calibration of the vertical measuring system of stylus instrument Perthometer S8P using roughness standard is presented. Calibration has been performed according to the recommendations from DKD Guideline 4-2. Based on the model shown in Directive DKD R 4-2, sheet 1, Annex A measurement uncertainty of a roughness parameter Rz determined using Perthometer S8P has been evaluated.

Ključne riječi

calibration; metrological traceability; roughness standard; stylus instrument

Hrčak ID:

116586

URI

https://hrcak.srce.hr/116586

Datum izdavanja:

21.2.2014.

Podaci na drugim jezicima: hrvatski

Posjeta: 2.535 *