Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.5599/admet.1584

A simple and fast flow injection amperometry for the determination of methimazole in pharmaceutical preparations using an unmodified boron-doped diamond electrode

Adison Meoipun
Kantima Kaewjua
Orawon Chailapakul
Weena Siangproh


Puni tekst: engleski pdf 845 Kb

str. 303-315

preuzimanja: 59

citiraj

Preuzmi JATS datoteku


Sažetak

In this work, an automated flow injection analysis (FIA) connected to a boron-doped diamond electrode (BDDE) was originally developed for the analysis of methimazole in pharmaceutical preparations. At a modification-free BDDE, methimazole was easilly oxidized. For the analysis of the mechanisms occurring at the electrode surface, cyclic voltammetry was employed to evaluate the impact of fundamental experimental parameters, such as pH and scan rate, on the BDDE response. For the quantitative detection, the FIA amperometric approach was constructed and used as a fast and sensitive method. The suggested approach provided a broad linear range of 0.5–50 μmol/L and a low detection limit of 10 nmol/L (signal-to-noise ratio = 3). Furthermore, the BDDE was successfully utilized to quantify methimazole in genuine samples from a variety of medicines, and its performance remained steady after more than 50 tests. The findings of amperometric measurements exhibit excellent repeatability, with relative standard deviations of less than 3.9 and 4.7 % for intra-day and inter-day, respectively. The findings indicated that, compared with traditional approaches, the suggested method has the following advantages: quick analysis time, simplicity, highly sensitive output, and no need for complicated operational processes.

Ključne riječi

Methimazole; boron-doped diamond thin film electrode; flow injection analysis; drug formulations; amperometry

Hrčak ID:

304479

URI

https://hrcak.srce.hr/304479

Datum izdavanja:

4.6.2023.

Posjeta: 206 *




Introduction

Methimazole (2-mercapto-1-methyl imidazole, MMI) is used to treat hyperthyroidism or an overactive thyroid. It is also prescribed before undergoing thyroid surgery or radioactive iodine treatment. The most prevalent cause of hyperthyroidism is Graves’ disease, an autoimmune condition caused by antibodies that bind to receptors on thyroid hormone-producing cells in the thyroid gland, inducing excessive thyroid hormone synthesis. By mixing iodine with a protein called thyroglobulin, an enzyme called peroxidase generates thyroid hormones, such as thyroxine (T4) and triiodothyronine (T3). MMI stops iodine and peroxidase from producing T4 and T3 from their usual interactions with thyroglobulin. Therefore, thyroid hormone synthesis is reduced as a result of this activity. The conversion of T4 to T3 is also hampered by MMI. Because T3 is more effective than T4, it decreases thyroid hormone activity [1,2]. MMI was authorized by the FDA in March 1999 [3]. However, MMI can induce nephritis, liver cirrhosis, skin irritation, allergies, pharyngitis with fever, and other adverse effects [4]. In pharmaceutical, nutrition, and clinical chemistry, accurate drug determination using simple and rapid procedures is a prerequisite.

The analytical methods used for MMI determination are described, including thin-layer chromatography [5], high-performance liquid chromatography with tandem mass spectrometry [6], ultraviolet detection [7,8], spectroscopy [9-11], potentiometry [12], liquid chromatography with amperometric detection [13], and capillary zone electrophoresis [14]. Among these, liquid chromatography is one of the most commonly employed in pharmaceutical research. While it is often utilized in laboratories, there are instances when complicated procedures must be developed and validated. They could also take a lot of time and call for qualified employees. On the other hand, because they are quick, simple, accurate, and exceedingly sensitive, electroanalytical methods have proved effective in detecting a wide range of analytes, including those present in biological and pharmaceutical samples. However, electrochemistry has a limitation in that it is not an automated method causing it not to support routine work. The discovery of a flow-based system led to widespread application in quantitative pharmaceutical analysis. Flow-based technology has been useful and can be an option for routine analysis with a fast response and a substantially lower cost. Moreover, flow injection analysis (FIA) offers the possibility to work in continuous rather than in batch mode, with higher sample throughput and various detectors. Thus, electrochemical techniques have been widely used as a detector in flow injection techniques for pharmaceutical applications [15,16]. One of the most frequent working electrodes used in electrochemical research is the glassy carbon electrode (GCE). Unfortunately, MMI oxidation at the GCE generally has a weak electrochemical response. Therefore, chemically modified electrodes have been frequently documented for the sensitive and selective determination of MMI. Acetylene black/chitosan film [17], MWCNTs [18], MWCNTs/electro-copolymerized cobalt nanoparticle–poly (pivalic acid) composite film [19], and carbon paste electrodes modified with Schiff base complexes of vanadium and cobalt [20,21], nanocomposite of CdS NP–RGO/IL [22], and MWCNT–titanium dioxide nanoparticles [23] are among the modifiers. In several applications mentioned, time was wasted in the modification and pretreatment steps. Furthermore, there is still a significant issue with electrode preparation repeatability.

Based on our expertise, the chemical and physical properties of a boron-doped diamond electrode (BDDE) (low background currents, wide potential window, good resistance to fouling, chemical and mechanical stability, lack of a surface oxide film, and controllable surface termination) have been successfully used in a variety of applications [24-26].

In this regard, BDDE can be realized for the quantification analysis of MMI [27]. However, the difference of this study from the previous work related to the BDDE is that the modification-free or unpretreated BDDE was used to explore the electrochemical property of MMI via cyclic voltammetry as a continuation of our previous work on the usage of the BDDE in applications. As previously mentioned, thus, this is the first time that reported the use of an unmodified BDDE as a detector in an FIA system to determine MMI. The online FIA and the promising characteristics of the BDDE enable the detection of extremely low MMI levels. Good analytical detection performances were observed with detection limits in the nanomolar range, strong sensitivity, outstanding response accuracy, and stability. Therefore, the ideal protocol to work with BDDE for FIA enables the electrochemical pretreatment step and sequential detection to simplify regular applications and routine measurements.

Experimental

Chemicals and reagents

Without additional purification, all compounds were of analytical grade or above. All solutions and subsequent dilutions were prepared in deionized water. Phosphate solutions (0.1 mol/L, pH 5–8) were made by combining 0.1 mol/L potassium dihydrogen phosphate (Merck, Darmstadt, Germany) and 0.1 mol/L sodium hydrogen phosphate (Fluka, Germany). Phosphate solution (pH 2.5) was prepared from 0.1 mol/L potassium dihydrogen phosphate, and the pH was adjusted with orthophosphoric acid (85 %, Carlo Erba). Phosphate solution (pH 9), 0.1 mol/L, was prepared from 0.1 mol/L potassium dihydrogen phosphate, and the pH was adjusted with 0.1 mol/L sodium hydroxide. Prior to use, the standard methimazole (Sigma-Aldrich, Darmstadt, Germany) solutions were prepared daily using the same solution.

Electrodes

The equipment utilized for the diamond film generation and the experimental setup have been extensively documented elsewhere [28]. The films were created by microwave plasma-assisted chemical vapor deposition of BDD thin films on highly conductive n-Si (111) substrates. A layer thickness of around 30 mm was attained after 10 hours of deposition. The normal degree of boron doping in the film was around 1021 cm3, and the nominal B/C atomic ratio in the gas phase was 1:100. Before usage, the BDDE was cleansed with ultrapure water. For the glassy carbon electrode (GCE, (0.07 cm2, Bioanalytical System, Inc.), the pretreatment involved sequential polishing with 1 and 0.05 μm of alumina/water slurries on felt pads, followed by rinsing with ultrapure water prior to use.

Voltammetric study

An Autolab potentiostat 100 (PG100, Metrohm, Switzerland) with a typical three-electrode setup was used to record the electrochemical signals. The planar working BDDE was pressed against a smooth ground joint at the bottom of the cell, isolated by an O-ring. The geometric area of the working electrode was 0.07 cm2. A platinum wire served as the auxiliary electrode, and Ag/AgCl was used as the reference. The placement of the back side of the Si substrate on a brass plate led to an ohmic contact. In the comparative study with the BDDE, a GCE (0.07 cm2, Bioanalytical System, Inc.) was also utilized as a working electrode. Cyclic voltammetry was employed to probe the electrochemical reaction. The electrochemical measurements were performed in a Faradaic cage to reduce electronic noise. All experiments were conducted at room temperature.

Flow injection analysis with amperometric detection

The FIA system included an electrochemical detector (PG100), a peristaltic pump (Ismatec), a thin-layer flow cell (all from Bioanalytical System, Inc.), and an injection port with a 20-μL injection loop (Rheodyne7725). A reagent delivery module regulated the carrier stream, which was 0.1 mol/L phosphate solution (pH 9) at a flow rate of 1 mL/min. To minimize the pulsation caused by the peristaltic pump’s roller alternation, a pulse dampener was used in series. A silicone gasket served as a spacer, and the thin-layer flow cell had an Ag/AgCl reference electrode, a stainless-steel tube as an auxiliary electrode, and an outlet. To decrease electrical noise, the experiments were conducted in a copper Faradaic cage. Before performing the amperometric measurement, a hydrodynamic voltammogram was acquired to obtain the suitable detection potential. After each injection, the peak current and the matching background current were recorded. Hydrodynamic voltammograms were created by plotting this data as a function of applied voltage. The hydrodynamic voltammograms provided a maximum signal-to-background (S/B) ratio since the amperometric measurements were performed at the potential of +0.3 V.

Sample preparations

Tablets containing MMI were acquired from local pharmacies in Bangkok, Thailand. The medication tablet was homogenized in an agar mortar prior to measuring. The accurate amount of the powdered mass analyte was dissolved in 100 mL of 0.1 mol/L phosphate solution (pH 9) by stirring for 10 min. The macromolecule excipient was separated by filtration with Whatman filter paper (No. 1), and the residue was washed three times with phosphate solution (pH 9). To determine the MMI amount of the tablets, the final test solution was diluted with phosphate solution (pH 9) to keep the MMI concentration within the linear dynamic range (0.5–50 μmol/L).

Results and discussion

Voltammetric study

The electrochemical responses of two electrodes (the BDDE and GCE) to MMI were first examined and compared via cyclic voltammetry.Figure 1a presents the voltammetric response of the BDDE in a 0.1 mol/L phosphate solution (pH 9) containing 1 mmol/L MMI, whereasFigure 1b shows the voltammetric response of the GCE in the same circumstances. The response at each electrode exhibited no detectable oxidative signals across the investigated voltage range, according to the cyclic voltammograms obtained before the addition of MMI (0.0–1.2 V). Furthermore, the BDDE consistently exhibited lower capacitance than the GCE when background currents at each electrode were evaluated. The response at the BDDE changed when MMI was added to the solution, with a new well-defined electrochemical signal developing at +0.45 V. The MMI response at the GCE, on the other hand, had two pairs of ill-defined peaks observed as an increase in oxidation current at +0.35 V and +0.98 V. For comparison, the CV response of MMI at the BDDE is more significant and has a stronger anodic peak current than bare GCE. This may be concerned with the fact that a BDDE provided a very lower background current than a GCE, suggesting that the BDDE has shown an efficient catalytic activity toward the electrooxidation of MMI. In addition, no additional reductive processes were identified in the voltage range investigated at each electrode when the scan direction was reversed. These irreversible behaviors provide consistent results with a chemically modified electrode reported previously.

Effect of pH value on the electrochemical behavior of MMI

Investigating the effect of solution pH on the electrochemical responses of MMI is essential for achieving the best results in high responsiveness. Thus, the electrochemical characteristics of MMI were investigated using a phosphate solution with a concentration of 0.1 mol/L as the supporting electrolyte. The tests were conducted at pH levels of 2.5, 5.0, 7.0, 8.0, and 9.0. According to the findings, MMI may be oxidized in both neutral and alkaline conditions. When the pH value increased, the peak potential shifts toward less positive potentials (Figure 2). Additionally, the oxidation of sulfur-containing compounds at the BDDE can explain this electrooxidation. The oxidation process involved the dissociation of the proton on the thiol group, followed by the electrochemical oxidation of the MMI anion. As a result, MMI easily lost the proton at higher pH levels and formed a more stable reduced form. The use of a phosphate solution at pH 9 yielded the largest oxidation current peak. Hence, this pH was selected as the starting point for all the following studies.

Scan rate and concentration dependence

To confirm that this oxidative process is diffusion-controlled, an investigation into the effects of scan rate on the electrochemical signal was carried out. As presented inFigure 3, the electrooxidation responses of MMI varied with the scan rate, with the current plateau rising linearly with the square root of the scan rate (R2 = 0.9988). Such a relationship indicates that MMI oxidation at a BDDE surface is really controlled by the diffusion process.

In order to perform a preliminary study on the analytical effectiveness of MMI at the BDDE, cyclic voltammetry was used to measure the change in voltammetric current as a function of various MMI concentrations. It was discovered that the response of the BDDE was linear over the concentration range of 0.025-3.0 mmol/L, with a limit of detection of 50 μmol/L at signal-to-background (S/B) = 3. The data is depicted inFigure 4. However, the obtained linear range and LOD are insufficient for the quantitative detection of MMI in applications such as biological samples or pharmaceutical preparations. Therefore, the use of a highly sensitive electrochemical technique is required.

Electrochemical behavior of MMI at the BDD electrode

To examine the mechanism of electrooxidation of MMI at the BDDE interface, the electrooxidation behavior of MMI was measured via cyclic voltammetry at various potential scan rates. The oxidation current linearly increases with the square root of the scan rate, as presented inFigure 2. In addition, the obtained linear relationship between Ipa and the ν1/2 (y = 75.513x −1.0473; R2 = 0.9988) was used to calculate the diffusion coefficient of MMI (D0) using Andrieux and Savèant’s theoretical model (Eq. 1):

(1)
ADMET-11-1584-e001.jpg

where F is Faraday’s constant = 96,485 C/mol, R = 8.31447 J/mol·K, T = 298 K, A = 0.07 cm2; Cs is the MMI concentration = 1 × 10−6 mol/cm3, Ip is the oxidative current of MMI, and ν is the scan rate. The calculated diffusion coefficient (D0) of MMI was found to be 1.30 × 10−5 cm2/s. The Randles–Ševčík relationship for the irreversible process (Eq. 2) was employed to calculate the number of electron transfers at the electrode interface:

(2)
ADMET-11-1584-e002.jpg
(3)
ADMET-11-1584-e003.jpg

The factor (1-α)nα was firstly calculated from the linear regression (Eq. 3) between Epa and ln ν (y = 0.0223x + + 0.7576; R2 = 0.9963) and was found to be 1.15. After calculating the important factors, the number of electrons involved in the electrooxidation of MMI was calculated withEq. 2 and was found to be n = 0.92 (nc1).

Moreover, it was found that the peak potential of MMI shifted to the negative potential by increasing pH. This is predictable due to the participation of proton in the oxidation process of MMI [18,19]. The Nernst equation (Eq. 4) was used to describe the relationship between Epa and pH (inset inFigure 2) The linear relationship between the pH values and oxidation peak potentials was plotted (y = −0.0572x + 1.0074; R2 = 0.9652).

(4)
ADMET-11-1584-e004.jpg

where m and n are the number of protons and electrons at the electrode interface, respectively. The slope was found to be 57.2 mV, which is in agreement with the theoretical slope (2.303 mRT/2F) of 59 mV/pH. Based on the calculation, m = n = 1, since the results show that the quantities of electrons and protons exchanged at the electrode surface are equal. Consequently, the possible mechanism of MMI at the BDDE involves one electron and one proton.

Hydrodynamic voltammetry

Hydrodynamic voltammetry is a helpful technique to figure out what potential to use for FIA/amperemetric detection. A standard solution of MMI was repeatedly injected, while the FIA/amperemetry working potential was adjusted at +0.20, +0.30, +0.35, +0.40 and +0.45 V, respectively.Figure 5a shows the hydrodynamic voltammogram of the standard solution, which contains 100 mol/L of MMI. The current response of MMI increased as the applied voltage increased; however, it did not reach its maximum value. To reach the maximum potential point, the S/B ratio was computed (Figure 5a) at each point to construct a graph as a function of S/B ratios and applied potential (Figure 5b). In this work, +0.30 V was selected as the optimum detection potential for subsequence work owing to high sensitivity and low noise. Remarkably, the reference electrode used for cyclic voltammetry and flow system is different. Thus, it may lead to a little different detection potential being obtained between batch and flow systems.

FIA coupled with amperometric detection

After setting the suitable detection potential for the FIA technique, amperometric measurements were carried out in the phosphate solution (pH 9) containing various MMI concentrations in order to get the analytical curve. The FIA current-time response for different MMI concentrations (Figure 6) demonstrated a linear relationship between the current values (at +0.30 V) obtained and the MMI concentrations exhibited from 0.5 to 50 μmol/L with a correlation coefficient of 0.9969.

The linearity is disrupted at concentrations greater than 50 μmol/L. The experiment revealed that the detection limit was found to be 10 nmol/L (signal-to-noise (S/N) ratio = 3). As demonstrated inTable 1, the proposed method gave the lowest detection limit compared to previously published studies, particularly those using chemically modified electrodes. This means that a BDDE may detect MMI with extreme sensitivity while maintaining its speed and ease of use. Furthermore, because it allows for rapid real-time measurements and eliminates the need for electrode cleaning, this system is ideal for pharmaceutical analysis.

Determination of MMI in commercially available pharmaceuticals

The use of the FIA amperometric approach for the assessment of MMI in commercially available drug samples was performed to test the analytical validity of the suggested method. However, the active ingredients of MMI tablets contain lactose monohydrate, magnesium stearate, starch, corn, and talc [35]. These substances are non-electroactive species that cannot be oxidized at the same oxidation potential as MMI at a BDDE. To ensure that the effects of matrices that may be present in real samples can be reduced, the MMI content was calculated using a standard addition method. The recovery of this method was determined by calculating the percentage of recovery after adding known amounts of standard MMI to the sample solutions.Table 2 presents the recovery for intra-day and inter-day from the determination of MMI using the proposed method. On the basis of intra-assay, the method's precision was established. Three concentrations of the added solution (0.33, 0.65, and 1.31 μg/mL) were selected for the investigation to check the results obtained from the low, medium, and high concentrations with respect to the probable range of interest in the samples. The results obtained from 10 injections were within 3.9 % of the relative standard deviation (RSD, %). The RSD values for the day-to-day assays of MMI were also investigated. It was discovered that during a week in the same laboratory, the %RSD measurements did not vary by more than 4.7 %. These findings indicate that BDDE detection was very reproducible and accurate. In addition, the BDDE were tested for their long-term stability in continuous operation (50 injections). They generated highly repeatable responses from day to day, and the responses were also repeatable after several days in the laboratory environment. The intra-day recovery from the proposed technique was between 95.2 and 101.7 %, whereas the inter-day recovery of MMI was between 93.5 and 103.8 %.

As presented inTable 3, for all the five supplements tested, the experimentally determined values were quite close to the manufacturer’s claims. These results were compared with those obtained using the HPLC-UV technique, and statistical analysis was conducted using a paired t-test with a 95 % confidence interval. The null hypothesis was supported by the paired two-tailed test, which yielded a computed t-value of 0.906, less than the crucial t-value (2.776). A summary of the findings is presented inTable 3, the results of the FIA combined with BDDE amperometry are comparable and not significantly different from those obtained using the traditional HPLC-UV method.

Conclusions

A simple and fast electrochemical method for the determination of MMI was successfully developed using an automatic FIA connected to an unmodified BDDE. The results indicate that BDDE is a promising material for further research in terms of repeatability, sensitivity, and low background current. Among other interesting features of the approach, it was described as rapid, simple, precise, and extremely sensitive, making it useful for the MMI analysis and quality control of pharmaceutical products. Furthermore, this method may be successfully used to determine MMI in commercially available medications, with results that match the label and standard procedure. This evidence will help understand specific drug pathways and metabolism, leading to their development in the various phases of pharmaceutical research as well as clinical investigations of the MMI levels.

Acknowledgments

The authors would like to acknowledge the partial financial support from Srinakharinwirot University and the National Research Council of Thailand (NRCT) through grant number N41A640073. The authors would like to thank Enago (www.enago.com) for the English-language review.

Notes

[1] Conflicts of interest Conflicts of interest: All the authors declare no conflict of interest.

References

[1] 

Marchant B.; Lees J.F.; Alexander W.D.. Antithyroid drugs. Pharmacology & Therapeutics, Part B: General and Systematic Pharmacology 3 (1978) 305-348. https://doi.org/10.1016/S0306-039X(78)80001-4. https://doi.org/10.1016/S0306-039X(78)80001-4

[2] 

Nakamura H.; Noh J.Y.; Itoh K.; Fukata S.; Miyauchi A.; Hamada N.. Comparison of methimazole and propylthiouracil in patients with hyperthyroidism caused by Graves’ disease. The Journal of Clinical Endocrinology & Metabolism 92 (2007) 2157-2162. https://doi.org/10.1210/jc.2006-2135. https://doi.org/10.1210/jc.2006-2135

[3] 

Methimazole, https://www.medicinenet.com/methimazole/article.htm) (accessed August 17, 2022).

[4] 

Baretić M.; Balić S.; Gudelj G.. Miastenična kriza kao nuspojava liječenja metimazolom: Prikaz slučaja. Acta Clinica Croatica 49 (2010) 67-71. https://hrcak.srce.hr/56826.

[5] 

Aletrari M.; Kanari P.; Partassides D.; Loizou E.. Study of the British Pharmacopeia method on methimazole (thiamazole) content in carbimazole tablets. Journal of pharmaceutical and biomedical analysis 16 (1998) 785-792. https://doi.org/10.1016/S0731-7085(97)00119-2. https://doi.org/10.1016/S0731-7085(97)00119-2

[6] 

Ito Y.; Sakai H.; Nagao H.; Suzuki R.; Odawara M.; Minato K.. Development of a high-performance liquid chromatography–tandem mass spectrometric method for the determination of Methimazole in human blood matrices. Journal of Chromatography B 1144 (2020) 122083. https://doi.org/10.1016/j.jchromb.2020.122083. https://doi.org/10.1016/j.jchromb.2020.122083

[7] 

Moretti G.; Betto P.; Cammarata P.; Fracassi F.; Giambenedetti M.; Borghese A.. Determination of thyreostatic residues in cattle plasma by high-performance liquid chromatography with ultraviolet detection. Journal of Chromatography B: Biomedical Sciences and Applications 616 (1993) 291-296. https://doi.org/10.1016/0378-4347(93)80397-M. https://doi.org/10.1016/0378-4347(93)80397-M

[8] 

Zakrzewski R.. Determination of methimazole in urine with the iodine-azide detection system following its separation by reversed-phase high-performance liquid chromatography. Journal of Chromatography B 869 (2008) 67-74. https://doi.org/10.1016/j.jchromb.2008.05.021. https://doi.org/10.1016/j.jchromb.2008.05.021

[9] 

El-Bardicy M.; El-Saharty Y.; Tawakkol M.. Determination of carbimazole and methimazole by first and third derivative spectrophotometry. Spectroscopy letters 24 (1991) 1079-1095. https://doi.org/10.1080/00387019108018174. https://doi.org/10.1080/00387019108018174

[10] 

Sanchez-Pedreno C.; Albero M.; Garcia M.; Rodenas V.. Flow-injection spectrophotometric determination of carbimazole and methimazole. Analytica Chimica Acta 308 (1995) 457-461. https://doi.org/10.1016/0003-2670(94)00606-M. https://doi.org/10.1016/0003-2670(94)00606-M

[11] 

García M.S.; Albero M.I.; Sánchez-Pedreño C.; Tobal L.. Kinetic determination of carbimazole, methimazole and propylthiouracil in pharmaceuticals, animal feed and animal livers. Analyst 120 (1995) 129-133. https://doi.org/10.1039/AN9952000129. https://doi.org/10.1039/AN9952000129

[12] 

Pinzauti S.; Papeschi G.; La Porta E.. Potentiometric titration of thiols, cationic surfactants and halides using a solid-state silver-silver sulphide electrode. Journal of pharmaceutical and biomedical analysis 1 (1983) 47-53. https://doi.org/10.1016/0731-7085(83)80007-7. https://doi.org/10.1016/0731-7085(83)80007-7

[13] 

Zhang S.; Sun W.-l.; Zhang W.; Qi W.-y.; Jin L.-t.; Yamamoto K.; Tao S.; Jin J.. Determination of thiocompounds by liquid chromatography with amperometric detection at a Nafion/indium hexacyanoferrate film modified electrode. Analytica Chimica Acta 386 (1999) 21-30. https://doi.org/10.1016/S0003-2670(98)00799-5. https://doi.org/10.1016/S0003-2670(98)00799-5

[14] 

Wang A.; Zhang L.; Zhang S.; Fang Y.. Determination of thiols following their separation by CZE with amperometric detection at a carbon electrode. Journal of pharmaceutical and biomedical analysis 23 (2000) 429-436. https://doi.org/10.1016/S0731-7085(00)00326-5. https://doi.org/10.1016/S0731-7085(00)00326-5

[15] 

Uslu B.; Ozkan S.A.. Electroanalytical methods for the determination of pharmaceuticals: a review of recent trends and developments. Analytical letters 44 (2011) 2644-2702. https://doi.org/10.1080/00032719.2011.553010. https://doi.org/10.1080/00032719.2011.553010

[16] 

Özkan S.A.; Uslu B.; Aboul-Enein H.Y.. Analysis of pharmaceuticals and biological fluids using modern electroanalytical techniques. Critical reviews in analytical chemistry 33 (2003) 155-181. https://doi.org/10.1080/713609162. https://doi.org/10.1080/713609162

[17] 

Yazhen W.. Electrochemical determination of methimazole based on the acetylene black/chitosan film electrode and its application to rat serum samples. Bioelectrochemistry 81 (2011) 86-90. https://doi.org/10.1016/j.bioelechem.2011.04.001. https://doi.org/10.1016/j.bioelechem.2011.04.001

[18] 

Xi X.; Ming L.; Liu J.. Electrochemical determination of thiamazole at a multi-wall carbon nanotube modified glassy carbon electrode. Journal of Applied Electrochemistry 40 (2010) 1449-1454. https://doi.org/10.1007/s10800-010-0122-x. https://doi.org/10.1007/s10800-010-0122-x

[19] 

Kutluay A.; Aslanoglu M.. Multi-walled carbon nanotubes/electro-copolymerized cobalt nanoparticles-poly (pivalic acid) composite film coated glassy carbon electrode for the determination of methimazole. Sensors and Actuators B: Chemical 171 (2012) 1216-1221. https://doi.org/10.1016/j.snb.2012.06.083. https://doi.org/10.1016/j.snb.2012.06.083

[20] 

Fahimeh J.; Loghman M.; Mahmoud R.. Electrocatalytic determination of anti-hyperthyroid drug, using methimazole, a modified carbon-paste electrode. African Journal of Pharmacy and Pharmacology 7 (2013) 269-274. https://doi.org/10.5897/AJPP12.717. https://doi.org/10.5897/AJPP12.717

[21] 

Shahrokhian S.; Ghalkhani M.. Voltammetric determination of methimazole using a carbon paste electrode modified with a schiff base complex of cobalt. Electroanalysis: An International Journal Devoted to Fundamental and Practical Aspects of Electroanalysis 20 (2008) 1061-1066. https://doi.org/10.1002/elan.200704149 https://doi.org/10.1002/elan.200704149

[22] 

Norouzi P.; Gupta V.K.; Larijani B.; Ganjali M.R.; Faridbod F.. A new Methimazole sensor based on nanocomposite of CdS NPs–RGO/IL–carbon paste electrode using differential FFT continuous linear sweep voltammetry. Talanta 127 (2014) 94-99. https://doi.org/10.1016/j.talanta.2014.03.061. https://doi.org/10.1016/j.talanta.2014.03.061

[23] 

Fouladgar M.; Mohammadzadeh S.. Determination of methimazole on a multiwall carbon nanotube titanium dioxide nanoparticle paste electrode. Analytical letters 47 (2014) 763-777. https://doi.org/10.1080/00032719.2013.855782. https://doi.org/10.1080/00032719.2013.855782

[24] 

Thangphatthanarungruang J.; Lomae A.; Chailapakul O.; Chaiyo S.; Siangproh W.. A Low-cost Paper-based Diamond Electrode for Trace Copper Analysis at On-site Environmental Area. Electroanalysis 33 (2021) 226-232. https://doi.org/10.1002/elan.202060305. https://doi.org/10.1002/elan.202060305

[25] 

Lomae A.; Nantaphol S.; Kondo T.; Chailapakul O.; Siangproh W.; Panchompoo J.. Simultaneous determination of β-agonists by UHPLC coupled with electrochemical detection based on palladium nanoparticles modified BDD electrode. Journal of Electroanalytical Chemistry 840 (2019) 439-448. https://doi.org/10.1016/j.jelechem.2019.04.003. https://doi.org/10.1016/j.jelechem.2019.04.003

[26] 

Pungjunun K.; Chaiyo S.; Jantrahong I.; Nantaphol S.; Siangproh W.; Chailapakul O.. Anodic stripping voltammetric determination of total arsenic using a gold nanoparticle-modified boron-doped diamond electrode on a paper-based device. Microchimica Acta 185 (2018) 1-8. https://doi.org/10.1007/s00604-018-2821-7. https://doi.org/10.1007/s00604-018-2821-7

[27] 

Seyidahmet S.; Dönmez F.; Yardım Y.; Şentürk Z.. Simple, rapid, and sensitive electrochemical determination of antithyroid drug methimazole using a boron-doped diamond electrode. Journal of the Iranian Chemical Society 16 (2019) 913-920. https://doi.org/10.1007/s13738-018-1562-1. https://doi.org/10.1007/s13738-018-1562-1

[28] 

Yano T.; Tryk D.; Hashimoto K.; Fujishima A.. Electrochemical behavior of highly conductive boron-doped diamond electrodes for oxygen reduction in alkaline solution. Journal of the Electrochemical Society 145 (1998) 1870. https://doi.org/10.1149/1.1838569 https://doi.org/10.1149/1.1838569

[29] 

Economou A.; Tzanavaras P.D.; Notou M.; Themelis D.G.. Determination of methimazole and carbimazole by flow-injection with chemiluminescence detection based on the inhibition of the Cu (II)-catalysed luminol–hydrogen peroxide reaction. Analytica Chimica Acta 505 (2004) 129-133. https://doi.org/10.1016/S0003-2670(03)00176-4. https://doi.org/10.1016/S0003-2670(03)00176-4

[30] 

Sun J.; Zheng C.; Xiao X.; Niu L.; You T.; Wang E.. Electrochemical detection of methimazole by capillary electrophoresis at a carbon fiber microdisk electrode. Electroanalysis: An International Journal Devoted to Fundamental and Practical Aspects of Electroanalysis 17 (2005) 1675-1680. https://doi.org/10.1002/elan.200403272. https://doi.org/10.1002/elan.200403272

[31] 

Martinez N.A.; Messina G.A.; Bertolino F.A.; Salinas E.; Raba J.. Screen-printed enzymatic biosensor modified with carbon nanotube for the methimazole determination in pharmaceuticals formulations. Sensors and Actuators B: Chemical 133 (2008) 256-262. https://doi.org/10.1016/j.snb.2008.02.025. https://doi.org/10.1016/j.snb.2008.02.025

[32] 

Zakrzewski R.. Determination of methimazole in pharmaceutical preparations using an HPLC method coupled with an iodine-azide post-column reaction. Journal of Liquid Chromatography & Related Technologies® 32 (2008) 383-398. https://doi.org/10.1080/10826070802631451. https://doi.org/10.1080/10826070802631451

[33] 

Molero L.; Faundez M.; del Valle M.A.; del Río R.; Armijo F.. Electrochemistry of methimazole on fluorine-doped tin oxide electrodes and its square-wave voltammetric determination in pharmaceutical formulations. Electrochimica Acta 88 (2013) 871-876. https://doi.org/10.1016/j.electacta.2012.10.142. https://doi.org/10.1016/j.electacta.2012.10.142

[34] 

Jalali F.; Hatami Z.. Fast electrocatalytic determination of methimazole at an activated glassy carbon electrode. Iranian journal of pharmaceutical research: IJPR 15 (2016) 735. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5316251/. PMC5316251

[35] 

METHIMAZOLE tablet https:/dailymed.nlm.nih.gov/dailymed/drugInfo.cfm, (accessed December 14, 2022).

Floating objects

Figure 1. Cyclic voltammograms for a) BDDE and b) GCE versus Ag/AgCl in 1.0 mmol/L MMI in 0.1 mol/L phosphate solution pH 9 (solid lines) and 0.1 mol/L phosphate solution pH 9 (dashed lines). Sweep rate, 0.05 V/s; area of electrode, 0.07 cm2
ADMET-11-1584-g001.jpg
Figure 2. Cyclic voltammogram for 1-mmol/L MMI in 0.1 mol/L phosphate solution at different pH solutions at a BDDE (potential scan rates: 0.05 V/s and area of electrode: 0.07 cm2). The calibration curve of the relationship between potential (V) and pH is also presented in the inset of this figure.
ADMET-11-1584-g002.jpg
Figure 3. Bar graph for 1 mmol/L MMI in 0.1 mol/L phosphate solution (pH 9) at BDDE for a series of potential scan rate; area of electrode, 0.07 cm2. The calibration curve of the relationship between current (μA) and (scan rate)1/2 is also presented in the inset of this figure. The measurement obtained based on three replicates (n=3)
ADMET-11-1584-g003.jpg
Figure 4. Cyclic voltammogram for MMI in 0.1 mol/L phosphate solution (pH 9) at BDDE for a series of methimazole concentrations. The potential scan rate was 0.05 V/s; the area of electrode, was 0.07cm2. The calibration curve is presented in the inset. The measurement obtained based on three replicates (n=3).
ADMET-11-1584-g004.jpg
Figure 5. (a) Hydrodynamic voltammogram of (-•-) 0.1mol/L phosphate solution (pH 9, background current) and (-▲-) 100 μmol/L of MMI in 0.1 mol/L phosphate solution (pH 9) with repeated four injections of analysis using 0.1 mol/L phosphate solution (pH 9) as a carrier solution. (b) The hydrodynamic of the signal-to-background (S/B) ratio. The flow rate was 1 mL/min.
ADMET-11-1584-g005.jpg
Figure 6. The flow injection analysis results of increasing concentrations of methimazole at a BDDE in 0.1 mmol/L PBS at pH 9. Methimazole concentrations: 0.5, 5, 15, 40, and 50 μmol/L. Inset: A plot of anodic peak currents versus increasing concentration of methimazole.
ADMET-11-1584-g006.jpg
Table 1. Comparison of analytical performance for the determination of MMI using different analytical methods
Techniques and detectorsLinearity range, mol/LDetection limit, mol/LRef.
Flow-Injection: chemiluminescence1.75110-5 – 8.75 10-48.75810-6[29]
Capillary Electrophoresis: a carbon fiber micro disk electrode10-7 – 2.0 10-45.00510-8[30]
Amperometry: Screen-printed enzymatic biosensor modified with carbon nanotube7.4710-8 – 6.35 10-55.60510-8[31]
Differential pulse voltammetry: Carbon-paste electrode modified with a chiff base complex of cobalt10-6 – 1.0 10-45.00510-7[21]
HPLC: Spectrophotometer detector5.0510-8 – 5.5 10-77.40710-5[32]
Differential pulse voltammetry: acetylene black/chitosan/GCE10-7 – 2.0 10-52.00210-8[17]
Square wave voltammetry: Fluorine-doped tin oxide electrodes6.0610-6 – 240 10-61.98110-6[33]
Square wave voltammetry: GCE7710-6 – 1.3 10-53.70310-6[34]
Square wave voltammetry: Pretreated anodically BDDE4.38×10−6– 2.19×10−47.45710−7[27]
Amperometry: Flow-injection analysis/unmodified BDDE0.5010-6 – 50 10-610−8This work
Table 2. The intra- and inter-precisions and recoveries of the FIA-BDDE method (n = 10)
SamplesSpiked level, Sg/mLIntra-dayInter-day
Mean of recovery, % (± SD)RSD, %Mean of recovery, % (± SD)RSD, %
Sample 10.3395.3 ± 1.53.993.5 ± 2.14.5
0.6596.5 ± 2.63.497.4 ± 2.33.6
1.31101.4 ± 1.12.6100.8 ± 1.82.8
Sample 20.3398.4 ± 1.23.899.2 ± 0.74.7
0.6599.1 ± 1.62.6101.2 ± 2.23.1
1.31101.2 ± 1.21.5100.7 ± 1.82
Sample 30.3396.4 ± 2.23.893.5 ± 2.44.1
0.65101.6 ± 1.92.6103.7 ± 2.23.1
1.3198.8 ± 1.41.6100.1 ± 1.51.9
Sample 40.3395.4 ± 3.33.594.8 ± 2.73.9
0.65101.2 ± 2.12.1103.8 ± 1.92.9
1.3199.3 ± 1.32.5102.3 ± 1.72.2
Sample 50.3395.2 ± 2.53.393.5 ± 1.84.4
0.65100.8 ± 2.02.999.8 ± 1.92.7
1.31101.7 ± 1.41.2103.8 ± 1.21.3
Table 3. Determination of the MMI levels in different drug samples (n = 5) using the traditional HPLC-UV method and the developed FIA-BDDE method reported herein.
SampleAmount drug label, mg/tabletAmount found, mg/capsule (± SD)Recovery, %
FIA-BDDE aHPLC-UV bFIA-BDDE aHPLC-UV b
Sample 154.93 ± 0.24.88 ± 0.598.6 ± 1.397.6 ± 1.5
Sample 254.90 ± 0.84.93 ± 1.798.0 ± 1.798.6 ± 1.3
Sample 354.99 ± 1.75.22 ± 1.599.8 ± 3.5104.4 ± 2.3
Sample 454.96 ± 2.54.95 ± 2.499.2 ± 2.499.0 ± 1.9
Sample 555.03 ± 1.45.05 ± 1.3100.6 ± 2.7101.0 ± 2.3

[i] a = This developed method

[ii] b = Traditional method


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.