Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.5599/admet.2357

Fast and simple voltammetric sensing of avanafil in the pharmaceutical formulation by using unmodified boron-doped diamond electrode

Hoshyar Saadi Ali
Hemn A. H. Barzani
Yavuz Yardım


Puni tekst: engleski pdf 636 Kb

str. 529-542

preuzimanja: 0

citiraj

Preuzmi JATS datoteku


Sažetak

Abstract
Background and purpose
Erectile dysfunction is a common issue among adult males involving difficulty in maintaining an erection, and it is often treated with fast-acting, low-side-effect drugs like avanafil (AVN), among other phosphodiesterase-5 inhibitors. Hence, developing fast, simple, and sensitive methods to detect AVN is crucial.
Experimental approach
This study conducts an electroanalytical inquiry and provides a new voltammetric method for accurately analyzing AVN utilizing a boron-doped diamond (BDD) electrode without any modifications.
Key results
In the Britton-Robinson buffer (BR, 0.04 mol L-1, pH 4.0), cyclic voltammetry showed a clearly defined and irreversible anodic peak at around +1.44 V relative to Ag/AgCl. The pH of the solution was shown to have an impact on the voltammetric signals of the oxidation peaks. A good linear response for AVN quantification was achieved using square-wave voltammetry. This was done in a 0.04 mol L-1 BR (pH 4.0) solution at a potential of +1.33 V (vs. Ag/AgCl). The method exhibited a wide dynamic range of 0.5 to 30.0 μg mL-1 (1.0 to 62 μmol L-1) and a low limit of detection of 0.14 μg mL-1 (0.29 μmol L-1). The method proposed demonstrated suitability for determining AVN content in pharmaceutical formulations. The accuracy of the approach was demonstrated by comparing the results obtained using the developed method with those achieved using the UV-Vis spectrometry method.
Conclusion
Our method simplifies the analytical process by eliminating the need for electrode modification, reducing both time and resource requirements while enhancing overall feasibility.

Ključne riječi

Electrochemical sensing; drug analysis; square-wave voltammetry

Hrčak ID:

320012

URI

https://hrcak.srce.hr/320012

Datum izdavanja:

5.8.2024.

Posjeta: 0 *




Introduction

Erectile dysfunction (ED) is a prevalent sexual disorder in adult males characterized by chronic or recurring difficulty in achieving and/or sustaining a penile erection adequate for sexual activity [1]. Moderate to severe ED affects an estimated 5-20 % of men throughout their sexual lives. There are a number of phosphordiesterase-5 (PDE5) inhibitors that are employed to treat ED, including vardenafil, tadalafil, sildenafil, and avanafil (AVN). AVN is particularly popular because it operates quickly and has few visible adverse effects compared with the other inhibitors of PDE5 [2]. AVN, (S)-4-((3-chloro-4-methoxybenzyl) amino)-2-(2-(hydroxymethyl) pyrrolidin-1-yl)-N-(pyrimidin-2-ylmethyl) pyrimidine-5- carboxamide is shown inScheme 1 [3]. Specifically, AVN blocks the action of the PDE5 enzyme. The European Medicines Agency (EMA) (2013) and the United States Food and Drug Administration (FDA) (2012) both gave their approval for the treatment of ED. The therapy of ED has made extensive use of AVN. Ejaculatory latency is enhanced because it promotes blockage of the serotonin transporter, such that the postsynaptic cleft becomes more active with serotonin. AVN is distinguished by its quicker initiation of impacts and its reduced pictorial adverse effects compared to other PDE5 inhibitors. AVN exhibits limited solubility in water, leading to its reduced bioavailability of 38 to 41 % [4]. The typical approach to enhance the oral bioavailability of weakly water-soluble medications is by formulating them as nanoparticles [5]. These nanoparticles, as well-defined by Junghanns and Müller (2008), are subdivisions with a size smaller than 1 μm [6].

Scheme 1. Proposed oxidation pathway of AVN on the BDD electrode surface.
ADMET-12-2357-g002.jpg

During the literature review conducted for AVN, it was found that there are only a few analytical techniques available for its evaluation. These methods can be categorized into three primary groups: HPLC, utilized for analyzing AVN in blood samples [7] and pharmaceutical formulations [8-13]; UV spectrophotometry for AVN analysis [14], as well as LC-MS/MS estimation of AVN in human plasma [15-19], LC-quadrupole-TOF-MS/MS [20], and electrochemistry [21-23].

Extensive studies conducted in the past twenty years have revealed that boron-doped diamond (BDD) electrodes, a recently developed carbon-based material, have garnered significant consideration in various fields, including biological and material science, chemical analysis, environmental investigations, among others [24-31]. BDD exhibits remarkable characteristics without undergoing any chemical alteration, in comparison to both noble metals (such as Au and Pt) and common sp2 carbon electrode materials (like pyrolytic graphite, glassy carbon, carbon paste, and others). The BDD electrode offers numerous advantages, including an extended electrochemical potential range in both non-aqueous and aqueous environments, consistent corrosion resistance, low background current, minimal pollutant adsorption, and remarkable mechanical durability [24,28,32,33].

Based on our comprehensive review of the literature, only three research papers in the field of electrochemistry have been used for the estimation of AVN. Notably, all of these investigations employed modified electrodes. Our study is the first to estimate AVN using bare BDD electrodes without any modification. The proposed method's applicability for determining AVN in pharmaceutical formulations was successfully demonstrated. Furthermore, the outcomes achieved for pharmaceutical formulation will undergo comparison via those derived from the UV-spectrophotometric approach application.

Experimental

Reagents

No further processing was performed on the AVN (ReagentPlus®, 98.0 % standard reference) purchased from ChemScene LLC (USA) before its use in this experiment. The samples were prepared utilizing analytical-grade reagents and water was filtered using a Millipore Milli-Q system. The electrolytes used were Britton-Robinson (BR) buffer, phosphate buffer solution (PBS), HNO3 (0.1 mol L-1), and H2SO4 (0.1 mol L-1). The pH range of the buffers was 2 to 12, and the reagents were analytical-grade (obtained from Sigma-Aldrich, Germany). The AVN was dissolved in a 0.1 mol L-1 HCI solution to create a stock solution with a concentration of 1.0 mg mL-1. While not in use, this item was stored in a refrigerator at 4 °C and protected from sunlight in the lab. A diluting of the stock solutions in the provided supporting electrolyte allowed for the measurement of the AVN concentrations.

Instrumentation and measurements

The electrochemical evaluation was carried out on an Autolab Type III apparatus (Metrohm Autolab B.V., Utrecht, The Netherlands), and the data was collected and analysed with the assistance of the GPES software (version 4.9). In order to correct the baseline of all the stripping voltammograms, the moving averages algorithm was employed. Before this, a Savicky and Golay algorithm was utilized to smoothen the voltammograms, with a peak width of 0.01 V. The three-electrode setup employed in the electrochemical cell used BDD as a working electrode (Windsor Scientific Ltd.) with a diameter of 3 mm and a boron doping level of 1000 ppm, an Ag/AgCl (3 mol L-1 NaCl) reference electrode (BAS, Model RE-1) and a Pt counter electrode (BAS, MW-4130, USA). At 25 °C, the pH readings were taken by a WTW inoLab720 pH meter (Xylem Analytics, Germany) that used a combination glass electrode.

Implementing an activation method similar to the ones outlined in the literature [34] is necessary. As per our earlier research work, the BDD surface underwent pretreatment with it [35-37]. Each day of the experiment commenced with subjecting the BDD electrode to anodic (APT) and cathodic (CPT) pretreatment in 0.5 mol L-1 H2SO4. This was done using a dedicated voltammetric cell with an applied voltage of +1.8 and -1.8 V. In the time between each measurement, the equipment was hand-cleaned. In our previous study, we found that the selected activation protocol enhanced the surface activity of the BDD electrode. The electroactive surface area was determined to be 0.036 cm2 [38]. After gently rubbing the electrode for less than a minute with a moist, smooth cleaning cloth (BAS polishing pad), the residual byproducts were cleaned away using deionized water. This protocol ensured the electrode surface is free of AVN and its oxidation byproducts.

The electrochemical oxidation performance of AVN, as well as the reaction kinetics occurring at the BDD electrode within the supporting electrolyte, were initially investigated by cyclic voltammetry (CV). Utilizing the square-wave voltammetry (SWV) mode, we then optimized the experimental parameters, including the supporting electrolyte and SWV settings, while simultaneously enhancing the sensitivity and selectivity of the AVN analysis. Additionally, the technique's analytical performance and practical applicability were evaluated using the same pulse approach, i.e., SWV. Before doing any quantifiable measurement of AVN by SWV, the three electrodes were immersed in electrochemical cell containing AVN and BR buffer (pH 4.0). Subsequently, anodic scanning was conducted using the SWV technique, spanning from 0 to 1.60 V. The quantitative measurement of AVN was achieved by optimised parameters of 12 mV scan increment, 60 mV pulse amplitude, and 50 Hz frequency using the SWV method.

In order to obtain comparison outcomes, a quantitative analysis of AVN in tablets was performed utilizing a UV/Vis spectrophotometric technique. The experiment utilized an AE-S90 UV-Vis spectrophotometer (AE Lab Guangzhou Co., Ltd, China) equipped with a quartz cell of 1 cm in diameter. The recording of absorption spectra was conducted using 0.1 mol L-1 HCl as the blank solution, covering a wavelength range of 200 to 800 nm. The regression equation was generated utilizing a calibration graph that was produced using the absorbance value of AVN at its maximum absorption wavelength of 245 nm as well as its corresponding concentrations in μg mL-1.

Preparation of sample

Avanair® was procured online and utilized as the sample for AVN analysis in the form of a fixed-dose combination tablet. Every tablet contained 100 milligrams of AVN. The method involved using a mortar and pestle to finely grind ten pills at a precise weight. A Variable quantity of the acquired powder, corresponding to 5.0 mg AVN, was weighed and placed within a 50 mL calibrated dark flask with 0.1 mol L-1 HCl. Complete dissolution was achieved by continually stirring for around 5 min. A volume of 100 μL of the solution (comprising 0.04 mol L-1 BR buffer at pH 4.0) was introduced into the voltammetric cell using a micropipette. The analysis was performed after the addition of AVN at concentrations of 1, 2.5, 5, 10, 15, and 20 μg mL-1. The AVN content throughout the sample was determined using the standard addition technique.

The next phase demonstrated the proposed method's applicability to human urine samples. Urine samples were collected from a healthy 33-year-old male volunteer who had not taken any drugs. A 9.0 mL urine sample was placed in a test tube and spiked with 1.0 mL of an AVN standard working solution (1.0 mg mL-1). After thorough mixing for 1 min, 0.05 mL of the mixture was transferred to a voltammetric cell and diluted to 10 mL with the chosen supporting electrolyte (0.04 mol L-1 BR at pH 4.0). The urine sample without AVN served as a blank. AVN detection was then performed using the standard addition method.

Results and discussion

The electrochemical oxidation of AVN on the BDD electrode surface was explored by the CV. Three successive CV cycles were taken for 50 μg mL-1 AVN in a 0.04 mol L-1 BR buffer (pH 4.0) at a potential range from 0 to 1.70 V. The scan rate was 100 mV s-1. On the first anodic scan, AVN exhibited a clear anodic oxidation peak at approximately 1.44 V, as shown inFigure 1A. However, since the reverse scan did not show a cathodic signal for AVN in the same potential range, it can be concluded that AVN's redox activity is irreversible in these experiments. The current peak height decreases with cycling, which can be explained by the passivation of the electrode and the surface contamination resulting from the AVN oxidation.

Figure 1. The BDD electrode was used to evaluate a concentration of 50 μg mL-1 of AVN in BR buffer with a pH of 4.0. Cyclic voltammograms were performed at a scan rate of 100 mV s-1 (A), and the scan rate was adjusted in subsequent experiments to 50, 75, 100, 200, 300, and 400 mV s-1 (B). The recorded scans are shown in sequence by the arrow. We may see the background current as dotted lines. B: The inset displayed the linear dependences of ip with respect to v1/2 and log ip with respect to log v.
ADMET-12-2357-g001.jpg

The kinetics of AVN oxidation at the BDD electrode were determined by employing the CV technique in a 0.04 mol L-1 BR buffer at a pH of 4.0. The scanning rates ranged from 50 to 400 mV s-1, with a total of six scanning rates. This experiment aimed to investigate the effect of potential scan rates (v) on the anodic peak current to 50 μg mL-1 AVN, and the obtained cyclic voltammograms are shown inFig 1B. With an increase in the scan rate, anodic peak current height is increased.

The linear correlation among the square root of the scan rate (v 1/2) and the oxidation peak current (ip) of the AVN was determined using the equation ip / μA = 0.276 v1/2 (mV s-1)1/2 - 0.431, via a correlation coefficient (r) of 0.994. The following equation, with a correlation coefficient of 0.996, was utilized to establish the linear relationship among log ip and log v: log ip / μA = 0.553 log v / mV s−1 - 0.739. The observed slope, approximately 0.553, falls within the expected range of 1.0 to 0.5 for methods influenced by adsorption and diffusion effects. To evaluate the number of electrons (n) involved in AVN oxidation on the BDD electrode, the equation αn = 47.7/(Ep-Ep/2) was used, with Ep (electrochemical potential difference) determined to be 77 mV. From this equation, the number of electrons involved in the reaction can be calculated. With the assumption of charge transfer coefficient (α) being 0.5, n equals 1.24, which approximates 1. This finding aligns with the observations reported in a prior investigation about the oxidation mechanism of AVN [22].

By employing the SWV, we examined the impact of various electrolytes and pH on the AVN oxidation peak current at the BDD electrode. The effects were investigated in BR buffer solution with a pH range of 2.0-12.0 and 5 μg mL-1 AVN concentration. The observed trend inFigure 2A indicates that the peak potentials shift towards less positive values as the solution pH increases from 2.0 to 11.0. From pH 2 to 4, the peak current increased; however, at pH 4, the peak current began to gradually decline. Under these experimental conditions, there was no oxidation peak at pH 12.0. The linear correlation between AVN anodic peak potentials and pH in the range of 3.0 to 8.0 can be described by the equation Ep / V = -0.055 pH + 1.609 (r = 0.991). The data suggests that an equivalent amount of electrons and protons is involved in the reaction mechanism, as seen by the slope of 0.055 V/pH being close to the theoretical value of 0.059 V. According to the suggested oxidation mechanism, a radical cation is produced when AVN is oxidized at the nitrogen atom of the amide group [22,23].Scheme 1 depicts the oxidation mechanism of AVN.

Figure 2. SW voltammograms were used to examine a 5 μg mL-1 AVN solution in BR buffer (pH 2.0-11.0) (A) and different electrolytes at different pH levels (B) at the BDD electrode. SWV parameters: frequency, 50 Hz; step potential, 8 mV; pulse amplitude, 30 mV. The inset in A contains two graphs that depict the correlation between the linear relationships of Ep / V and pH within the ranges of 2-11 (A) and 3.0-8.0 (B).
ADMET-12-2357-g003.jpg

As shown inFigure 2B, the SW voltammograms for different electrolytes are presented. The anodic peak potentials of +1.56, 1.57, 1.39, and 1.15 V were obtained in 0.1 mol L-1 HNO3, H2SO4, PBS pH 2.5, and 7.4, respectively. The corresponding peak currents were 1.92, 1.54, 2.61, and 0.63 μA, consistent with the findings reported in BR buffer. Also, fromFigures 2A and2B, a peak splitting and/or additional current peaks appearing as shoulders are evident in working pHs other than pH 4 and 5. The voltammogram on the BDD electrode with 5 μg mL-1 AVN in BR buffer at pH 4.0 produced the most impressive results, showing the highest peak current and a well-defined, sharp peak (Figure 2 A andB). Consequently, this voltammogram was chosen for further examination.

The sensitivity of two pulse techniques, SWV and differential pulse voltammetry (DPV), in identifying the anodic peak currents of AVN was evaluated. When comparing the anodic peak currents of AVN measured by SWV and DPV, the experimental results showed that the former was around 8.6 times higher (Figure 3). Following this, more investigation will be done using the SWV method.

Figure 3. In the BDD electrode, DPV (a) and SWV measurements were taken at a concentration of 5 5 μg mL-1 AVN in BR buffer (pH 4.0). DPV parameters: modulation amplitude; 50 mV; step potential; 8 mV and modulation time; 0.05 s. SWV parameters: frequency 50 Hz; step potential 8 mV and pulse amplitude 30 mV.
ADMET-12-2357-g004.jpg

SWV parameters, including frequency (f), pulse amplitude (ΔEsw), and step potential (ΔEs), were optimized to achieve the highest peak current and enhance sensitivity. All other factors were kept constant except one. Via the ΔEs and ΔEsw held at 8 mV and 30 mV, respectively, the f value varied in a 25 to 150 Hz range. The anodic peak current increased as the f value increased. On the other hand, the peak width was significantly widened for f > 50 Hz. By applying this finding to SW voltammetric responses, we have shown that it gives higher analytical selectivity. The remaining experiments used f = 50 Hz. In the 30 to 70 mV region, the impact of the ΔEsw (remaining parameters: ΔEs= 8 mV, f = 50 Hz) on the intensity of the oxidation peak current was also measured. The oxidation peak current increased linearly as a function of ΔEsw in the experimental range. A noticeable widening of the SW voltammograms was observed when ΔEsw was raised to values greater than 60 mV. While the ΔEs varied from 6 to 16 mV, and the other parameters remained constant (f = 50 Hz, ΔEsw = 60 mV), the voltammetric signal that was recorded rose unless it reached a value of 12 mV, and then it increased at a slower rate from 12 to 16 mV. On the other hand, the SW curves exhibited a broadening when the ΔEs values were more than 12 mV. That is why ΔEs = 12 mV was selected for further experiments. Hence, the optimal SWV parameters for AVN oxidation on the BDD electrode in a 0.04 mol L-1 BR buffer at pH 4.0 were obtained with f = 50 Hz, ΔEs = 12 mV, and ΔEsw = 60 mV. The aforementioned graphic does not portray these values.

The evaluation of the analytical performance involved a relationship between the oxidation peak current and the AVN concentrations. The evaluation was conducted after optimising the working conditions, encompassing the experimental settings and instrumental variables. In accordance with this, the AVN stock solution was added to the voltammetric cell in small amounts at regular intervals, and the SWV responses were recorded after each addition. Concentration ranging from 0.5 to 30.0 μg mL-1 (1.0 to 62 μmol L-1) was employed to obtain the SW curves. The calibration curve for this purpose is depicted inFigure 4. A strong linear correlation is observed between the SW voltammograms, which exhibit an AVN peak potential of +1.33 V, following the successive addition of AVN: ip / μA = 0.550 C / μg mL-1 - 0.058 (r = 0.999, n = 10).

Figure 4. SW voltammograms for AVN levels of 0.5, 1.0, 2.5, 5.0, 7.5, 10.0, 15.0, 20.0, 25.0 and 30.0 μg mL-1 (1-10) in 0.04 mol L-1 BR buffer at pH 4.0 on the BDD electrode. SWV parameters are as follows: frequency, 50 Hz; step potential, 12 mV; pulse amplitude, 60 mV. Inset shows the corresponding calibration plot for the quantitation of AVN.
ADMET-12-2357-g005.jpg

The variables ip, C, r, and n represent the anodic peak current, AVN concentration, correlation coefficient, and number of experiments, respectively.

The analytical curve determined the limit of detection (LOD) at 0.14 μg mL-1 (0.29 μmol L-1) as well as the limit of quantification (LOQ) at 0.46 μg mL-1 (0.95 μmol L-1). A 3 s/m approach was used to evaluate the LOD. The slope (m) of the corresponding analytical curve and the standard deviation (s) of the 10 least concentration observations within the calibration range were taken into account.

To the best of our information, previous studies have not utilized bare working electrodes for the determination of AVN. Our analysis, employing non-modified BDD electrodes, yielded results with greater sensitivity in terms of the LOD (0.29 μmol L-1) compared to previous outcomes [21]. However, other studies show greater sensitivity than our study [22,23] (Table 1). Nonetheless, the proposed methodology demonstrates superior practicality, cost-effectiveness, and efficiency in measuring AVN compared to earlier research efforts.

Table 1. Comparison among the developed work and the previously reported studies.
AnalyteElectrodeLinear range, μmol L-1LOD, μmol L-1Reference
AVNZnO-NPs/MWCNT/CP2.7–0.330.71[21]
AVN and nimodipineNiO-NPs/poly(sulf)/PGE0.05 – 1.00.016[22]
AVN and doxorubicinTiO2/MWCNTs/CPE0.01 – 8.00.003[23]
AVNBDDE1.0-620.29This study

[i] ZnO-NPs/MWCNT/CP; Zinc oxide nanoparticles multi-walled carbon nanotubes carbon paste electrode, NiO-NPs/poly(sulf)/PGE; NiO nanostructured/ sulfanilamide polymeric film pencil graphite electrode, TiO2/MWCNTs/CPE; titanium dioxide nanoparticles multi-walled carbon nanotubes carbon paste electrode, BDDE; Boron-doped diamond electrode

The precision of the developed approach was evaluated by quantifying its consistency under identical circumstances, both within and across different days. Ten separate measurements of 0.5 μg mL-1 AVN solution were used to evaluate the repeatability of the anodic peak current's amplitude within a single day. The results were reproducible with an RSD of 5.6 %. After that, the RSD was calculated by measuring the peak current response over five consecutive business days. The results showed an inter-day RSD of 7.2 %. These results prove that a BDD electrode is suitable for reproducible determination of AVN in pharmaceutical formulations and samples.

Preliminary tests were conducted using SWV at a concentration of 5 μg mL-1 AVN under identical experimental conditions as the real samples to detect possible interfering compounds. Pharmaceutical formulations or biological components are the primary sources of their presence. The acceptance limit was determined at a concentration that caused an inaccuracy of approximately ±5% in the oxidation peak current of the AVN. Sodium (Na+), potassium (K+), calcium (Ca2+), magnesium (Mg2+), zinc (Zn2+), iron (Fe3+), and titanium (Ti4+) ions, among others, were found to have no discernible impact on the oxidation peak current of AVN while applied in 10-fold excess. The AVN oxidation peaks were also unaffected by sugars such as maltose, glucose, sucrose, and fructose, even in a concentration 10× higher than the concentration of AVN. It was also found that pharmaceutical formulation components such microcrystalline magnesium stearate, cellulose, and cornstarch had little impact on the oxidation current responses of AVN. The effects of ascorbic acid (AA), dopamine (DOP), and uric acid (UA), potentially present in body fluids, were investigated at molar concentrations of 1:1, 1:5, and 1:10 (AVN solution:interfering compound). At UA, DOP, and AA concentrations relevant to this study, we find that individual solutions do not influence the anodic peak currents of AVN (Figure 5). According to the findings, the created approach exhibits high selectivity and opens up possibilities for using it on actual samples.

Figure 5. SW voltammograms of AVN (5.0 μg mL-1) mixture in the existence of (A) equimolar concentration, 5 and 10-fold excess AA, (B) equimolar concentration, 5 and 10-fold excess DOP and (C) equimolar concentration, 5 and 10-fold excess UA. Other operating conditions as indicated inFigure 4.
ADMET-12-2357-g006.jpg

By measuring AVN in pharmaceutical formulations already on the market, we were able to ascertain the method's potential usefulness. Methods for conducting experiments, such as collecting samples and taking measurements, are detailed in the Experimental Section. We examined the tablet samples without extracting, filtering, or evaporating any materials. As shown inFigure 6, the sample and the standards that were used for analysis are shown in the SW voltammograms. Additionally, the graphical assessment is presented, which includes the equation [ip / μA = 0.595 C /μg mL− 1 + 0.614 (r = 0.998)]. The amount of AVN in the pill was found to be 103.2 mg (RSD of 3.8 %), which is quite similar to the manufacturer-reported figure (100.0 mg) after accounting for the sample's consecutive dilutions.

Figure 6. SW voltammograms were obtained for the diluted drug sample (shown by dashed lines) and afterward addition standard amounts of the drug to accomplish final concentrations of 1.0, 2.5, 5.0, 10.0, 15.0, and 20.0 μg mL-1 AVN in a 0.04 mol L-1 BR buffer at pH 4.0. on the BDD electrode. Inset depicts the result of analysis by standard addition method. Other operating conditions as indicated inFigure 4.
ADMET-12-2357-g007.jpg

A voltammetric cell was used to perform spike/recovery studies with commercial samples to confirm that the newly discovered approach could be used in real-world applications. Voltammetric responses were evaluated after adding 10 mL of sample solution to standard AVN solutions with concluding amounts of 1, 5, and 10 μg mL-1, as shown inTable 2. According toTable 2, there was no noticeable impact of the matrix on the proposed method.

Table 2. Pharmaceutical formulation samples spiked with AVN standard solutions, and their recovery values measured using the suggested voltammetric method
Concentration of AVN, μg mL-1Recovery ± RSD, %
AddedaExpectedaFounda,b
01.03- ± 3.8
1.02.031.9596.1 ± 3.5
5.06.036.19102.7 ± 3.1
10.011.0310.8298.1 ± 2.9

[i] aConcentration in the measured solution

[ii] bAverage of three replicate measurements

The results of the developed voltammetric procedure for commercial pharmaceutical formulations were juxtaposed with those successfully acquired using a straightforward UV spectrophotometry method. A typical absorption spectrum of AVN in 0.1 mol L-1 HCl is shown inFigure 7. Calibration studies were conducted over the AVN concentration range of 1.0 to 20.0 μg mL-1 (equivalent to 2.1 to 41 μmol L-1), with measurements taken at 245 nm. The regression equation for the calibration graph was determined as [A = 0.082 C / μg mL-1 + 0.070 (r = 0.998, n = 6)], where A represents the absorbance value of AVN, and C denotes the AVN concentration. Following three replicate measurements, the average AVN content was computed as 101.5 mg per tablet, with an RSD of 1.4 %. Evaluation of the findings suggests similar results were obtained for AVN content using both methods. Furthermore, the results from both the suggested and comparison approaches were subjected to a Student's t-test for validation purposes. There were no significant discrepancies between the values acquired using the two approaches and the genuine, labelled values, since the computed (experimental) t value = 2.32, at a 95 % confidence range, was lower than the tabulated (theoretical) one (2.78, for α = 0.05).

Figure 7. Absorption spectrum for 10 μg mL-1 AVN in 0.1 mol L-1 HCl using 1 cm quartz cell
ADMET-12-2357-g008.jpg

To evaluate the effectiveness of the proposed approach, the analysis was extended to determine AVN in urine samples.Figure 8 shows SW voltammograms for urine sample analysis, including spiked concentrations of 0.5 μg mL-1 in the electrochemical cell. In the unspiked urine sample, an anodic peak appeared around +0.79 V, likely due to UA oxidation. Using the standard addition method, the intensity of the peak increased proportionally with AVN concentration, resulting in a linear calibration plot: ip / μA = 0.681 C / μg mL-1 + 0.315 (r = 0.997), as shown in the inset ofFigure 8. The electrochemical analysis using this approach for AVN determination showed satisfactory results in terms of recovery percentage and RSD in urine samples.Table 3 highlights robust recovery and RSD values, confirming the potential application of this developed method for the direct quantification of AVN in urine samples.

Figure 8. SW voltammograms of the urine sample: the dashed line depicts the diluted urine sample in the presence of 0.5 μg mL-1 AVN, with (1-5) showing the standard additions of 2.5, 5.0, 7.5, 10.0 and 15.0 μg mL−1 AVN in BR buffer at pH 4.0. Inset shows the result of analysis by standard addition method.Figure 4 shows the other operational parameters.
ADMET-12-2357-g009.jpg
Table 3. Using the developed approach, measurement results for the determination and recovery of AVN from the urine sample.
AVD content, μg mL-1Recovery ± RSD, %
AddedFounda
0.500.4692.0 ±4.3

[i] aCalculated by the use of standard addition method.

Conclusions

The current investigation used an unmodified BDD electrode, which presents a significant departure from the previous analytical procedures in the determination of AVN. This novel approach underscores a departure from the conventional reliance on electrode modification, presenting a fresh perspective in AVN analysis. Importantly, our proposed methodology breaks new ground and showcases several distinct advantages over previous research endeavours. Our approach's inherent practicality, cost-effectiveness, and efficiency stand out prominently, offering a streamlined and resource-efficient avenue for AVN measurement. By eliminating the need for electrode modification, our method simplifies the analytical process, reducing both time and resource requirements while enhancing overall feasibility. The proposed method's applicability for determining AVN in pharmaceutical formulations was successfully demonstrated.

Notes

[5] Conflicts of interest Conflict of interest: Authors declare that there is no conflict of interest.

[6] Funding: The authors listed in the manuscript certify that they have no affiliations with or involvement in any organization or entity that has any financial or non-financial interest in the subject matter or materials discussed in this manuscript.

References

[1] 

Shamloul R.; Ghanem H. Erectile dysfunction. The Lancet 381 (2013) 153-165. https://doi.org/10.1016/S0140-6736(12)60520-0 https://doi.org/10.1016/S0140-6736(12)60520-0

[2] 

Burnett A.L. Erectile dysfunction. The Journal of Urology 75 (2006) S25-S31. https://doi.org/10.1016/S0022-5347(05)00309-5 https://doi.org/10.1016/S0022-5347(05)00309-5

[3] 

Abou-Omar M.N.; Annadi A.M.; El Zahar N.M.; Youssef A.O.; Amin M.A.; Attia M.S.; Mohamed E.H. Simultaneous determination of Avanafil and Dapoxetine in human plasma using liquid chromatography/tandem mass spectrometry (LC-MS/MS) based on a protein precipitation technique. RSC Advances 11 (2021) 29797-29806. https://doi.org/10.1039/D1RA05492A https://doi.org/10.1039/D1RA05492A

[4] 

Burke R.M.; Evans J.D. Avanafil for treatment of erectile dysfunction: review of its potential. Vascular Health and Risk Management 1 (2012) 517-523. https://doi.org/10.2147/VHRM.S26712 https://doi.org/10.2147/VHRM.S26712

[5] 

Ashmawy A.Z.G.A.; Alyami M.H.; Eissa N.G.; Balata G.F.; Nahas H.M.E. Oral bioavailability enhancement of doxazosin mesylate: Nanosuspension versus self-nanoemulsifying drug delivery systems. ADMET & DMPK 12 (2024) 167-176. https://doi.org/10.5599/admet.2022 https://doi.org/10.5599/admet.2022

[6] 

Corona G.; Rastrelli G.; Burri A.; Jannini E.A.; Maggi M. The safety and efficacy of Avanafil, a new 2nd generation PDE5i: comprehensive review and meta-analysis. Expert Opinion on Drug Safety 15 (2016) 237-247. https://doi.org/10.1517/14740338.2016.1130126 https://doi.org/10.1517/14740338.2016.1130126

[7] 

Soliman K.A.; Ibrahim H.K.; Ghorab M.M. Effects of different combinations of nanocrystallization technologies on avanafil nanoparticles: in vitro, in vivo and stability evaluation. International Journal of Pharmaceutics 517 (2017) 148-156. https://doi.org/10.1016/j.ijpharm.2016.12.012 https://doi.org/10.1016/j.ijpharm.2016.12.012

[8] 

Jadhav M.V.; Akash A.; Munipalli V.; Singh R.M.; Nayak S.; Vaidhun B. Development and Validation of Reverse Phase High Performance Liquid Chromatography Method for Quantitative Estimation of Avanafil in Tablet Dosage Form. Journal of Pharma Innovation 1(1) (2020) 29-38. https://scholar.google.com/scholar_url?url=https://www.academia.edu/download/64740720/Avanafil_In_Tablet_Dosage_Form.pdf&hl=en&sa=X&ei=neN6Zs_tD7bKy9YPsbqkkAU&scisig=AFWwaeZfZHwf2Xo5NI7fkaECpsDw&oi=scholarr

[9] 

Hegazy M.; Kessiba A.; Abdelkawy M.; El-Gindy A.E. RP-HPLC with time programmed fluorescence detection for quantitation of avanafil and dapoxetine hydrochloride: Application to pharmaceutical dosage form and biological fluid. Journal of Liquid Chromatography & Related Technologies 38 (2015) 1660-1665. https://doi.org/10.1080/10826076.2015.1087860 https://doi.org/10.1080/10826076.2015.1087860

[10] 

Kumar N.; Sangeetha D.; Kalyanraman L.; Sainath K. Stability-indicating HPLC method for simultaneous determination of degradation products and process-related impurities of avanafil in avanafil tablets. Acta Chromatographica 30 (2018) 158-163. https://doi.org/10.1556/1326.2017.00116 https://doi.org/10.1556/1326.2017.00116

[11] 

Zhao M.; Wu X.; Yu Z.; Sun Y.; Liu Z.; Yuan J.; Liu H.; Jin Y. Identification, synthesis and characterization of avanafil process impurities and determination by UPLC. RSC Advances 12 (2022) 9256-9262. https://doi.org/10.1039/D2RA01224C https://doi.org/10.1039/D2RA01224C

[12] 

Shah D.A.; Vegad K.L.; Patel E.D.; Prajapati H.K.; Patel R.N.; Patel Y.K. Analytical method validation for estimation of avanafil and dapoxetine hydrochloride tablet dosage form by HPTLC method. Pharmaceutical and Biological Evaluations 4 (2017) 171-179. https://doi.org/10.26510/2394-0859.pbe.2017.26 https://doi.org/10.26510/2394-0859.pbe.2017.26

[13] 

Lee J.H.; Park O.R.; Yang Y.J.; Han J.H.; Jung E.J.; Kim N.S.; Baek S.Y. Detection of 94 PDE-5is and Their Analogs Including N-Desmethylthiosildenafil in Various Formulations of Dietary Supplements and Food Samples Using HPLC and LC-Q-TOF/MS. Journal of Chromatographic Science 60 (2022) 953-962. https://doi.org/10.1093/chromsci/bmac015 https://doi.org/10.1093/chromsci/bmac015

[14] 

Patel M.N.; Kothari C.S. Multivariate approaches for simultaneous determination of avanafil and dapoxetine by UV chemometrics and HPLC-QbD in binary mixtures and pharmaceutical product. Journal of AOAC International 99 (2016) 649-663. https://doi.org/10.5740/jaoacint.15-0259 https://doi.org/10.5740/jaoacint.15-0259

[15] 

Jung J.; Choi S.; Cho S.H.; Ghim J.L.; Hwang A.; Kim U.; Kim B.S.; Koguchi A.; Miyoshi S.; Okabe H.; Bae K.S. Tolerability and pharmacokinetics of avanafil, a phosphodiesterase type 5 inhibitor: a single-and multiple-dose, double-blind, randomized, placebo-controlled, dose-escalation study in healthy Korean male volunteers. Clinical Therapeutics 32 (2010) 1178-1187. https://doi.org/10.1016/j.clinthera.2010.06.011 https://doi.org/10.1016/j.clinthera.2010.06.011

[16] 

Abou-Omar M.N.; Annadi A.M.; El Zahar N.M.; Youssef A.O.; Amin M.A.; Attia M.S.; Mohamed E.H. Simultaneous determination of Avanafil and Dapoxetine in human plasma using liquid chromatography/tandem mass spectrometry (LC-MS/MS) based on a protein precipitation technique. RSC Advances 11 (2021) 29797-29806. https://doi.org/10.1039/D1RA05492A https://doi.org/10.1039/D1RA05492A

[17] 

Er E.O.; Akkaya E.; Özbek B.; Bakırdere S. Development of an analytical method based on citric acid coated magnetite nanoparticles assisted dispersive magnetic solid-phase extraction for the enrichment and extraction of sildenafil, tadalafil, vardenafil and avanafil in human plasma and urine prior to determination by LC-MS/MS. Microchemical Journal 147 (2019) 269-276. https://doi.org/10.1016/j.microc.2019.03.043 https://doi.org/10.1016/j.microc.2019.03.043

[18] 

Kammoun A.K.; Khedr A.; Ahmed O.A. LC-MS/MS determination of avanafil and its metabolites in rat plasma and brain: pharmacokinetic study after oral administration and transdermal film application. RSC Advances 10 (2020) 9407-9413. https://doi.org/10.1039/D0RA00569J https://doi.org/10.1039/D0RA00569J

[19] 

Ibrahim A.E.; Hashem H.; Elhenawee M.; Saleh H. Core–shell particles and monolithic columns; tools for simultaneous LC analysis of avanafil, sildenafil, apomorphine, trazodone, yohimbine, tramadol and dapoxetine in pharmaceutical dosage forms, counterfeit products and human plasma. RSC Advances 10 (2020) 1379-1387. https://doi.org/10.1039/C9RA08717F https://doi.org/10.1039/C9RA08717F

[20] 

Öztürk Er E.; Özbek B.; Bakırdere S. Accurate and sensitive determination of sildenafil, tadalafil, vardenafil, and avanafil in illicit erectile dysfunction medications and human urine by LC with quadrupole-TOF-MS/MS and their behaviors in simulated gastric conditions. Journal of Separation Science 42 (2019) 475-483. https://doi.org/10.1002/jssc.201800734 https://doi.org/10.1002/jssc.201800734

[21] 

Ali M.B.; Abdel-Raoof A.M.; Hendawy H.A.; Talaat W.; Omran G.A.; Morshedy S. An Eco-Friendly Solid-State Electrode Modified With ZnO Nanoparticles Decorated With MWCNT as an Electrochemical Sensor for the Determination of Avanafil in Pure Form, Dosage Form and Human Plasma. Journal of the Electrochemical Society 168 (2021) 087510. https://doi.org/10.1149/1945-7111/ac1d7d https://doi.org/10.1149/1945-7111/ac1d7d

[22] 

Ali A.M.B.H.; Abdel-aal F.A.; Rageh A.H.; Mohamed A.M.I. Hybrid NiO nanostructured/sulfanilamide polymeric film for studying possible pharmacokinetic interaction between avanafil and nimodipine in real human serum by their simultaneous determination using square-wave voltammetry. Microchemical Journal 172 (2022) 106895. https://doi.org/10.1016/j.microc.2021.106895 https://doi.org/10.1016/j.microc.2021.106895

[23] 

Ali A.M.B.H.; Rageh A.H.; Abdel-aal F.A.; Mohamed A.M.I. Anatase titanium oxide nanoparticles and multi-walled carbon nanotubes-modified carbon paste electrode for simultaneous determination of avanafil and doxorubicin in plasma samples. Microchemical Journal 185 (2023) 108261. https://doi.org/10.1016/j.microc.2022.108261 https://doi.org/10.1016/j.microc.2022.108261

[24] 

Peckova K.; Musilova J.; Barek J. Boron-doped diamond film electrodes-new tool for voltammetric determination of organic substances. Critical Reviews in Analytical Chemistry 3 (2009) 148e172. https://doi.org/10.1080/10408340903011812 https://doi.org/10.1080/10408340903011812.

[25] 

Yence M.; Cetinkaya G.; Ozcelikay J.; Kaya S.I.; Ozkan S.A. Boron-doped diamond electrodes: recent developments and advances in view of electrochemical drug sensors. Critical Reviews in Analytical Chemistry 52 (2022) 1122-1138. https://doi.org/10.1080/10408347.2020.1863769 https://doi.org/10.1080/10408347.2020.1863769

[26] 

Yang N.; Foord J.S.; Jiang X. Diamond electrochemistry at the nanoscale: a review. Carbon 99 (2016) 90e110. https://doi.org/10.1016/j.carbon.2015.11.061 https://doi.org/10.1016/j.carbon.2015.11.061.

[27] 

Cobb S.J.; Ayres Z.J.; Macpherson J.V. Boron doped diamond: a designer electrode material for the twenty-first century. Annual Review of Analytical Chemistry 11 (2018) 463-484. https://doi.org/10.1146/annurev-anchem-061417-010107 https://doi.org/10.1146/annurev-anchem-061417-010107

[28] 

Muzyka K.; Sun J.; Fereja T.H.; Lan Y.; Zhang W.; Xu G. Boron-doped diamond: current progress and challenges in view of electroanalytical applications. Analytical Methods 11 (2019) 397e414. https://doi.org/10.1039/C8AY02197J https://doi.org/10.1039/C8AY02197J.

[29] 

Lourencao B.C.; Brocenschi R.F.; Medeiros R.A.; Fatibello-Filho O.; Rocha-Filho R.C. Analytical applications of electrochemically pretreated boron-doped diamond electrodes. ChemElectroChem 7 (2020) 1291e1311. https://doi.org/10.1002/celc.202000050 https://doi.org/10.1002/celc.202000050

[30] 

Meoipun A.; Kaewjua K.; Chailapakul O.; Siangproh W. A simple and fast flow injection amperometry for the determination of methimazole in pharmaceutical preparations using an unmodified boron-doped diamond electrode. ADMET & DMPK 11 (2023) 303-315. https://doi.org/10.5599/admet.1584 https://doi.org/10.5599/admet.1584

[31] 

Khadhri N.; Mosbah M.B.; Rabaaoui N.; Khadimallah M.A.; Moussaoui Y. Advancements of Electrochemical Removal of o-Methylphenol from Aqueous Using BDD Anode Compared to Pt One: Kinetics and Mechanism Determination. Iranian Journal of Chemistry and Chemical Engineering 40 (2021) 1446-1456. https://doi.org/10.30492/ijcce.2020.39880 https://doi.org/10.30492/ijcce.2020.39880

[32] 

Švorc Ľ.; Stanković D. M.; Mehmeti E.; Kalcher K. Sensitive electrochemical determination of yohimbine in primary bark of natural aphrodisiacs using boron-doped diamond electrode. Analytical Methods 6 (2014) 4853-4859. https://doi.org/10.1039/C4AY00704B https://doi.org/10.1039/C4AY00704B

[33] 

Cinková K.; Kianičková K.; Stanković D. M.; Vojs M.; Marton M.; Švorc Ľ. The doping level of boron-doped diamond electrodes affects the voltammetric sensing of uric acid. Analytical Methods 10 (2018) 991-996. https://doi.org/10.1039/C7AY02720F https://doi.org/10.1039/C7AY02720F

[34] 

Sarakhman O.; Dubenska L.; Svorc L. First voltammetric behavior study of non-narcotic analgesic drug nefopam and its reliable determination on boron-doped diamond electrodes. Journal of Electroanalytical Chemistry 858 (2020) 113759. https://doi.org/10.1016/j.jelechem.2019.113759 https://doi.org/10.1016/j.jelechem.2019.113759

[35] 

Ali H.S.; Yardım Y. Simultaneous estimation of total phenolic and alkaloid contents in the tea samples by utilizing the catechin and caffeine oxidation signals through the square-wave voltammetry technique. Food Chemistry 441 (2024) 138262. https://doi.org/10.1016/j.foodchem.2023.138262 https://doi.org/10.1016/j.foodchem.2023.138262

[36] 

Barzani H.A.; Ali H.S.; Yunusoğlu O.; Yardım Y. Sensing ivacaftor accomplished using the square-wave voltammetric technique with the assistance of a cationic surfactant on a boron-doped diamond electrode. Diamond and Related Materials 144 (2024) 110932. https://doi.org/10.1016/j.diamond.2024.110932 https://doi.org/10.1016/j.diamond.2024.110932

[37] 

Pınar P.T.; Barzani H.A.; Ali H.S.; Yardım Y. Quantification of ferulic acid using square-wave voltammetric method at an unmodified boron-doped diamond electrode. Monatshefte für Chemie-Chemical Monthly 154 (2023) 1225-1233. https://doi.org/10.1007/s00706-023-03126-8 https://doi.org/10.1007/s00706-023-03126-8

[38] 

Allahverdiyeva S.; Keskin E.; Talay Pınar P.; Yunusoğlu O.; Yardım Y.; Şentürk Z. Electroanalytical investigation and determination of hepatitis C antiviral drug ledipasvir at a non-modified boron-doped diamond electrode. Diamond and Related Materials 108 (2020) 107962. https://doi.org/10.1016/j.diamond.2020.107962 https://doi.org/10.1016/j.diamond.2020.107962


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.