Constant Cardiac Output Monitoring Using the Picco and Lidco Methods Versus Pak in Septic Patients: When to Do Calibration?

Mladen Širanović
Josip Kovač
Aleksandar Gopčević
Mijo Kelečić
Marinko Vučić
Nataša Kovač
Bojan Rode


Puni tekst: engleski pdf 205 Kb

str. 267-272

preuzimanja: 2.098

citiraj


Sažetak

The accuracy of cardiac output measurement by two most widely used methods of less invasive hemodynamic monitoring and by the standard technique of thermodilution with pulmonary catheter was assessed. The measurements were carried out in septic surgical patients immediately after and between system calibrations. Study results showed satisfactory compatibility of measurements performed by the two methods and by pulmonary catheter in both phases, thus system calibration being recommendable in hemodynamically unstable septic patients.

Ključne riječi

Cardiac output – physiology; Monitoring, physiologic – methods; Sepsis; Intensive care units