Sažetak
Important objectives of External Quality Assessment (EQA) is to detect analytical errors and make corrective actions. The aim of this paper is to describe knowledge required to interpret EQA results and present a structured approach on how to handle deviating EQA results. The value of EQA and how the EQA result should be interpreted depends on five key points: the control material, the target value, the number of replicates, the acceptance limits and between lot variations in reagents used in measurement procedures. This will also affect the process of finding the sources of errors when they appear. The ideal EQA sample has two important properties: it behaves as a native patient sample in all methods (is commutable) and has a target value established with a reference method. If either of these two criteria is not entirely fulfilled, results not related to the performance of the laboratory may arise. To help and guide the laboratories in handling a deviating EQA result, the Norwegian Clinical Chemistry EQA Program (NKK) has developed a flowchart with additional comments that could be used by the laboratories e.g. in their quality system, to document action against deviations in EQA. This EQA-based trouble-shooting tool has been developed further in cooperation with the External quality Control for Assays and Tests (ECAT) Foundation. This flowchart will become available in a public domain, i.e. the website of the European organisation for External Quality Assurance Providers in Laboratory Medicine (EQALM).
Ključne riječi
external quality assessment; commutability, reference method; EQA error; between lot variation
Hrčak ID:
176197
URI
Datum izdavanja:
15.2.2017.
Posjeta: 2.691 *