Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.2478/aiht-2021-72-3468

Application of a new in situ calibration technique for gamma spectrometry and comparison of in situ and laboratory measurements

Davor Rašeta ; Institute for Medical Research and Occupational Health, Zagreb, Croatia
Branko Petrinec orcid id orcid.org/0000-0001-9272-3828 ; Institute for Medical Research and Occupational Health, Zagreb, Croatia
Dinko Babić ; Institute for Medical Research and Occupational Health, Zagreb, Croatia
Marko Šoštarić ; Institute for Medical Research and Occupational Health, Zagreb, Croatia


Puni tekst: engleski pdf 687 Kb

str. 29-35

preuzimanja: 287

citiraj

Puni tekst: hrvatski pdf 687 Kb

str. 35-35

preuzimanja: 208

citiraj


Sažetak

In situ gamma ray spectrometry was developed to quickly measure large areas of land following nuclear accidents. However, a proper calibration of detectors for in situ measurements is a long and complicated process. One tool designed to make this calibration quick is the InSiCal software. We compared 5,000 s in situ measurements with two different HPGe detectors calibrated using the InSiCal software and laboratory measurements of samples collected at the same locations. Our findings suggest that in situ gamma spectrometry using InSiCal software can provide reasonably accurate data, but some improvements are needed.

Ključne riječi

emergency; HPGe spectrometers; InSiCal software; radiation; radionuclide measurements

Hrčak ID:

253760

URI

https://hrcak.srce.hr/253760

Datum izdavanja:

16.3.2021.

Podaci na drugim jezicima: hrvatski

Posjeta: 1.526 *