Skip to the main content

Original scientific paper

Xps analysis of biomaterials using a new method of integrated charge compensation

Kevin S. Robinson ; Kratos Analytical Ltd., Urmston, Manchester M41 7LD, England
Hans Schmiedel ; VSI Vacuum Science Instruments GmbH, 65307 Bad Schwalbach, Germany


Full text: english pdf 251 Kb

versions

page 49-54

downloads: 50

cite


Abstract

The major problem in X-ray photoelectron spectroscopy of polymeric and other insulating materials is that of ``charge neutralisation'' or `` charge compensation''. Due to variable potentials on the surface of the investigated specimen broadening of peaks, or even multiple peaks, are observed. Various methods are applied to achieve the charge compensation. The AXIS series of instruments provide a secure and reliable charge compensation method, which uses the characteristic properties of the magnetic lens. The system enables confident determination and use of small binding energy shifts in both spectroscopy and imaging.

Keywords

Hrčak ID:

299596

URI

https://hrcak.srce.hr/299596

Publication date:

2.5.1996.

Article data in other languages: croatian

Visits: 185 *