Skoči na glavni sadržaj

Kratko priopćenje

Comparison of Two Methods Using Atomic Absorption Spectrometry for Determination of Selenium in Food

Marijana Matek
Maja Blanuša
Jerica Grgić


Puni tekst: engleski pdf 91 Kb

str. 283-288

preuzimanja: 5.664

citiraj


Sažetak

This study describes and compares two methods of atomic absorption spectrometry (AAS) for the selenium (Se) analysis in food: electrothermal AAS (ET AAS) and hydride generation method of AAS (HG AAS). The accuracy of the two methods was established by analysing two biological reference materials: Wheat Flour 1567a and Bovine Liver 1577b from the National Institute of Standards and Technology, USA. Good agreement with certified values was obtained for both methods. The accuracy of ET AAS method is 109% and 103% for Wheat Flour and Bovine Liver, respectively. The respective accuracies for HG AAS method were 88% and 87%. The detection limit obtained for ET AAS was 1µg Se/L and for HG AAS 0.02 µg Se/L. The repeatability of ET AAS method was 5–11% and that of HG AAS 14–17%. Both methods are similar in accuracy and repeatability, but hydride generation (HG AAS) is more sensitive than graphite furnace technique (ET AAS) for determination of selenium in food.

Ključne riječi

accuracy; atomic absorption spectrometry; certified values; detection limit; electrothermal atomic absorption spectrometry; food; hydride generation method; repeatability; selenium

Hrčak ID:

2713

URI

https://hrcak.srce.hr/2713

Datum izdavanja:

14.9.1999.

Podaci na drugim jezicima: hrvatski

Posjeta: 7.421 *