Skip to the main content

Original scientific paper

https://doi.org/10.15255/CABEQ.2014.2164

Application of a New Potentiometric Sensor for Determination of Anionic Surfactants in Wastewater

O. Galović ; Department of Chemistry, Josip Juraj Strossmayer University of Osijek, Ulica cara Hadrijana 8/A, HR-31000 Osijek, Croatia
M. Samardžić ; Department of Chemistry, Josip Juraj Strossmayer University of Osijek, Ulica cara Hadrijana 8/A, HR-31000 Osijek, Croatia
S. Petrušić ; Department of Chemistry, Josip Juraj Strossmayer University of Osijek, Ulica cara Hadrijana 8/A, HR-31000 Osijek, Croatia
M. Sak-Bosnar ; Department of Chemistry, Josip Juraj Strossmayer University of Osijek, Ulica cara Hadrijana 8/A, HR-31000 Osijek, Croatia


Full text: english pdf 1.502 Kb

page 307-313

downloads: 897

cite


Abstract

A new screen-printed potentiometric microsensor (SPMS) for the determination of anionic surfactants (ASs) based on the dimethyldioctadecylammonium-tetraphenylborate
(DDA-TPB) ion pair as a sensing material was applied to real industrial effluents. The sensor is accurate, inexpensive, simple and very sensitive. The solutions of cetylpyridinium chloride (CPC) of the concentrations down to 5· 10–5 mol dm–3 were used as a titrant.
Standard addition method was used for checking the precision and accuracy of measurements providing satisfactory results with recoveries between 98.3 to 102.0 % in pure AS solutions and 99.0 to 114.5 % in wastewater. Standard spectrophotometric method Methylene Blue Active Substances (MBAS) was used as control for comparison of the results obtained with the new sensor. The results obtained using SPMS were compared with those obtained by the polymeric membrane ion-selective electrodes (ISEs), containing
the same sensing material, and exhibited good mutual agreement.

Keywords

anionic surfactant; screen-printed; microsensor; potentiometric; wastewater