Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy

Dora Vuljanić
Ana Dojder
Valentina Špoljarić
Andrea Saračević
Lora Dukić
Jasna Leniček Krleža
Jelena Vlašić Tanasković
Ivana Maradin
Ana Grzunov
Željka Vogrinc
Ana-Maria Šimundić


Full text: english pdf 118 Kb

page 123-132

downloads: 806

cite


Abstract

Introduction: Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic
errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks
in Croatia, examine their accuracy for glucose and total protein and to test their repeatability.
Materials and methods: A total of 75 urine samples were used to examine comparability and accuracy of 12 dipstick brands (Combur 10 TestM,
ChoiceLine 10, Combur 10 TestUX, ComboStik 10M, ComboStik 11M, CombiScreen 11SYS, CombiScreen 10SL, Combina 13, Combina 11S, Combina
10M, UriGnost 11, Multistix 10SG). Agreement between each dipstick and the reference (Combur 10 TestM) was expressed as kappa coefficient
(acceptable κ ≥ 0.80). Accuracy for glucose and total protein was tested by comparison with quantitative measurements on analysers: AU400
(Beckman Coulter, USA), Cobas 6000 c501 (Roche Diagnostics, Germany) and Architect plus c4000 (Abbott, USA). Repeatability was assessed on 20
replicates (acceptable > 90%).
Results: Best agreement was achieved for glucose, total protein and nitrite (11/11, k > 0.80) and the lowest for bilirubin (5/5, k < 0.60). Sensitivities
for total protein were 41-75% (AU400) and 56-92% (Cobas and Architect); while specificities were 41-75% (AU400, Cobas, Architect). Dipsticks’
sensitivity and specificity for glucose were 68-98%. Most of the dipsticks showed unacceptable repeatability (6/12, < 90%) for one parameter, most
prominently for pH (3/12, < 90%).
Conclusions: Most commonly used dipsticks in Croatia showed low level of agreement between each other. Moreover, their repeatability varies
among manufacturers and their accuracy for glucose and proteins is poor.

Keywords

verification; urine dipsticks; comparability; accuracy; repeatability