Skip to the main content

Other

https://doi.org/10.15836/ccar2022.176

Fluoroless cryoballoon ablation of atrial fibrillation using the novel mapping system

Ana Jordan orcid id orcid.org/0000-0003-4996-2203
Ante Lisičić orcid id orcid.org/0000-0002-4365-9652
Sanda Sokol Tomić orcid id orcid.org/0000-0002-4551-9231
Nikša Bušić
Ivica Benko orcid id orcid.org/0000-0002-1878-0880
Fran Rode orcid id orcid.org/0000-0002-8787-2455
Šime Manola orcid id orcid.org/0000-0001-6444-2674
Nikola Pavlović orcid id orcid.org/0000-0001-9187-7681


Full text: english pdf 131 Kb

page 176-176

downloads: 147

cite

Download JATS file


Abstract

Keywords

cryoablation; flouroless; pulmonary vein isolation

Hrčak ID:

286928

URI

https://hrcak.srce.hr/286928

Publication date:

8.12.2022.

Visits: 517 *



Introduction: Pulmonary vein isolation (PVI) is the cornerstone of atrial fibrillation (AF) ablation. Radiofrequency point-by-point ablation (RFA) in conjunction with 3D mapping systems uses minimal or no fluoroscopy, while cryoballon ablation (CBA) still requires longer use of fluoroscopy. (1) KODEX-EPD (3D mapping system) is a cardiac imaging system that can be used to guide CBA. With its ability to assess the dielectric properties of structures around an inserted catheter, 3D maps can be made, and the system can assess the occlusion of PVs without the use of fluoroscopy or iodine contrast (2). The aim of our study was to test the feasibility and safety of this relatively new dielectric imaging system in reducing fluoroscopy and contrast use during CBA.

Patients and Methods: Consecutive patients undergoing CBA were enrolled with the intention to perform flouroless procedures. The KODEX-EPD navigation system was used to image the anatomy of the right atrium (RA). A single transseptal puncture was performed guided by intracardiac echocardiography. Mapping of the pulmonary veins and LA was performed by navigation with an inner lumen circular mapping catheter. The degree of PV occlusion with the inflated CB was verified using the “occlusion tool” software module and additionally with the use of saline injection.

Results: During the two months, 15 consecutive patients undergoing CBA were enrolled (age 60±11; 6 women; mean left atrial diameter 39±4 mm). The average procedural and fluoroscopy times were 105±32 minutes and 3.5 ±3 1.8 minutes, respectively. In 7 patients (46,6%) procedure was performed with no use of fluoroscopy, while in 8 (53,4%) fluoroscopy was used due to the difficult crossing of the interatrial septum. No fluoroscopy was used after achieving LA access in all patients. In all patients, PVI was achieved with a mean total freezing time of 18.5± 4 minutes. PV occlusion was verified using the system’s occlusion tool successfully in all but one patient in whom 16 ml of iodine contrast was used to verify the PV occlusion. No complications have occurred.

Conclusion: Although still a small number of patients were included, the results show promising effects of this novel electroanatomic system in performing CBA with no or minimal use of fluoroscopy and a significant reduction in the use of iodine contrast.

LITERATURE

1 

Tovia Brodie O, Rav-Acha M, Wolak A, Ilan M, Orenstein DJ, Abuhatzera S, et al. Anatomical accuracy of the KODEX-EPD novel 3D mapping system of the left atrium during pulmonary vein isolation: A correlation with computer tomography imaging. J Cardiovasc Electrophysiol. 2022 April;33(4):618–25. https://doi.org/10.1111/jce.15391 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/35118762

2 

Schillaci V, Stabile G, Arestia A, Shopova G, Agresta A, Poggi S, et al. Fluoroscopy and contrast media use in cryoballoon ablation of atrial fibrillation using a novel imaging system. Heart Vessels. 2022 January;37(1):115–20. https://doi.org/10.1007/s00380-021-01902-2 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/34240266