Original scientific paper

https://doi.org/10.15644/asc58/1/3

Monowave vs. Polywave Light – Curing Units: Effect on Light Transmission of Composite without Alternative Photoinitiators

Danijela Marović orcid id orcid.org/0000-0003-3104-5749 ; Department of Endodontics and Restorative Dentistry, University of Zagreb School of Dental Medicine, Zagreb, Croatia
Paulina Daničić orcid id orcid.org/0009-0009-9102-9848 ; Student of the University of Zagreb School of Dental Medicine, Zagreb, Croatia
Gloria Bojo orcid id orcid.org/0009-0005-3178-4121 ; Student of the University of Zagreb School of Dental Medicine, Zagreb, Croatia
Matej Par orcid id orcid.org/0000-0002-2846-1840 ; Department of Endodontics and Restorative Dentistry, University of Zagreb School of Dental Medicine, Zagreb, Croatia
Zrinka Tarle ; Department of Endodontics and Restorative Dentistry, University of Zagreb School of Dental Medicine, Zagreb, Croatia


Full text: croatian pdf 1.578 Kb

page 30-38

downloads: 428

cite

Full text: english pdf 1.578 Kb

page 30-38

downloads: 145

cite


Abstract

Objective: The aim of this study was to compare the light transmission of monowave and polywave-curing devices by a bulk-fill composite containing only camphorquinone as a photoinitiator. Materials and methods: Three light-curing devices were used to cure bulk-fill composite QuiXfil: one monowave (Translux® Wave) and two polywave (VALO Cordless and Bluephase® PowerCure). The NIST-calibrated spectrometer (MARC Resin Calibrator, BlueLight Analytics Inc.) was used to measure the incident and transmitted light through a 2-mm composite specimen over 20 s. Light transmittance was calculated from the ratio of the amount of transmitted and incident light. For data analysis (ANOVA, α = 0.05), total irradiation of the entire spectrum, irradiation with wavelengths of 360-420 nm for the violet spectrum, and 420-540 nm for the blue spectrum were selected. Results: Monowave curing unit Translux® Wave had the lowest light transmission (13.78 ± 0.5%), similar to the violet light transmission of polywave devices (12.02 ± 0.94% and 13.81 ± 1.72% for Valo Cordless and Bluephase PowerCure, respectively). Blue light transmittance (32.15-23.70%) was more than twofold higher than for the wavelengths in the violet region of the spectrum (13.81-12.02%) for the two polywave devices. VALO Cordless showed the highest total and blue light transmission (p<0.001). There was no significant difference in the transmission of the violet part of the spectrum between VALO Cordless and Bluephase® PowerCure (p = 0.465). Conclusion: Within the limitations of this study, we could conclude that polywave curing devices can be used for the polymerization of the bulk-fill composite with camphorquinone as the sole photoinitiator.

Keywords

Composite Resins; Dental Curing Lights; Dental Photoinitiators; Light-curing units; Monowave; Polywave; Light transmission; Photoinitiator

Hrčak ID:

315329

URI

https://hrcak.srce.hr/315329

Publication date:

25.3.2024.

Article data in other languages: croatian

Visits: 1.279 *