Determination of Dazomet in Basamid Granulat Using Reversed Phase HPLC

Biljana R. Petanovska-Ilievska ; Laboratory for Pesticide Formulation Control, Department of Chemistry, Faculty of Agriculture, Bul. Aleksandar Makedonski bb, 91000 Skopje, P.O. Box 297, Republic of Macedonia
Lila B. Vodeb ; Laboratory for Pesticide Formulation Control, Department of Chemistry, Faculty of Agriculture, Bul. Aleksandar Makedonski bb, 91000 Skopje, P.O. Box 297, Republic of Macedonia


Full text: english pdf 103 Kb

page 225-234

downloads: 838

cite


Abstract

The reverse phase HPLC determination of dazomet (tetrahydro-3,5-dimethyl-1,3,5-thiadiazine-2-tione) in a pesticide formulation has been studied. HS Pecosphere 3 x 3 C8 (3 μm, 3.3 x 0.46 cm) and LiChrosorb C8 (5 μm, 25 x 0.4 cm) analytical columns were tested at a flow rate of 1.0 cm3 min-1, column temperature of 25 °C and UV detection at 280 nm. The best separation of dazomet from its dehydro-dimer forms was achieved with a mobile phase containing acetonitrile-water in the volume ratio 15 : 85 on a HS Pecosphere column, and acetonitrile-water in the volume ratio 30 : 70 on a LiChrosorb C8 column. The HS Pecosphere column showed a better peak symmetry, separation factor, and multiple correlation coefficient. The retention time and peak area for the HS Pecosphere column were precise within a day and between days as indicated by the ANO VA test, in contrast to the LiChrosorb column, which showed lost of efficiency.

Keywords

reversed-phase HPLC; tetrahydro-3,5-dimethyl-1,3,5-thiadiazine-2-tione

Hrčak ID:

127498

URI

https://hrcak.srce.hr/127498

Publication date:

4.2.2002.