Skip to the main content

Professional paper

Determination of aluminium in serum by electrothermal atomic absorption spectrophotometry

P. Radošević ; Vojnomedicinska akademija, Beograd, Srbija


Full text: serbian pdf 2.135 Kb

page 55-64

downloads: 432

cite


Abstract

A simple and sensitive method for the determination of aluminium in human serum by electrothermal atomic absorption spectrophotometry has been developed. In view of contradictory data in literature all relevant analytical conditions have been tested. The method involves simple dilution with a 0.05% solution of Tri ton X-100 and atomization on the wall of a pyrolytically coated graphite tube. The sensitivity of the method is about 1 µg/L and precision < 11 %. The method of standard addition is recommended for calibration, but the use of a calibration curve based on scrum with a low content of aluminium is also possible.

Keywords

atomic absorption spectrophotometer; electrothermal atomizer; haemodialysis; serodiagnosis; aluminium toxicity; toxicology

Hrčak ID:

151061

URI

https://hrcak.srce.hr/151061

Publication date:

5.6.1991.

Article data in other languages: serbian

Visits: 974 *