Skip to the main content

Preliminary communication

Calibration of the Wiechert seismographs relative to a reference seismometer

Marijan Herak
Ivo Allegretti
Davorka Herak
Seweryn J. Duda


Full text: english pdf 1.897 Kb

page 31-59

downloads: 291

cite


Abstract

The horizontal 1000 kg Wiechert seismograph has been calibrated on the basis of a comparison of its records to the ones that were obtained by the reference seismometer. The resulting magnification curve shows notable deviations from the theoretical one. Its most prominent feature is a sudden magnification increase observed for frequencies above 5 Hz, which is attributed to frame oscillations excited by the seismic waves. Also, the observed average response curve exhibits larger magnification for frequencies below the resonance frequency. The results indicate a non-linear behavior of the Wiechert instrument related to the degree of impulsiveness of the excitation.

Keywords

Seismograph calibration; historical instrumentation; Wiechert seismograph

Hrčak ID:

18775

URI

https://hrcak.srce.hr/18775

Publication date:

1.12.1996.

Visits: 849 *