Skip to the main content

Original scientific paper

Interferometer as seismometer displacement transduce

Ivica Sović


Full text: croatian pdf 1.103 Kb

page 47-59

downloads: 358

cite


Abstract

Displacements of the pendulum of S-13 Teledyne Geotech seismometer were measured using a small Michelson interferometer. The oscillations of the pendulum were simultaneously recorded by the PDAS-100 digital seismograph. These records were integrated and compared with those made by the interferometer. The pendulum was simulated by sine (0.5, 1 and 2 Hz) and random signals. Michelson interferometer was simplified by using the cube beam splitter with mirror film deposited onto one of its faces. The source of radiation was thermally stabilised collimated laser diode (LD) driven with constant current. Wavelength of LD was 670 nm. Fringes were counted with 16-bit presentable up/down CMOS counter. Counter state was sampled by Atari ST 1040 computer with sampling frequency equal to 33.4 Hz. Resolution of the transducer was 335 nm. The transducer would be able to record oscillations of the pendulum caused by an earthquake with magnitudes between 1.1 and 5.2 at epicentral distance of 0.1°, and could therefore find the application in a displacement strong-motion seismograph.

Keywords

Interferometer; seismometer

Hrčak ID:

20130

URI

https://hrcak.srce.hr/20130

Publication date:

1.12.1993.

Article data in other languages: croatian

Visits: 1.501 *