Skip to the main content

Original scientific paper

https://doi.org/10.37023/ee.6.2.6

Solid-phase extraction of neonicotinoids residue from water: comparison between extraction cartridges

Gorica Vuković ; Institute of Public Health of Belgrade, Serbia
Dušan Marinković ; University of Novi Sad, Faculty of Agriculture, Novi Sad, Serbia
Aleksandra Petrović ; University of Novi Sad, Faculty of Agriculture, Novi Sad, Serbia
Bojan Kostadinović ; University of Novi Sad, Faculty of Agriculture, Novi Sad, Serbia
Tijana Stojanović ; University of Novi Sad, Faculty of Agriculture, Novi Sad, Serbia
Nikola Puvača ; University Business Academy, Faculty of Economics and Engineering Management, Novi Sad, Serbia
Vojislava Bursić ; University of Novi Sad, Faculty of Agriculture, Novi Sad, Serbia


Full text: english pdf 393 Kb

page 84-89

downloads: 404

cite


Abstract

The Commission Implementing Decision (EU) 2018/840 established the watch list of substances for Union-wide monitoring in the field of water policy pursuant to Directive 2008/105/EC and Commission Implementing Decision (EU) 2015/495. This Decision establishes additional substances and their maximum acceptable method detection limits including neonicotinoids (9-500 ng/L). During the pesticides extraction from water, the solid-phase extraction (SPE) column was used. Bakerbond spe™ SDB-1 SPE and Bond Elut Plexa were evaluated for the SPE of neonicotinoids (imidacloprid, thiamethoxam, acetamiprid, thiacloprid and clothianidin) from water. The comparation of the results of these two types of columns, was performed by spiking water samples at two levels (0.05 and 0.1 μg/L) in three replicates. The Bakerbond™ column quantitatively adsorbed these pesticides, with the obtained recoveries: 36.9 % for imidacloprid, 43.2 % for thiacloprid, 119.3 % for clothianidin, 64.6 % for acetamiprid and 53.6 % for thiamethoxam. Using Bond Elut Plexa the obtained recoveries were 66.9 % for imidacloprid, 72.9% for thiacloprid, 103.4 % for clothianidin, 67.4 % for acetamiprid and 45.2% for thiamethoxam. Both colums have low recovery values for thiametoxam, while the highest values were obtained by Bond Elut Plexa for thiacloprid, imidacloprid and acetamiprid. Great recoveries were achieved for clothianidin using both SPE columns.

Keywords

neonicotinoids; SPE extraction; water; LC-MS/MS

Hrčak ID:

229236

URI

https://hrcak.srce.hr/229236

Publication date:

9.12.2019.

Visits: 929 *