A quasi adiabatic reaction and solution calorimeter was constructed
and tested using tris(hydroxymethyl)aminomethane dissolution
in 0.1 mol dm-3 HCl as the reference (standard) process.
The precision of the calorimeter amounts to approximately ± 1 per
cent and its sensitivity to 4 J dm-a.