The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures

Nick Polychronakis ; Assistant Professor, Division of Removable Prosthodontics, Dental School, National and Kapodistrian University of Athens, Greece
Stavros Yannikakis ; Department of Dental Technology, Division of Removable Prosthodontics, Technological Educational Institution of Athens, Greece
Alcibiades Zissis ; Associate Professor, Division of Removable Prosthodontics, Dental School, National and Kapodistrian University of Athens, Greece

Puni tekst: engleski, pdf (286 KB) str. 279-284 preuzimanja: 347* citiraj
APA 6th Edition
Polychronakis, N., Yannikakis, S. i Zissis, A. (2014). The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures. Acta stomatologica Croatica, 48 (4), 279-284. https://doi.org/10.15644/asc48/4/5
MLA 8th Edition
Polychronakis, Nick, et al. "The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures." Acta stomatologica Croatica, vol. 48, br. 4, 2014, str. 279-284. https://doi.org/10.15644/asc48/4/5. Citirano 24.02.2021.
Chicago 17th Edition
Polychronakis, Nick, Stavros Yannikakis i Alcibiades Zissis. "The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures." Acta stomatologica Croatica 48, br. 4 (2014): 279-284. https://doi.org/10.15644/asc48/4/5
Harvard
Polychronakis, N., Yannikakis, S., i Zissis, A. (2014). 'The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures', Acta stomatologica Croatica, 48(4), str. 279-284. https://doi.org/10.15644/asc48/4/5
Vancouver
Polychronakis N, Yannikakis S, Zissis A. The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures. Acta stomatologica Croatica [Internet]. 2014 [pristupljeno 24.02.2021.];48(4):279-284. https://doi.org/10.15644/asc48/4/5
IEEE
N. Polychronakis, S. Yannikakis i A. Zissis, "The Effect of Repeated Microwaving Disinfection on the Dimensional Stability of Acrylic Dentures", Acta stomatologica Croatica, vol.48, br. 4, str. 279-284, 2014. [Online]. https://doi.org/10.15644/asc48/4/5
Puni tekst: hrvatski, pdf (286 KB) str. 279-284 preuzimanja: 136* citiraj
APA 6th Edition
Polychronakis, N., Yannikakis, S. i Zissis, A. (2014). Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza. Acta stomatologica Croatica, 48 (4), 279-284. https://doi.org/10.15644/asc48/4/5
MLA 8th Edition
Polychronakis, Nick, et al. "Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza." Acta stomatologica Croatica, vol. 48, br. 4, 2014, str. 279-284. https://doi.org/10.15644/asc48/4/5. Citirano 24.02.2021.
Chicago 17th Edition
Polychronakis, Nick, Stavros Yannikakis i Alcibiades Zissis. "Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza." Acta stomatologica Croatica 48, br. 4 (2014): 279-284. https://doi.org/10.15644/asc48/4/5
Harvard
Polychronakis, N., Yannikakis, S., i Zissis, A. (2014). 'Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza', Acta stomatologica Croatica, 48(4), str. 279-284. https://doi.org/10.15644/asc48/4/5
Vancouver
Polychronakis N, Yannikakis S, Zissis A. Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza. Acta stomatologica Croatica [Internet]. 2014 [pristupljeno 24.02.2021.];48(4):279-284. https://doi.org/10.15644/asc48/4/5
IEEE
N. Polychronakis, S. Yannikakis i A. Zissis, "Učinak ponovljenih mikrovalnih dezinfekcija na dimenzijsku stabilnost akrilatnih proteza", Acta stomatologica Croatica, vol.48, br. 4, str. 279-284, 2014. [Online]. https://doi.org/10.15644/asc48/4/5

Rad u XML formatu

Sažetak
Objective. The aim of this study was to evaluate the effect of repeated microwave disinfections on the dimensional stability of acrylic dentures. Materials and Methods. Three groups of dentures made of a heat polymerized acrylic resin were tested. I: dentures kept in water (control group). II: dentures microwaved daily while being immersed into water (wet disinfection). III: dentures microwaved daily without being immersed into water (dry disinfection). Measurements were taken across three reference points, on two occasions: after curing and immersion in water for 24 hours, and one week later. Data obtained were analyzed using one-way analysis of variance (ANOVA)
and Scheffe’s multiple range test. Results.The results showed that the microwave disinfection provokes dimensional changes of the same pattern (shrinkage). The dentures which underwent wet disinfection exhibited the greatest shrinkage (p<0.05). Conclusions. Disinfection using microwave energy may cause dimensional changes (shrinkage) of complete dentures. The microwave “dry disinfection” method can be safely applied in everyday practice since the dimensional changes
which occurred seem to be of no clinical significance.

Ključne riječi
Complete Dentures, Acrylic Resins; Disinfection; Microwaves; Dimensional Measurement Accuracy

Hrčak ID: 131867

URI
https://hrcak.srce.hr/131867

▼ Article Information



Introduction

Dentures in all stages of their construction are potential sources of cross contamination to and from patients, clinical personnel and dental technicians. Thus, dental clinic personnel and technicians should protect both themselves and patients applying relatively simple sterilization-disinfection procedures in everyday practice (1-6).

Many disinfection-sterilization methods have been suggested such as: immersion in 2% alkaline glutaraldehyde for 10 minutes, immersion in 1% sodium hypochlorite for 10 to 30 minutes, and immersion in 3% aqueous formaldehyde for 30 minutes (3). Also, immersion in a solution of 5.25% sodium hypochlorite for 5 minutes can sterilize a denture (7). Nevertheless, these chemical disinfection methods hold some disadvantages e.g. time-consuming, denture bleaching and staining, expiration date and possible oral tissue reaction (8-10). Furthermore, the chemical solutions may alter some physical-mechanical properties of denture base materials (11-13). Microwave energy is claimed to be an alternative to these traditional sterilization-disinfection methods and has been reported to be a reliable, cost effective and time saving method for complete denture disinfection (8, 14-22).

When using the microwave disinfection method, a denture is placed either into water (wet disinfection) or not (dry disinfection) and the microwave oven irradiation is mainly between 450 and 650 W. Furthermore, the time periods of irradiation range from 2 to 10 min.

A study conducted by Rohrer et al. (8) has shown that microwave irradiation at 720 W for 8 minutes, in dry conditions, can sterilize acrylic dentures contaminated with aerobic bacteria and Candida albicans. Webb et al. (14) reported that microwave irradiation at 604 W (±92) or 331 W (±34) for 6 minutes produced sterilization of dry complete upper dentures contaminated by Candida albicans and S. gordonii.

Silva et al. (16) reported that complete dentures contaminated with Candida albicans and S. aureus were sterilized after microwave irradiation at 650 W for 6 minutes. Mima et al. (17) revealed that microwave energy at 650 W for 3 minutes can be used for hard chairside reline resin sterilization. Recently, Senna et al. (22) showed that microwave irradiation at 450 W for 3 minutes or 450 W for 2 or 3 minutes combined with alkaline peroxide containing enzyme denture cleanser is an effective method for disinfection of dentures contaminated with Candida albicans. In all the above mentioned studies (16, 17, 22) for the disinfection/sterilization procedure at 650 or 450W all dentures were placed into water. A review of the literature revealed that although many studies have been conducted to evaluate the effect of microwave disinfection on the physical-mechanical properties of denture base resins (23-31), little information is available on the effect of repeated microwave disinfections on the dimensional changes of denture base materials (30-32), particularly in dry conditions (13).

The aim of this study was to evaluate the effect of repeated microwave disinfections on the dimensional stability of acrylic denture bases.

Materials and Methods

A permanent silicone elastomer mold was made (Panadent Ltd, London, England) from a standard brass die simulating an upper edentulous ridge where index marks had been incorporated by drilling holes in specific locations (33) (in the right and left posterior area (A, B) respectively and the central incisor region (C)). Eighteen casts were obtained by pouring dental stone (Vel-Mix Stone, Kerr Mfg Co, MIIn). Each stone cast was waxed up with a 3 mm thick wax sheet (Tenatex red Associated Dental Products Ltd, Purton, Swindon, Wiltshire SNS 4HT, UK). Eighteen denture bases were constructed using a heat polymerized resin (Paladon 65, Kulzer, GmbH, Germany) following a curing cycle at 74oC for 7 hours and 1 hour at 100oC in water bath (Figure 1).

Figure 1 Denture base with index marks.
ASC_48(4)_279-294-f1

The denture bases were randomly divided into 3 groups according to the disinfection procedure:

  • Group I: the denture bases were kept in distilled water (37°C) during the experiment period (control group).

  • Group II: the denture bases were individually microwaved for 6 minutes at 650 W daily for one week. During disinfection the dentures were immersed in 150 ml of distilled water (wet disinfection).

  • Group III: the dentures were individually microwaved following the same procedure as group II without immersing them in water (dry disinfection).

Denture bases of group II and III were microwaved every day at the same time using a domestic microwave oven (Model HF 1210, Siemens GmbH, Germany). A plastic cup, filled with 150 ml of water, was placed into the microwave oven during dry disinfection (group III), to protect the device from damage. All denture bases were kept in distilled water at 37°C between the disinfection procedures.

Method of measurement

Measurements were taken across the dimensions AB, AC, and BC, with a digital micrometer accurate to 0.01 mm (Mitutoyo, Tokyo, Japan). For each dimension, five readings were made and the mean value was calculated. The coefficient of variation for the repeated measurements did not exceed 0.9%. In order to take the measurements, the tip of the micrometer was placed on the centre of the index marks (Figure 1).

The measurements were taken on two occasions: after curing and immersion in water for 24 hours (considered to be the starting point), and one week later.

STATISTICAL ANALYSIS

Data obtained were analyzed using one-way analysis of variance (ANOVA) and Scheffe’s multiple range test.

Results

The results of this study, in bar graphs, are presented in Figures 2, 3, and 4.

Figure 2 Dimensional changes for the distance AB.
ASC_48(4)_279-294-f2
Figure 3 Dimensional changes for the distance AC.
ASC_48(4)_279-294-f3
Figure 4 Dimensional changes for the distance BC.
ASC_48(4)_279-294-f4

Generally, all microwaved dentures bases showed a similar pattern of dimensional changes, a shrinkage for all distances measured (Figures 2, 3, 4).

After seven dry disinfections, the denture bases showed smaller dimensional changes compared to those submitted to wet disinfections for all measured distances (Figures 2, 3, 4). This finding was statistically significant only for the distance AB (p<0.05) (Figure 2).

Discussion

The main finding of this in vitro study was that all denture bases submitted to microwaving disinfection, either wet or dry, exhibited shrinkage after seven disinfections. This may be due to the fact that when acrylic denture bases are heated, as occurs during microwaving, internal stresses incorporated during the processing are released (25).

Our results revealed that denture bases which had undergone wet disinfections showed greater linear changes, compared to those occurring during dry disinfections at the posterior region of the denture bases. This can be explained by the fact that the boiling water (water starts to boil after 90 sec of irradiation) raises the temperature of the denture base, which in turn causes the residual monomer (34) conversion into polymer (25).

Denture bases which had undergone wet microwave disinfection presented linear changes up to -1.16%. This finding is in agreement with the study by Sartori et al. (24) claiming that repeated wet microwave irradiation negatively affects the internal adaptation of denture bases to their casts. Fleck et al. (26) showed that repeated wet disinfection had harmful effects on denture bases’ adaptation to the casts. Furthermore, our results are comparable with those of Seo et al. (32) who reported shrinkage of 2.36% after daily disinfections for seven days. Wagner et al. (31) revealed a decrease of 1.92% at the distance between buccal flanges after two disinfections at 700 W for 3 minutes. On the other hand, Consani et al. (35) reported that repeated wet microwave disinfections did not adversely affect the denture base adaptation.

Our finding that dry disinfection caused minor shrinkage is in agreement with the report of Polyzois et al. (13) stating that acrylic rectangular specimens which underwent dry disinfection showed negligible dimensional changes (-0.005 to 0.009%). Also, it confirms the report of Burns et al. (36) who revealed a minor shrinkage up to 0.03% in cylindrical specimens after a 15-minute at 650 W microwave disinfection. Furthermore, the finding corroborates the study of Pavan et al. (10) that microwave irradiation at 500 W for 3 min did not affect the adaptation of maxillary acrylic denture bases. However, microwaving of denture bases at 604 W for 10 min adversely affects their fit to the stone casts.

In everyday practice, a denture base material is considered to be acceptable when the linear changes across first molars occurring during processing and use are under the limit of 1%, as it was suggested by Woelfer et al. (37) about 50 years ago.

Concerning the above mentioned reports and our findings that dry disinfection caused shrinkage less than 1%, we may conclude that when there is a need for dentures to be repeatedly disinfected, a dry microwaving disinfection is preferable since it does not affect the dimensional stability of denture bases to a degree which would cause clinical problems.

ConclusionS

Under the limitations of this laboratory study, the following conclusions can be drawn:

  1. Disinfection using microwave energy may cause dimensional changes (shrinkage) of complete dentures.

  2. Denture bases immersed in water during microwave irradiation exhibited greater dimensional changes compared to those which were not immersed in water.

  3. Denture bases submitted to repeating disinfection, without being immersed in water, showed dimensional stability comparable to the untreated denture bases.

  4. Microwaving disinfection is preferable to be carried out in dry environment because the dimensional changes occurred seem to be of no clinical significance.

Notes

[1] Conflicts of interest Conflict of interest: None declared

References

1 

No authors listed. Infection control in dental office. J Am Dent Assoc. 1978 Oct;97(4):673–7. DOI: http://dx.doi.org/10.14219/jada.archive.1978.0370 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/281409

2 

Kahn RC, Lancaster MV, Kate WJ. The microbiologic cross-contamination of dental prostheses. J Prosthet Dent. 1982 May;47(5):556–9. DOI: http://dx.doi.org/10.1016/0022-3913(82)90309-2 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/6953241

3 

No authors listed. Guidelines for infection control in the dental office and the commercial dental laboratory. Council on Dental Therapeutics. Council on Prosthetic Services and Dental Laboratory Relations. J Am Dent Assoc. 1985 Jun;110(6):969–72. DOI: http://dx.doi.org/10.14219/jada.archive.1985.0016 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/3860557

4 

Powell GL, Runnells RD, Saxon BA, Whisenant BK. The presence and identification of organisms transmitted to dental laboratories. J Prosthet Dent. 1990 Aug;64(2):235–7. DOI: http://dx.doi.org/10.1016/0022-3913(90)90185-F PubMed: http://www.ncbi.nlm.nih.gov/pubmed/2202822

5 

King AH, Matis B. Infection control of in-office dental laboratories. Dent Clin North Am. 1991 Apr;35(2):415–26. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/2032584

6 

Chau VB, Saunders TR, Pimsler M, Elfring DR. In-depth disinfection of acrylic resins. J Prosthet Dent. 1995 Sep;74(3):309–13. DOI: http://dx.doi.org/10.1016/S0022-3913(05)80140-4 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/7473287

7 

Rudd RW, Senia ES, McCleskey FK, Adams ED. Sterilization of complete dentures with sodium hypochloride. J Prosthet Dent. 1984 Mar;51(3):318–21. DOI: http://dx.doi.org/10.1016/0022-3913(84)90212-9 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/6584598

8 

Rohrer MD, Bulard RA. Microwave sterilization. J Am Dent Assoc. 1985 Feb;110(2):194–8. DOI: http://dx.doi.org/10.14219/jada.archive.1985.0250 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/3884686

9 

Baysan A, Whiley R, Wright PS. Use of microwave energy to disinfect a long-term soft lining material contaminated with candida albicans or staphylococcus aureus. J Prosthet Dent. 1998 Apr;79(4):454–8. DOI: http://dx.doi.org/10.1016/S0022-3913(98)70161-1 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/9576322

10 

Pavan S, Arioli Filio JN, Dos Santos PH, Mollo Fde A Jr. Effect of microwave treatments on dimensional accuracy of maxillary acrylic resin denture base. Braz Dent J. 2005;16(2):119–23. DOI: http://dx.doi.org/10.1590/S0103-64402005000200006 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/16475605

11 

Robinson JG, McCabe JF, Storer R. Denture bases, the effects of various treatments on clarity, strength and structure. J Dent. 1987 Aug;15(4):159–65. DOI: http://dx.doi.org/10.1016/0300-5712(87)90143-6 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/3478378

12 

Asad T, Watkinson AC, Huggett R. The effect of disinfection procedures on flexural properties of denture base acrylic resins. J Prosthet Dent. 1992 Jul;68(1):191–5. DOI: http://dx.doi.org/10.1016/0022-3913(92)90303-R PubMed: http://www.ncbi.nlm.nih.gov/pubmed/1403907

13 

Polyzois GL, Zissis AJ, Yannikakis SA. The effect of glutaraldehyde and microwave disinfection on some properties of acrylic denture resin. Int J Prosthodont. 1995 Mar-Apr;8(2):150–4. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/7575966

14 

Webb BC, Thomas CJ, Harty DW, Willcox MD. Effectiveness of two methods of denture sterilization. J Oral Rehabil. 1998 Jun;25(6):416–23. DOI: http://dx.doi.org/10.1046/j.1365-2842.1998.00266.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/9687113

15 

Dixon DL, Breeding LC, Faler TA. Microwave disinfection of denture base materials colonized with Candida albicans. J Prosthet Dent. 1999 Feb;81(2):207–14. DOI: http://dx.doi.org/10.1016/S0022-3913(99)70250-7 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/9922435

16 

Silva MM, Vergani CE, Giampaolo ET, Neppelenbroek KH, Spolidorio DM, Machado AL. Effectiveness of microwave irradiation on the disinfection of complete dentures. Int J Prosthodont. 2006 May-Jun;19(3):288–93. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/16752628

17 

Mima EGO, Pavarina AC, Neppelenbroek KH, Vergani CE, Spolidorio DMP, Machado AL. Effect of different exposure times on microwave irradiation on the disinfection of a hard chairside relines resin. J Prosthodont. 2008 Jun;17(4):312–7. DOI: http://dx.doi.org/10.1111/j.1532-849X.2007.00277.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/18086140

18 

Neppelenbroek KH, Pavarina AC, Spolidorio DMP, Massucato EMS, Spolidorio LC, Vergani CE. Effectiveness of microwave disinfection of complete dentures on the treatment of candida – related stomatitis. J Oral Rehabil. 2008 Nov;35(11):836–46. DOI: http://dx.doi.org/10.1111/j.1365-2842.2008.01874.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/18466276

19 

Buergers R, Rosentritt M, Scheider-Brachert W, Behr M, Handel G, Hahnel S. Efficacy of denture disinfection methods in controlling Candida albicans colonization in vitro. Acta Odontol Scand. 2008 Jun;66(3):174–80. DOI: http://dx.doi.org/10.1080/00016350802165614 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/18568477

20 

Ribeiro DG, Pavarina AC, Dovigo LN, Palomari Spolidorio DM, Giampaolo ET, Vergani CE. Denture disinfection by microwave irradiation: A randomized clinical study. J Dent. 2009 Sep;37(9):666–72. DOI: http://dx.doi.org/10.1016/j.jdent.2009.04.009 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19481327

21 

Dovigo LN, Pavarina AC, Ribeiro DG, de Oliveira JA, Vergani CE, Machado AL. Microwave disinfection of complete dentures contaminated in vitro with selected bacteria. J Prosthodont. 2009 Oct;18(7):611–7. DOI: http://dx.doi.org/10.1111/j.1532-849X.2009.00489.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19523027

22 

Senna PM, Sotto-Mattior BS, Silva WJ, Del Bel Curry AA. Adding denture cleanser to microwave disinfection regimen to reduce the irradiation time and the exposure of dentures to high temperatures. Gerodontology. 2013 Mar;30(1):26–31. DOI: http://dx.doi.org/10.1111/j.1741-2358.2012.00641.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22420329

23 

Pavarina AC, Neppelenbroek KH, Guinesi AS, Vergani CE, Machado AC, Ciampaolo ET. Effect of microwave disinfection on the flexural strength of hard chairside reline resins. J Dent. 2005 Oct;33(9):741–8. DOI: http://dx.doi.org/10.1016/j.jdent.2005.02.003 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/16199282

24 

Sartori EA. Schmidt, Walber LE, Shinkai RSA. Effect of microwave disinfection on denture base adaptation and resin surface roughness. Braz Dent J. 2006;17(3):195–200. DOI: http://dx.doi.org/10.1590/S0103-64402006000300004 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17262124

25 

Goncalves AR, Machado AL, Giampaolo ET, Pavarina AC, Vergani CE. Linear dimensional changes of denture base and hard chair-side reline resins after disinfection. J Appl Polym Sci. 2006;102:1821–6. DOI: http://dx.doi.org/10.1002/app.24052

26 

Fleck G, Ferneda F, Ferreira da Silva DF, Mota EG, Shinkai RS. Effect of two microwave disinfection protocols on adaptation of poly (methyl methacrylate) denture bases. Minerva Stomatol. 2007 Mar;56(3):121–7. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17327816

27 

Sartori EA, Schmidt CB, Mota EG, Hirakata LM, Shinkai RSA. Cumulative effect of disinfection procedures on microhardness and tridimensional stability of a poly (methyl methacrylate) denture base resin. J Biomed Mater Res B Appl Biomater. 2008 Aug;86(2):360–4. DOI: http://dx.doi.org/10.1002/jbm.b.31027 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/18161805

28 

Machado AL, Breeding LC, Vergani CE, Perez LE. Hardness and surface roughness of reline and denture base acrylic resins after repeated disinfection procedures. J Prosthet Dent. 2009 Aug;102(2):115–22. DOI: http://dx.doi.org/10.1016/S0022-3913(09)60120-7 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19643225

29 

Consani RL, Azevedo DD, Mesquita MF, Mendes WB, Saquy PC. Effect of repeated disinfections by microwave energy on physical and mechanical properties of denture base acrylic resins. Braz Dent J. 2009;20(2):132–7. DOI: http://dx.doi.org/10.1590/S0103-64402009000200008 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19738946

30 

Basso MFM, Giampaolo ET, Vergani CE, Machado AL, Pavarina AC, Ribeiro RC. Influence of microwave disinfection on the dimensional stability of denture reline polymers. J Prosthodont. 2010 Jul;19(5):364–8. DOI: http://dx.doi.org/10.1111/j.1532-849X.2010.00583.x PubMed: http://www.ncbi.nlm.nih.gov/pubmed/20345741

31 

Wagner DA, Pipko DJ. The effect of repeated microwave irradiation on the dimentional stability of a specific acrylic denture resin. J Prosthodont. 2015 Jul 31 DOI: http://dx.doi.org/10.1111/jopr.12203 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/25082075

32 

Seo RS, Vergani CE, Pavarina AC, Compagnoni MA, Machado AL. Influence of microwave disinfection on the dimensional stability of intact and relined acrylic-resin denture bases. J Prosthet Dent. 2007 Sep;98(3):216–23. DOI: http://dx.doi.org/10.1016/S0022-3913(07)60058-4 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17854623

33 

Huggett R, Zissis A, Harrison A, Dennis A. Dimensional accuracy and stability of acrylic resin denture bases. J Prosthet Dent. 1992 Oct;68(4):634–40. DOI: http://dx.doi.org/10.1016/0022-3913(92)90379-O PubMed: http://www.ncbi.nlm.nih.gov/pubmed/1403942

34 

Harrison A, Huggett R. Effect of curing cycle on residual monomer levels of acrylic resin denture base polymers. J Dent. 1992 Dec;20(6):370–4. DOI: http://dx.doi.org/10.1016/0300-5712(92)90031-7 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/1452879

35 

Consani RL, Iwasaki RY, Mesquita MF, Mendes WB, Consani S. Effect of repeated simulated disinfections by microwave energy on complete denture base adaptation. Open Dent J. 2008;2:61–6. DOI: http://dx.doi.org/10.2174/1874210600802010061 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19088884

36 

Burns DR, Kazanoglu A, Moon PC, Gunsolley JC. Dimensional stability of acrylic resin materials after microwave sterilization. Int J Prosthodont. 1990 Sep-Oct;3(5):489–93. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/2088388

37 

Woelfel JB, Paffenbarger GC, Sweeney WT. Method of evaluating the clinical effect of warping a denture: Report of a case. J Am Dent Assoc. 1959 Aug;59(2):250–60. DOI: http://dx.doi.org/10.14219/jada.archive.1959.0171 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/13672726


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.

[hrvatski]

Posjeta: 756 *