Kratko priopćenje

A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids

B. B. Radak ; Radiation Chemistry Department, Boris Kidric Institute of Nuclear Sciences, Beograd-Vinča
V. I. Matić ; Radiation Chemistry Department, Boris Kidric Institute of Nuclear Sciences, Beograd-Vinča
V. M. Radak ; Institute of Chemistry and Physical Chemistry, Faculty of Science, Beograd, Serbia, Yugoslavia

Puni tekst: engleski, pdf (2 MB) str. 25-30 preuzimanja: 69* citiraj
APA 6th Edition
Radak, B.B., Matić, V.I. i Radak, V.M. (1977). A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids. Croatica Chemica Acta, 49 (1), 25-30. Preuzeto s https://hrcak.srce.hr/196408
MLA 8th Edition
Radak, B. B., et al. "A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids." Croatica Chemica Acta, vol. 49, br. 1, 1977, str. 25-30. https://hrcak.srce.hr/196408. Citirano 08.03.2021.
Chicago 17th Edition
Radak, B. B., V. I. Matić i V. M. Radak. "A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids." Croatica Chemica Acta 49, br. 1 (1977): 25-30. https://hrcak.srce.hr/196408
Harvard
Radak, B.B., Matić, V.I., i Radak, V.M. (1977). 'A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids', Croatica Chemica Acta, 49(1), str. 25-30. Preuzeto s: https://hrcak.srce.hr/196408 (Datum pristupa: 08.03.2021.)
Vancouver
Radak BB, Matić VI, Radak VM. A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids. Croatica Chemica Acta [Internet]. 1977 [pristupljeno 08.03.2021.];49(1):25-30. Dostupno na: https://hrcak.srce.hr/196408
IEEE
B.B. Radak, V.I. Matić i V.M. Radak, "A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids", Croatica Chemica Acta, vol.49, br. 1, str. 25-30, 1977. [Online]. Dostupno na: https://hrcak.srce.hr/196408. [Citirano: 08.03.2021.]

Sažetak
A quasi adiabatic reaction and solution calorimeter was constructed
and tested using tris(hydroxymethyl)aminomethane dissolution
in 0.1 mol dm-3 HCl as the reference (standard) process.
The precision of the calorimeter amounts to approximately ± 1 per
cent and its sensitivity to 4 J dm-a.

Hrčak ID: 196408