Skoči na glavni sadržaj

Izvorni znanstveni članak

Ac-Dc Characterization of Coaxial Current Shunts and Application of the hunt in the Digital Sampling Wattmeter

Jure Konjevod orcid id orcid.org/0000-0001-6233-9286 ; University of Zagreb Faculty of Electrical Engineering and Computing / FER Zagreb, Croatia
Roman Malarić ; University of Zagreb Faculty of Electrical Engineering and Computing / FER Zagreb, Croatia
Martin Dadić ; University of Zagreb Faculty of Electrical Engineering and Computing / FER Zagreb, Croatia
Ivica Kunšt ; University of Zagreb Faculty of Electrical Engineering and Computing / FER Zagreb, Croatia
Hrvoje Hegeduš ; University of Zagreb Faculty of Electrical Engineering and Computing / FER Zagreb, Croatia


Puni tekst: engleski pdf 1.179 Kb

preuzimanja: 293

citiraj


Sažetak

The purpose of this paper is to give a review of ac-dc characterization of the current shunts and application of the current shunt (nominal current 1A)
within the digital sampling wattmeter. It is described the ac-dc transfer difference measurement of six cage type ac shunts from 10 mA up to 10 A
using step-up measurement procedure. Furthermore, the substantial part of the measurement setup is fast switching system which is also described
in detail. For the purpose of measurement procedure, the application is developed in LabVIEW and whole process is fully automatized. Obtained results
are analyzed and shown on graphs. This paper is extended version of two papers: [1] and [2] which are presented on 1st International Colloquium
on Smart Grid Metrology. Thus, paper is extended with presented application of the shunt 1 A in the digital sampling wattmeter

Ključne riječi

current shunt; step-up procedure; thermal voltage converter; automated measurement procedure; digital sampling wattmeter

Hrčak ID:

213473

URI

https://hrcak.srce.hr/213473

Datum izdavanja:

1.5.2018.

Posjeta: 633 *