Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.11613/BM.2021.030701

Comparison of Freelite and N-Latex serum free light chain assays: a critical review

Massimo Daves ; Clinical Biochemical Laboratory
Andrea Piccin ; Northern Ireland Blood Transfusion Service (NIBTS)
Vincenzo Roccaforte ; Clinical Pathology Laboratory
Giuseppe Lippi ; Section of Clinical Biochemistry


Puni tekst: engleski pdf 216 Kb

preuzimanja: 304

citiraj

Preuzmi JATS datoteku


Sažetak

The measurement of serum free light chain (FLC) represents a fundamental aspect on the assessment of patients with monoclonal gammopathies (MG). Different analytical methods for FLC have become available with the possibility to obtain different value with a substantial impact on the assessment of patients with MG. This study aimed to evaluate FLC results obtained with two different assays and how the difference value obtained can impact in the patient’s assessment.
Ninety-three patient serum samples that underwent analysis for FLC with two different methods, Serum Freelite (The Binding Site, Birmingham, UK) and N-Latex FLC (Siemens, Marburg, Germany), were included in this retrospective study. Statistical analysis was performed to evaluate correlation, difference, and the grade of concordance between the results obtained with the two methods.
Significant statistical differences between the results obtained from the two methods were found (P < 0.05). A good correlation was found (0.99 for κ FLC, 0.95 for λ FLC, and 0.94 for the κ/λ ratio, respectively). We found a weighted kappa value of 0.65 for κ/λ ratio, 0.65 for λ FLC and 0.90 for κ FLC. A positive bias found with the Bland-Altman plot mirrors overestimation of κ FLC and κ/λ ratio with Freelite compared to N-Latex, whilst a negative bias underscores underestimation of λ FLC by Freelite compared to N-Latex.
Although in general the concordance between Freelite and N-Latex appears satisfactory, several discrepancies could be evidenced and consequently the two assays are not interchangeable.

Ključne riječi

free light chain; monoclonal gammopathies; analytical techniques and equipment; comparison; immunoassays

Hrčak ID:

264025

URI

https://hrcak.srce.hr/264025

Datum izdavanja:

15.10.2021.

Posjeta: 1.116 *




Introduction

Monoclonal gammopathies (MG) are characterized by clonal proliferation of plasma cells and the production of monoclonal entire immunoglobulin and/or fragments. Detection and measurement of monoclonal proteins or monoclonal components (MC) are an integral part of diagnosis and clinical follow-up and have been traditionally performed by serum and urine electrophoresis and immunofixation (IFE) (1). Human antibodies (Ab) are composed of two identical heavy chains and two identical light chains that are produced in excess of heavy chains, and excess light chains secreted in the blood are called free light chains (2). Analytical methods for routine free light chain (FLC) measurement in serum have become commercially available at the beginning of this century so the widespread availability of automated serum assay for measuring and quantifying FLC has generated a significant impact on laboratory assessment of patients with MG (3). In 2009 the guideline endorsed by the International Myeloma Working Group (IMWG) recommended the assessment of FLC along with serum electrophoresis and IFE, since these tests were considered the best screening panel for studying plasma cells disorder (4).

Now, this testing strategy represents an essential step in the evaluation of MC patients, being capable to detect and measuring light chain MC in almost all patients with non-secreting or oligo-secreting disease and amyloidosis (AL). Furthermore, the measurement of FLC and the derived index, i.e., the ratio between the κ FLC and λ FLC (κ/λ ratio) and involved FLC/non-involved FLC (iFLC/niFLC) ratio, are recommended for the risk stratification of disease progression and in the evaluation of the response to therapy in multiple myeloma or AL (4). It has also been suggested that the measurement of FLC may replace in some cases that of Bence Jones protein (BJP). However, in AL as well as in other clinical conditions characterized by elusive MC, urine-IFE must be performed for detecting BJP, such that maximum diagnostic sensitivity can be reached (5,6). Therefore, this study aimed to evaluate FLC results obtained with a new nephelometer recently commercially available and how the implementation of a different assay for FLC measurement can impact patient assessment.

Materials and methods

Study design

This retrospective study was conducted between 1st October 2020 and 30th November 2020 at the Clinical Biochemical Laboratory, Hospital of Bolzano, Italy. The study was planned after implementing the new nephelometer Atellica NEPH 630 (Siemens, Erlangen, Germany) to replace a turbidimeter Optilite (The Binding Site, Birmingham, UK) previously used for the same purposes. All FLC test results performed during the study period were extracted from the laboratory information system (LIS). Serum Freelite (The Binding Site, Birmingham, UK) analyses were measured using the turbidimeter whilst N-Latex FLC (Siemens, Erlangen, Germany) analyses were performed with the new nephelometer. All tests were performed according to the manufacturer’s instructions. This investigation was based on pre-existing data extracted from the LIS in fully anonymised form, so that informed consent was unnecessary. The study was performed in accordance with the Declaration of Helsinki and under the terms of all relevant local legislations.

Materials and methods

A total number of 93 patients (median age 72 years, range 20-88 years; male 53/93, 0.57) serum samples undergoing routine clinical analysis for FLC at the laboratory were finally included in the study. The concentration of FLC and κ/λ ratio were considered abnormal when results were outside of the reference interval provided by the manufacturers (Freelite κ FLC: 3.3-19.4 mg/L; λ FLC: 5.7-26.3 mg/L; κ/λ ratio: 0.26-1.65; N-Latex FLC: κ FLC: 6.7-22.4 mg/L; λ FLC: 8.3-27.0; κ/λ ratio: 0.31-1.56).

Statistical analysis

Shapiro-Wilk test for normal distribution showed that all the investigated parameters were non-normally distributed so that results were reported as median, 95% confidence interval (95% CI), and interquartile range (IQR). Wilcoxon’s test was used to assess the statistical significance of differences between the two methods. A P < 0.05 was set as statistically significant. Passing-Bablok regression, Bland-Altman plot, and Spearman’s rank correlation were carried out to analyse the numerical results of κ FLC, λ FLC and FLC-ratio obtained with the two FLC assays. Qualitative concordance of the two FLC assays was explored with the weighted kappa (κ) coefficient, where complete agreement was defined as κ coefficient = 1.00, high agreement as 0.81 ≤ κ coefficient < 1 and a good agreement when 0.61 ≤ κ coefficient < 0.8. The number of patients with iFLC/niFLC ratio ≥ 100 with the two assays was also calculated. Concordance with initial diagnoses was analysed when the κ/λ ratio obtained with the two methods was not concordant. Statistical analysis was done using MedCalc 17.4.4 statistical software (MedCalc Software, Ostend, Belgium).

Results

The main results of this investigation are shown inTable 1 andFigures 1, 2 and 3{ label needed for fig[@id='f2'] }{ label needed for fig[@id='f3'] }. The result of the Passing-Bablok regression analysis for κ FLC is as follows: N-Latex κ FLC = 4.08 (2.97 to 5.01) + 0.78 (0.74 to 0.83) Freelite κ FLC; the Cusum test for linearity showed no significant deviation from linearity (P = 0.130). The result of the Passing and Bablok regression analysis for the λ FLC is as follows: N-Latex λ FLC = 5.42 (3.00 to 7.51) + 1.19 (0.99 to 1.42) Freelite λ FLC; the Cusum test for linearity show no significant deviation from linearity (P = 0.470). The result of the Passing and Bablok regression analysis for the κ/λ ratio is as follows: N-Latex ratio = 0.36 (0.29 to 0.41) + 0.33 (0.29 to 0.36) Freelite ratio; the Cusum test for linearity show no significant deviation from linearity (P = 0.130). Significant statistical differences between values of κ FLC, λ FLC and κ/λ ratio obtained using the two assays were found with Wilcoxon’s test (P = 0.003 for κ FLC, P < 0.001 for λ FLC and P < 0.001 for κ/λ ratio, respectively).

Table 1 Comparison of the Freelite and N-Latex FLC methods
N LatexFrelite
sFLC κ (mg/L)sFLC λ (mg/L)κ/λ ratiosFLC κ (mg/L)sFLC λ (mg/L)κ/λ ratio
Lowest value1.52.20.01.11.30.01
Highest value1870.04360.0188.903105.71150.2889.90
Median (95% CI)32.9
(25.0 - 41.7)
23.2
(19.0 - 30.6)
1.10
(0.98 - 1.21)
31.5
(23.5 - 47.9)
15.0
(10.7 - 18.5)
1.80
(1.42 - 2.19)
IQR16.1 - 65.014.3 - 40.40.70 - 2.2015.3 - 79.37.7 - 36.91.00 - 6.20
IQR – interquartile range. Cl – confidence interval. sFLC κ – serum free light chain kappa. sFLC λ – serum free light chain lambda.
Figure 1 Comparison of N-Latex with Freelite for the determination of FLC kappa using the Bland Altman plot. A positive bias indicates higher values for the determination of FLC by Freelite compared with N Latex FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
bm-31-3-030701-f1
{ label needed for fig[@id='f2'] }
Figure 2 Comparison of N-Latex with Freelite for the determination of FLC lambda using the Bland Altman plot. A negative bias indicates higher values for the determination of FLC by N Latex compared with Freelite FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
bm-31-3-030701-f2
{ label needed for fig[@id='f3'] }
Figure 3 Comparison of N-Latex with Freelite for the determination of FLC κ/λ ratio with Bland Altman plot. A positive bias indicates higher values for the determination of FLC κ/λ ratio by Freelite compared with N Latex FLC. FLC – free light chain. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation.
bm-31-3-030701-f3

The Spearman’s rank correlation coefficient was found to be 0.99 (95% CI 0.98 to 0.99) (P < 0.001) for κ FLC, 0.95 (95% CI 0.93 to 0.97) (P < 0.001) for λ FLC, and 0.94 (95% CI 0.91 to 0.96) (P < 0.001) for the κ/λ ratio, respectively. The results of agreement analysis of κ/λ ratio, κ FLC and λ FLC are summarized inFigure 4. We found a weighted kappa value of 0.65 for κ/λ ratio (95% CI 0.52 to 0.78), 0.65 for λ FLC (95% CI 0.53 to 0.78) and 0.90 for κ FLC (95% CI 0.81 to 0.98). The overall number of patients with abnormal κ/λ ratio was 38 (9 patients with ratio < 0.31 and 29 with ratio > 1.56) with the N-Latex compared to 61 (10 patients with ratio < 0.26 and 51 with ratio > 1.65) with Freelite. Only one patient was found to have a pathological low value of κ/λ ratio with Freelite (0.13) and normal value with N-Latex (0.43). This patient especially displayed a large difference in λ FLC concentration, with a high pathological value of 119.2 mg/L measured with Freelite assay compared to only a modest increase (i.e., 39.7 mg/L) with N-Latex, whilst the difference of κ FLC was substantially insignificant (15.6 mg/L and 17.3 mg/L with Freelite and N-Latex, respectively). All patients with abnormally increased κ/λ ratio measured with N-Latex also exhibited abnormal increased with Freelite. Interestingly, 19 patients with increased κ/λ ratio with Freelite displayed normal κ/λ ratio with N-Latex. All these patients displayed a slightly increased value of κ/λ ratio, in a range comprised between 1.67 and 3.37 (median value, 2.16; IQR 1.94-2.44). Bland-Altman plots are shown inFigures 1, 2, and 3{ label needed for fig[@id='f2'] }{ label needed for fig[@id='f3'] }. A positive bias reflects overestimation of κ FLC and κ/λ ratio with Freelite compared to N-Latex (Figure 1 and 3{ label needed for fig[@id='f3'] }), whilst a negative bias underscores underestimation of λ FLC by Freelite compared to N-latex (Figure 2). Finally, a ratio ≥ 100 was found in two patients with both methods, in 3 patients with Freelite and in one patient with N-Latex, respectively.

Figure 4 Concordance analysis for the parameters investigated (N = 93). A: FLC κ/λ ratio (Weighted kappa 0.65 (95% CI 0.52 to 0.78). B: κ FLC (Weighted kappa 0.90 (95% CI 0.82 to 0.98). C: λ FLC (Weighted kappa 0.65 (95% CI 0.53 to 0.78). FLC – free light chain.
bm-31-3-030701-f4

Discussion

Taken together, the results of our study reveal a better correlation between the two methods for κ FLC compared to λ FLC and κ/λ ratio. The technical and analytical characteristics of the N-Latex FLC and Freelite assays have been thoughtfully described elsewhere (7-10). Previous evidence has shown that numerical values produced by the two different systems can have remarkable differences and this aspect has also been evidenced in our study, where higher concentrations are paralleled by an increase in absolute differences (11,12). Several causes can be preferred for explaining inter-assay variability. First, the Freelite method uses polyclonal Ab, whilst the N-Latex is based on a cocktail of monoclonal Ab. It is hence conceivable that less antigenic determinants may be better recognized by the N-Latex assay, whereas some FLC abnormally expressed in plasma cell dyscrasia could escape detection (2,11,12). Previous data underpins that the discrepancies observed between the two methods may be attributable to polymerization of FLC, leading to potential overestimation using the Freelite method, with concurrent underestimation using the N-Latex assay due to binding sites masking (13,14). The N- Latex FLC applications on the BN Nephelometer systems have built-in pre-reaction protocols to secure antigen excess protection. Although this protocol seems highly effective for both FLC, samples containing certain monoclonal FLC display a non-linear dilution trend, with a result of the following dilution yielding a higher result than expected (9). Furthermore, it is always possible that part of the monoclonal intact immunoglobulin would not be correctly folded, with part of the bound light chain unexposed. If one epitope is exposed, part of the reagent is sequestrated by this bound light chain, thus generating results in false-low FLC concentrations in the initial dilution of the assay (15).

Although the vast majority of currently available techniques are based on immunoassays, significant differences in the type of antibodies and in assay design exist. Therefore, their efficiency in coping with the risk of antigen excess, the possibility of missing rare epitopes, the overestimation due to light chain polymerization, the behaviour on serial dilution, and the reproducibility between batches may differ widely. Moreover, due to the lack of an international standard for FLC measurement, it is currently unfeasible to establish whether FLC tests results are underestimated with Freelite or, alternatively, if those obtained with N-Latex are overestimated (16).

Overall, our findings demonstrate that the agreement for identifying patients with pathological FLC values seems substantially good for λ FLC and κ/λ ratio, even excellent for κ FLC, though the absolute values differ considerably between the two methods. Similar evidence was provided by previous studies which compared different FLC assays, also revealing significant absolute differences in FLC concentration, especially in samples with high FLC concentrations (14-18). We also found that the clinical information that can be garnered may differ between these two methods, as we observed 20 patients with altered κ/λ ratio with only Freelite (19 patients with κ/λ ratio above the normal reference range and one below). Our study thus underscores those patients may be differently classified, as being considered at higher risk in case of monoclonal gammopathy of undetermined significance, supposed to have a worse response to treatment in case of MM, or with suspected MG in case of unknown pathological conditions. The iFLC/niFLC ratio is another important information that can be garnered from FLC measurement since a value ≥ 100 is now considered an index of malignancy (19). Therefore, this ratio shall be considered an important element in the initial assessment of patients with suspected myeloma. We found six patients with iFLC/niFLC ratio ≥ 100 with at least one assay in our investigation. It is also worthwhile mentioning here that Schieferdecker et al. have previously recommended that an iFLC/niFLC ratio ≥ 100 with Freelite may correspond to an iFLC/niFLC ratio ≥ 50 with N-Latex (11). By setting the value of 50 with N-Latex as equivalent to 100 with Freelite, the agreement between the two methods was only reached in one such cases, whilst disagreement remained for the other two.

Palladini et al. previously compared FLC values measured with Freelite and N-Latex during diagnosis, prognostication, and therapeutic monitoring of AL. The authors concluded that these two assays have similar diagnostic and prognostic performance, though highlighting that they are not interchangeable, so that patient follow-up should be unformed by using a single assay (20). Notably, some studies reported κ/λ ratio as a high false positive rate in patients without MG (especially in samples with polyclonal hyper-gamma globulinemia) (21). Moreover, a high false negative rate for κ/λ ratio was found in samples with detectable MG (e.g., in the case of lambda light chain monoclonal immunoglobulins) (22). These aspects further reinforce the importance of patient monitoring by always using the same assay.

In conclusion, although the concordance between Freelite and N-Latex appears globally satisfactory, several discrepancies could be evidenced. In keeping with previous reports, our findings show that these assays shall not be considered equivalent and are consequently not interchangeable. Caution must be specially used taken with test results of N-Latex for patient classification, considering that the current IMWG guidelines are based on FLC values obtained with Freelite.

Acknowledgements

We thank Antonia De Luisi, Cristina Vicidomini, Claudio Berenzi, Manuela Scapin and Nadia Trevisan for the great work they do every day in our institution.

Notes

[1] Conflicts of interest Potential conflict of interest

None declared.

References

1 

Graziani MS, Merlini G. Serum free light chain analysis in the diagnosis and management of multiple myeloma and related conditions. Expert Rev Mol Diagn. 2014;14:55–66. https://doi.org/10.1586/14737159.2014.864557 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/24308339

2 

Caponi L, Romiti N, Koni E, Di Fiore A, Paolicchi A, Franzini M. Inter-assay variability in automated serum free light chain assay and their use in the clinical laboratory. Crit Rev Clin Lab Sci. 2020;57:73–85. https://doi.org/10.1080/10408363.2019.1670133 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/31612753

3 

Bradwell AR, Carr-Smith HD, Mead GP, Tang LX, Showell PJ, Drayson MT, et al. Highly sensitive, automated immunoassay for immunoglobulin free light chains in serum and urine. Clin Chem. 2001;47:673–80. https://doi.org/10.1093/clinchem/47.4.673 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/11274017

4 

Dispenzieri A, Kyle R, Merlini G, Miguel JS, Ludwig H, Hajek R, et al. International Myeloma Working Group guidelines for serum-free light chain analysis in multiple myeloma and related disorders. Leukemia. 2009;23:215–24. https://doi.org/10.1038/leu.2008.307 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19020545

5 

Kumar S, Dispenzieri A, Katzmann JA, Larson DR, Colby CL, Lacy MQ, et al. Serum immunoglobulin free light-chain measurement in primary amyloidosis: prognostic value and correlations with clinical features. Blood. 2010;116:5126–9. https://doi.org/10.1182/blood-2010-06-290668 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/20798235

6 

Palladini G, Russo P, Bosoni T, Verga L, Sarais G, Lavatelli F, et al. Identification of amyloidogenic light chains requires the combination of serum-free light chain assay with immunofixation of serum and urine. Clin Chem. 2009;55:499–504. https://doi.org/10.1373/clinchem.2008.117143 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19131635

7 

Velthius H, Knop I, Stam P, van den Broek M, Bos HK, Hol S, et al. N Latex FLC-new monoclonal high-performance assays for the determination of free light chain kappa and lambda. Clin Chem Lab Med. 2011;49:1323–32. https://doi.org/10.1515/cclm.2011.624 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/21663464

8 

White-Al Habeeb NMA, Earle T, Spencer M, Blasutig IM. Evaluation of the N-latex serum free light chain assay on the Siemens BNII analyzer and agreement with The Binding Site FreeLite assay on the SPAPlus. Clin Biochem. 2018;51:90–6. https://doi.org/10.1016/j.clinbiochem.2017.05.009 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28512013

9 

Te Velthius H, Drayson M, Campbell P. Measurement of free light chains with assays based on monoclonal antibodies. Clin Chem Lab Med. 2016;54:1005–14. https://doi.org/10.1515/cclm-2015-0963 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/27010775

10 

Katzmann JA, Clark RJ, Abraham RS, Bryant S, Lymp JF, Bradwell AR, et al. Serum reference intervals and diagnostic ranges for free kappa and free lambda immunoglobulin light chains: relative sensitivity for detection of monoclonal light chains. Clin Chem. 2002;48:1437–44. https://doi.org/10.1093/clinchem/48.9.1437 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/12194920

11 

Schieferdecker A, Horber S, Ums M, Besemer B, Bokemeier K, Weisel K. Comparison of three different serum-free light chain assays - implications on diagnostic and therapeutic monitoring of multiple myeloma. Blood Cancer J. 2020;10:2–9. https://doi.org/10.1038/s41408-019-0267-8 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/31915365

12 

Messiaen AS, De Sloovere MMW, Claus PE, Vercammen M, Van Hoovels L, Heylen O, et al. Performance evaluation of serum free light chain analysis: nephelometry vs turbidimetry, monoclonal vs polyclonal reagents. Am J Clin Pathol. 2017;147:611–22. https://doi.org/10.1093/ajcp/aqx037 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28575180

13 

de Kat Angelino CM, Raymakers R, Teunesen MA, Jakobs JFM, Klasen IS. Overestimation of serum kappa free light chain concentration by immunonephelometry. Clin Chem. 2010;56:1188–90. https://doi.org/10.1373/clinchem.2010.143529 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/20413429

14 

Yang Y, Han X, Zheng G, Cai Z. Comparison of two serum free light chain assays for the diagnosis of primary plasma cell malignant proliferative disease. Health Sci Rep. 2019;2:e113. https://doi.org/10.1002/hsr2.113 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/31049418

15 

Jacobs JF, Hoedemakers RM, Teunissen E, van der Molen RG, Te VH. Effect of sample dilution on two free light chain nephelometric assays. Clin Chim Acta. 2012;413:1708–9. https://doi.org/10.1016/j.cca.2012.04.032 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22588162

16 

Graziani MS. Measurement of free light chains – pros and cons of current methods. Clin Chem Lab Med. 2016;54:1015–20. https://doi.org/10.1515/cclm-2015-1062 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/26812876

17 

Hutchison CA, Harding S, Hewins P, Mead GP, Townsend J, Bradwell AR, et al. Quantitative assessment of serum and urinary polyclonal free light chains in patients with chronic kidney disease. Clin J Am Soc Nephrol. 2008;3:1684–90. https://doi.org/10.2215/CJN.02290508 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/18945993

18 

Fleming CKA, Swarttouw T, de Jat Angelino CM, Jacobs FM, Russcher H. Method comparison of four clinically available assay for serum free light chain analysis. Clin Chem Lab Med. 2019;58:85–94. https://doi.org/10.1515/cclm-2019-0533 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/31730518

19 

Rajkumar SV, Dimopoulos MA, Palumbo A, Blade J, Merlini G, Mateos MV. International Myeloma Working Group updated criteria for the diagnosis of multiple myeloma. Lancet Oncol. 2014;15:e538–48. https://doi.org/10.1016/S1470-2045(14)70442-5 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/25439696

20 

Palladini G, Jaccard A, Milani P, Lavergne D, Foli A, Bender S, et al. Circulating free light chain measurement in the diagnosis, prognostic assessment and evaluation of response of AL amyloidosis: comparison of freelite and N latex FLC assays. Clin Chem Lab Med. 2017;55:1734–43. https://doi.org/10.1515/cclm-2016-1024 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28343171

21 

Singh G. Serum Free Light Chain Assay and κ/λ Ratio: Performance in Patients without Monoclonal Gammopathies: High False-Positive Rate. Am J Clin Pathol. 2016;146:207–14. https://doi.org/10.1093/ajcp/aqw099 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/27473738

22 

Singh G. Serum Free Light Chain Assay and κ/λ Ratio: Performance in Patients with Monoclonal Gammopathy-High False Negative Rate for κ/λ Ratio. J Clin Med Res. 2017;9:46–57. https://doi.org/10.14740/jocmr2802w PubMed: http://www.ncbi.nlm.nih.gov/pubmed/27924175


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.