Skip to the main content

Original scientific paper

https://doi.org/10.5599/jese.2747

Highly sensitive determination of methyldopa and hydrochloro­thiazide using CoMoO4 nanosheets-modified screen-printed electrode

Shams B. Ali orcid id orcid.org/0000-0003-0232-7341 ; Department of Laser and Optoelectronics Engineering, University of Technology, Iraq


Full text: english pdf 615 Kb

versions

downloads: 161

cite


Abstract

A novel screen-printed electrode (SPE) modified by CoMoO4 nanosheets (CoMoO4 NSs) was designed to analyse hydrochlorothiazide. Cyclic voltammetry, chronoamprometry and differential pulse voltammetry using the developed electrode were performed to evaluate its electrochemical behaviour. Further, the anodic peak currents for hydrochlorothiazide at CoMoO4 NSs modified SPE were almost 4 times higher than for the untreated SPE. Between the concentration range of 0.1 to 400.0 μM, the modified electrode's response was directly proportional to the concentration, and it could detect hydrochlorothiazide at a very low level of 0.05 μM (S/N = 3). Hydrochlorothiazide was also measured using the modified electrode when methyldopa was present. The sensor offers various advantages for the analysis of hydrochlorothiazide detection, including moderate cost, ease of preparation and consi­derable sensitivity.

Keywords

Voltammetry; drug analysis; real sample analysis; blood pressure drugs

Hrčak ID:

330701

URI

https://hrcak.srce.hr/330701

Publication date:

28.4.2025.

Visits: 340 *