Arsen Pavić ; Clinical Hospital Center Split, Spinčićeva 5, 21000 Split, Croatia
Janoš Kodvanj ; Faculty of Mechanical Engineering and Naval Architecture, Ivana Lučića 5, 10000 Zagreb, Croatia
Martin Surjak ; Faculty of Mechanical Engineering and Naval Architecture, Ivana Lučića 5, 10000 Zagreb, Croatia


Puni tekst: hrvatski pdf 1.850 Kb

str. 995-999

preuzimanja: 625

citiraj

Puni tekst: engleski pdf 1.850 Kb

str. 995-999

preuzimanja: 490

citiraj


Sažetak

External fixators are used to treat open and closed fractures of long bones. Stiffness and strength of the fixator greatly affect the final outcome of treatment. An adequate stability of bone fragments provides better fracture healing and early rehabilitation of patients. The aim of this study was to evaluate biomechanical stability of the newly developed unilateral external fixator for proximal tibial fractures. In order to compare the properties of the prototype of the novel fixator and commonly used external fixator Orthofix, static and dynamic tests were conducted. Two different types of static tests were carried out, axial compression test and bending test. In static tests displacements were measured and calculated by using the non-contact optical measuring system Aramis. The results indicated that the novel fixator has better outcomes in bending, while the conventional fixator outperformed the new one in axial compression. There was no significant difference in cyclic test.

Ključne riječi

external fixator; optical system Aramis; stability

Hrčak ID:

112318

URI

https://hrcak.srce.hr/112318

Datum izdavanja:

20.12.2013.