Skoči na glavni sadržaj

Kratko priopćenje

A Quasi-Adiabatic Reaction Calorimeter for Measurements in Liquids

B. B. Radak ; Radiation Chemistry Department, Boris Kidric Institute of Nuclear Sciences, Beograd-Vinča
V. I. Matić ; Radiation Chemistry Department, Boris Kidric Institute of Nuclear Sciences, Beograd-Vinča
V. M. Radak ; Institute of Chemistry and Physical Chemistry, Faculty of Science, Beograd, Serbia, Yugoslavia


Puni tekst: engleski pdf 2.148 Kb

str. 25-30

preuzimanja: 239

citiraj


Sažetak

A quasi adiabatic reaction and solution calorimeter was constructed
and tested using tris(hydroxymethyl)aminomethane dissolution
in 0.1 mol dm-3 HCl as the reference (standard) process.
The precision of the calorimeter amounts to approximately ± 1 per
cent and its sensitivity to 4 J dm-a.

Ključne riječi

Hrčak ID:

196408

URI

https://hrcak.srce.hr/196408

Datum izdavanja:

25.3.1977.

Posjeta: 609 *