What Can We Actually See Using Computer Aided Detection in Mammography?
Marina Džoić Dominković
; Department of Radiology, Orašje General Hospital, Orašje, Bosnia and Herzegovina
Gordana Ivanac
; Department of Diagnostic and Interventional Radiology, Dubrava University Hospital, Zagreb, Croatia; University of Zagreb, School of Medicine, Zagreb, Croatia
Niko Radović
; Department of Diagnostic and Interventional Radiology, Dubrava University Hospital, Zagreb, Croatia
Mislav Čavka
; Department of Diagnostic and Interventional Radiology, Dubrava University Hospital, Zagreb, Croatia; University of Zagreb, School of Medicine, Zagreb, Croatia
Sažetak
The main goal of this study was to compare the results of computer aided detection
(CAD) analysis in screening mammography with the results independently obtained by two radiologists
for the same samples and to determine the sensitivity and specificity of CAD for breast lesions.
A total of 436 mammograms were analyzed with CAD. For each screening mammogram, the changes
in breast tissue recognized by CAD were compared to the interpretations of two radiologists. The
sensitivity and specificity of CAD for breast lesions were calculated using contingency table. The
sensitivity of CAD for all lesions was 54% and specificity 16%. CAD sensitivity for suspicious lesions
only was 86%. CAD sensitivity for microcalcifications was 100% and specificity 45%. CAD mainly
‘mistook’ glandular parenchyma, connective tissue and blood vessels for breast lesions, and blood vessel
calcifications and axillary folds for microcalcifications. In this study, we confirmed CAD as an excellent
tool for recognizing microcalcifications with 100% sensitivity. However, it should not be used as
a stand-alone tool in breast screening mammography due to the high rate of false-positive results.
Ključne riječi
Breast neoplasm; Image interpretation computer-assisted; Mammography; Computer aided detection (CAD); CAD sensitivity; CAD specificity