Izvorni znanstveni članak
The dispersion of an axicon-scanned Fabry-Perot spectrometer
M. Platiša
; Institute of Physics, Beograd
Sažetak
A correlation is found between the linear dispersion of an Axicon-scanned Fabry-Perot ( A-F-P) spectrometer and a Fabry-Perot (F-P) interferometer. The position of the axicon, as the scanning element, is related to the parameters: the focal length of the imaging lens, the wavelength, the spacing of the F -P plates, the refraction index and the vertex angle of the axicon. A functional relation is established between the wavelength and the displacement of the axicon, from the focal plane stop. The linear dispersion of the A-F-P spectrometer is derived and shown that, at a given wavelength, it increases linearly with increasing focal length of the imaging lens as the square root of the spacing of the F-P plates. The dispersion also increases through the cot function if the vertex angle and. the refractive index of the axicon decrease.
Ključne riječi
Hrčak ID:
321981
URI
Datum izdavanja:
8.5.1973.
Posjeta: 0 *