Comparison of CT and angiography-derived common femoral artery dimensions in patients with vascular injury following transcatheter aortic valve implantation

Sara Dolički orcid id orcid.org/0009-0002-7543-8157 ; University of Zagreb, School of Medicine, Zagreb, Croatia
Zvonimir Ostojić orcid id orcid.org/0000-0003-1762-9270 ; University Hospital Center Zagreb, Zagreb, Croatia
Anamaria Ostović ; University Hospital Center Zagreb, Zagreb, Croatia
Luka Perčin orcid id orcid.org/0000-0003-0497-6871 ; University Hospital Center Zagreb, Zagreb, Croatia
Hrvoje Jurin ; University Hospital Center Zagreb, Zagreb, Croatia
Tomislav Krčmar orcid id orcid.org/0000-0003-4689-1673 ; University Hospital Center Zagreb, Zagreb, Croatia
Joško Bulum orcid id orcid.org/0000-0002-1482-6503 ; University of Zagreb, School of Medicine, Zagreb, Croatia


Puni tekst: engleski pdf 144 Kb

str. 225-225

preuzimanja: 6

citiraj

Preuzmi JATS datoteku


Sažetak

Ključne riječi

angiography; aortic valve disease; transcatheter aortic valve replacement; vascular system injuries

Hrčak ID:

337267

URI

https://hrcak.srce.hr/337267

Datum izdavanja:

30.10.2025.

Posjeta: 13 *



Introduction: Access site-related vascular injury (ASRVI) following percutaneous transfemoral transcatheter aortic valve implantation (TAVI) remains a notable complication. Most ASRVI cases can be effectively treated percutaneously using balloon dilatation or stent graft (SG) implantation (1). Device sizing is typically based on angiographic imaging of the common femoral artery (CFA) at the procedure’s start. This study aimed to compare angiographic CFA dimeter with those derived from computed tomography (CT) to evaluate whether CT-based measurements could guide device sizing for ASRVI treatment.

Patients and Methods: This single-center analysis included all TAVI patients who developed ASRVI and were treated with balloon dilatation or SG implantation. CFA dimensions from pre-TAVI CT scans were compared to angiography performed at the beginning of the procedure. Diameter differences were calculated to assess percentage of the correspondence.

Results: Among 54 patients with ASRVI post-TAVI, all underwent balloon or SG treatment. Device diameter was determined using angiography. Mean CFA diameter via angiography was 8.01 ± 1.24 mm, significantly larger than CT-based mean diameter (7.16 ± 1.60 mm, p=0.003), but not different from CT-based maximal diameter (7.93 ± 1.42 mm, p=0.75). Case-by-case comparison between maximal CT and angiography derived CFA diameter showed: good accordance (< 0.3 mm) in 16 cases (29.63%), moderate (0.3 – 0.8 mm) in 18 cases (33.33%), and discordance (> 0.9 mm) in 20 cases (37.4%). In the moderate group, CT overestimated CFA in 10 cases by mean of 0.57 mm, and underestimated in 8 by mean of 0.47 mm. In the discordant group, CT overestimated CFA in 7 cases (mean 1.5 mm) and underestimated in 13 (mean 1.23 mm).

Conclusion: CT-based CFA measurements show poor agreement with angiography and should not be used for device sizing in ASRVI treatment post-TAVI.