Key incidents identified by the Royal College of Pathologists of Australasia Quality Assurance Programs Key Incident Monitoring and Management EQA program

Anushka Jayanetti ; Royal College of Pathologists of Australasia Quality Assurance Programs, Sydney, Australia
David Roxby ; Royal College of Pathologists of Australasia Quality Assurance Programs, Sydney, Australia; Molecular Medicine and Pathology, College of Medicine and Public Health, Flinders Medical Centre, Adelaide, Australia
Tony Badrick ; Royal College of Pathologists of Australasia Quality Assurance Programs, Sydney, Australia *

* Dopisni autor.


Puni tekst: engleski pdf 244 Kb

str. 50-61

preuzimanja: 66

citiraj

Preuzmi JATS datoteku


Sažetak

Introduction: The Royal College of Pathologists of Australasia Quality Assurance Programs (RCPAQAP) Key Incident Monitoring and Management EQA Program (KIMMS) aims to monitor the laboratory quality system’s pre- and post-analytical phases. The purpose of this paper is to describe the most common incidents from 2024.
Materials and methods: The KIMMS program has four surveys a year, collecting data from the previous three months, with preanalytical and postanalytical incident reporting of 35 incident types. Participants are asked to capture the number of episodes and the number of incidents per quarter of the year.
Results: The four 2024 surveys received an average of 111 responses, with 55,329,998 episodes recorded and 1,496,708 incidents identified. The findings from the 2024 program are that the incident “No specimen received” appears to have the highest 80th percentile across the patient sources. The commonest site of error is the Emergency Department (ED), with an 80th percentile overall.
Conclusions: The KIMMS data provide valuable, regular and reproducible benchmarking data for the pre- and post-analytical phases of the total testing cycle.

Ključne riječi

preanalytical phase; incident reporting; risks management; quality improvements; Benchmarking

Hrčak ID:

342241

URI

https://hrcak.srce.hr/342241

Datum izdavanja:

15.2.2026.

Posjeta: 207 *




Introduction

Pre- and postanalytical errors represent most errors in the total testing cycle leading to increased risk of misdiagnosis and treatment of patients (1-3). Identification, phlebotomy, collection tube and sample preservation and transport related errors are common sources of error, particularly when the specimen collection staff are not laboratory-controlled (4). Postanalytical errors mainly involve delayed or incomplete delivery of results to referring doctors. Despite this, few benchmarking EQA programs are available for medical laboratories to assess these error rates objectively (5,6). Indicators of the extra-analytical phases of the Total Testing Process (TTP) have been developed in several countries, such as Australia and New Zealand, The US Brazil, and Spain/Catalonia, and other surveys and programs have been promoted in the UK, China and Croatia (7-14). In 2008, the IFCC launched a Working Group named “Laboratory Errors and Patient Safety” (WG: LEPS) to identify Quality Indicators (QIs) and related quality specifications that (i) produce benchmarks from comparing laboratories, (ii) promote error reduction and (iii) increase patient safety. The IFCC has developed QIs, which laboratories in several countries have evaluated, and the WG: LEPS has reported preliminary results (15-17).

To reduce pre- and post-analytical error, laboratories need to focus on these phases of the total testing cycle with relevant QIs and benchmarks from external sources (18-20). As with any form of benchmarking, laboratories must collect and track their QIs and then compare them against the relevant peer group. Initially, this ensures they collect the same data as their peers, then compare their performance. If there are significant differences after data collection error has been eliminated, then a review of the process is warranted. To ensure there is confidence in this comparison, a sufficiently large peer group collecting the same QIs needs to be available, and the data collection needs to be regular enough to detect trends (21).

But collecting QI data in pathology can be tedious. Benchmarking processes are far more difficult than standard analytical EQA because measurement may be complex and not standardised, data collection may not be easy, and targets are unavailable. As with many benchmarking exercises, there have been problems with the definitions and robustness of the QIs. Key Incident Monitoring and Management EQA Program has tried to ensure that participants collect and submit the same data for each of the QIs. While progress has been made in this direction, the data is still not always homogeneous. One of the aims of the KIMMS program is to produce a list of key incidents that should be tracked by an LIS. Also, changing a process is more difficult than identifying an error in an assay system.

The Royal College of Pathologists of Australasia Quality Assurance Programs (RCPAQAP) Key Incident Monitoring and Management EQA Program (KIMMS) started in 2011 and aims to monitor the pre- and post-analytical phase of the laboratory quality system (7). Key Incident Monitoring and Management EQA Program measures risks from incidents recorded by participant laboratories and compares these to those of peer organisations. Incidents and risks are related but not equal (22). The program assesses risk based on the consequences to the patient, the probability that the event will occur, and how easy it is to detect the incident. A reduction in probability or an increase in detectability can reduce the risk. A two-phase risk calculation is used in KIMMS, where a harm factor equals consequences multiplied by likelihood (21). The risk score equals harm times detectability (23).

As an example of this risk score process, we present this case where the overall performance summary highlights any “incident” that recorded a rate greater than 80% of all participants. This is presented in a table, which includes columns for the consequence, probability and detectability rating (Tables 1-3{ label needed for table-wrap[@id='t2'] }{ label needed for table-wrap[@id='t3'] }). The KIMMS Advisory Committee has pre-determined the consequence rating, so this field is pre-populated.

Table 1 Example risk score table from the KIMMS report
Parameter out of allowable limitsConsequence ratingProbability ratingDetectability ratingRisk scoreOrganisational response
Inpatient - hemolysis rate = 40025110Example: Although this is a low-risk incident, high compared to peers and to our collectors (0.2). Meeting with hospital organised 7/10/22
{ label needed for table-wrap[@id='t2'] }
Table 2 KIMMS consequence scale
ScaleNameDefinition
1Negligible/MinimalMinimal, delay, inconvenience
2Marginal/MinorRecollect required
3Significant/ModerateDelayed management (non-malignant) and/or medical treatment
4Serious/MajorDelayed diagnosis (malignant) and/or surgical treatment
5Critical/CatastrophicSerious harm to multiple patients and/or patient death
{ label needed for table-wrap[@id='t3'] }
Table 3 KIMMS probability scale
ScaleNameExample Definition
1Rare< 1/year
2Unlikely1 per year
3Occasional1 per month
4Likely1 per week
5Frequent1 per day or more

Results review

It is recommended that participants review their results by applying the probability and detectability rating to determine the risk score (Tables 2-4{ label needed for table-wrap[@id='t3'] }{ label needed for table-wrap[@id='t4'] }). The risk score is calculated by multiplying the three ratings (consequence rating x probability rating x detectability rating). The risk score can be used as a guide to organisational risk; from that, a decision is made as to the next step(s).

{ label needed for table-wrap[@id='t4'] }
Table 4 KIMMS detectability scale
ScaleNameDefinition
1Detected> 95%
2Most detected75-95%
3Half detected25-75%
4Most not detected< 5-25%
5Not detected< 5%

In the above example, the consequence, probability and detectability rating give a risk score of 10. This would be considered low risk. Apart from flagging that the risk is higher than other participants’ inpatient collections, it is unlikely that any further immediate action is required. This incident should be monitored closely to ensure it does not deteriorate.

Risk score

The score is calculated by multiplying the consequences (Table 1) x probability (Table 2) x detectability (Table 3).

The purpose of this paper is to report the most common incidents that have been detected in the 2024 KIMMS program.

Materials and methods

The KIMMS program has four surveys a year, collecting data from the previous three months, with preanalytical and postanalytical incident reporting of 35 incident types that have been pre-determined by the RCPAQAP KIMMS Advisory Committee. Data is submitted via the RCPAQAP myQAP portal, where result entry instructions are included. Participants are asked to capture the number of episodes and the number of incidents per quarter of the year.

Data is then analysed using RCPAQAP in-house data analysis software, where each specific incident reported to KIMMS is calculated per 100,000 episodes (number of incidents divided by the number of episodes, multiplied by 100,000), known as the rate. Incident rates are compared to peer groups and flagged red if outside the best performers, which is set at 80% of all results. This summary will focus on 2024 data.

A single request usually covers an episode and may contain one or more samples. It may include more than one request, but the samples are collected simultaneously, i.e., samples received together in a bag with a request form equal one episode.

Participants range from small to large pathology laboratories, where size is determined by the number of episodes per quarter (small: < 100,000, medium: 100,000-500,000, large: > 500,000). This data is also collected for each submission. The participants include a mix of public and private organisations from across Australia and a couple of international sites.

Patient sources

The KIMMS program was renewed in 2023 to allow for further comparison by separating data into categories such as the patient source (Inpatient/Outpatient/Community/Emergency Department (ED)).

The Inpatient patient source refers to pathology testing performed on an admitted hospital patient (public or private hospital). The Outpatient source includes pathology requests and specimens that are collected while a patient attends a clinic run by a hospital. This includes day patients and cancer treatment centres. The Emergency Department patient source refers to pathology testing performed on a patient attending an Emergency Department (ED)/Emergency Room (ER)/Accident and Emergency (A&E)/Casualty ward/department. The community patient source is for when the pathology request is generated anywhere, but the specimens are collected in settings other than the previously mentioned patient sources. This includes doctors’ surgeries, designated collection centres (even within a hospital), nursing homes, home visits, and workplace collections.

The KIMMS program also offers a Mixed patient source option for those organisations that are unable to differentiate their data between patient sources. In the final calculations, these are analysed with a total of all results.

Results

The four 2024 surveys received an average of 111 responses, with 55,329,998 episodes recorded and 1,496,708 incidents identified.

Table 5 summarises the total episodes and incidents reported per survey.Table 6 summarises the key results for the top three incidents from all sites, andTable 7 shows the key results per size of organisation.Tables 8-10{ label needed for table-wrap[@id='t9'] }{ label needed for table-wrap[@id='t10'] } display the average 80th percentile cut-off rates and standard deviations (SDs) per patient source and the Mixed/Total category for the Test Request/Collection incident categories, as well as the Test Registration/Analytical/Postanalytical incident categories from the four 2024 KIMMS surveys.

Table 5 The total number of episodes and incidents reported in each 2024 KIMMS survey
Total episodesTotal incidents
Survey 111,930,241274,277
Survey 213,123,033295,720
Survey 314,832,081484,514
Survey 415,444,643442,197
Table 6 The top three incidents with the highest incident rates in 2024, per the source of the patient
InpatientOutpatientCommunityEmergency DepartmentMixed/Total
Incorrect specimen type or container or acid (200)Incorrect specimen type or container or acid (160)No specimen received (272)Incorrect specimen type or container or acid (344)No specimen received (271)
Specimen clotted or other clotting issue (250)Essential clinical indication for a test is not provided (227)Essential clinical indication for the test is not provided (130)Specimen clotted or other clotting issue (428)Insufficient specimen (126)
Specimen hemolysed (230)No specimen received (320)Insufficient specimen (92)Specimen hemolysed (711)Specimen hemolysed (127)
Values in brackets represent the rates.
Table 7 The top three incidents with the highest incident rates in 2024, per the size of the organisation (episodes per quarter)
Small organisation
(< 100,000 episodes)
N = 89
Medium organisation (100,000-500,000 episodes)
N = 13
Large organisation
(> 500,000 episodes)
N = 9
Essential collection date/time not provided or discrepant between specimen and request (401)No specimen received (407)No specimen received (528)
Incorrect specimen type or container or acid (589)Insufficient specimen (322)Essential clinical indication for the test is not provided (331)
Specimen clotted or other clotting issue (466)Specimen clotted or other clotting issue (411)Specimen hemolysed (280)
The “N” value refers to the average number of submissions per survey in 2024 that selected that particular size of organisation.
Table 8 Average 80th percentile cut-off rates per patient source for the Test Request/Collection incident categories from the four 2024 KIMMS surveys
InpatientOutpatientCommunityED
Clarification of tests required109.4140.20.9263.6
Insufficient requester details or signature missing3.20.07.58.1
Unlabelled specimen or request210.4247.5128.6272.2
Insufficient patient ID on specimen and/or request113.491.385.2100.4
ID mismatch between specimen and/or request161.8100.961.9200.6
Specimens from wrong patient (WSIT)35.524.11.644.8
Essential collection date/time not provided or discrepant between specimen and request14.19.32.334.3
Essential signature missing or discrepant on transfusion sample and/or request113.683.017.9185.6
Essential clinical indication for test not provided41.778.840.318.5
Essential specimen type/site not provided6.00.010.93.3
Incorrect patient preparation8.20.012.29.0
Incorrect specimen type or container or acid228.7158.075.5208.3
No specimen received421.4521.7378.5589.4
Insufficient specimen335.5252.6147.8458.2
Specimen incorrect fill leading to incorrect specimen:additive ratio153.788.842.0199.3
Specimen clotted or other clotting issue385.4161.651.1679.7
Specimen contaminated - non-microbiological40.717.84.633.6
Specimen contaminated - microbiological0.00.00.11.8
Specimen hemolysed75.137.144.1416.0
Specimen leaking0.00.03.813.1
Incorrect transport / storage temperature / handling46.565.250.933.0
Transport delay leading to specimen being too old to test109.7135.145.882.6
{ label needed for table-wrap[@id='t9'] }
Table 9 Average 80th percentile cut-off rates for the Mixed/Total patient source for the Test Request/Collection incident categories from the four 2024 KIMMS surveys
Mixed/Total
Clarification of tests required44.3
Insufficient requester details or signature missing9.1
Unlabelled specimen or request227.2
Insufficient patient ID on specimen and/or request78.0
ID mismatch between specimen and/or request120.5
Specimens from wrong patient (WSIT)28.0
Essential collection date/time not provided or discrepant between specimen and request12.9
Essential signature missing or discrepant on transfusion sample and/or request102.4
Essential clinical indication for test not provided53.1
Essential specimen type/site not provided11.1
Incorrect patient preparation11.0
Incorrect specimen type or container or acid178.7
No specimen received425.7
Insufficient specimen290.4
Specimen incorrect fill leading to incorrect specimen: additive ratio129.4
Specimen clotted or other clotting issue359.8
Specimen contaminated - non-microbiological26.7
Specimen contaminated - microbiological0.0
Specimen hanalytemolysed83.9
Specimen leaking8.6
Incorrect transport / storage temperature / handling49.4
Transport delay leading to specimen being too old to test94.9
{ label needed for table-wrap[@id='t10'] }
Table 10 Average 80th percentile cut-off rates for all results for the Test Registration/Analytical/Postanalytical incident categories from the four 2024 KIMMS surveys
Total
Incorrect unique specimen identifier8.4
Patient ID - wrong patient7.1
Errors in transcription of patient demographic information31.7
Incorrect or missed tests183.1
Incorrect or missing specimen type, site, collection time26.0
Incorrect requesting or copy doctor131.7
Internal laboratory process incident - unfixable43.0
Within laboratory ID error6.4
Intra or inter-laboratory specimen lost or misplaced - irreplaceable0.2
Intra or inter-laboratory specimen lost or misplaced - replaceable10.0
Failures of clinical handover high-risk (critical) results0.1
Failures of clinical handover non-critical results0.1
Amended reports, significant patient impact47.7

Figures 1-3{ label needed for fig[@id='f2'] }{ label needed for fig[@id='f3'] } are bar graphs highlighting the average 80th percentile cut-off rates per patient source and the Mixed/Total category for the Test Request/Collection incident categories, as well as the Test Registration/Analytical/Postanalytical incident categories from the four 2024 KIMMS surveys.

Figure 1 Average 80th percentile cut-off rates per patient source for the Test Request/Collection incident categories from the four 2024 KIMMS surveys.
bm-36-1-010701-f1
{ label needed for fig[@id='f2'] }
Figure 2 Average 80th percentile cut-off rates for the Mixed/Total patient source for the Test Request/Collection incident categories from the four 2024 KIMMS surveys.
bm-36-1-010701-f2
{ label needed for fig[@id='f3'] }
Figure 3 Average 80th percentile cut-off rates for all results for the Test Registration/Analytical/Post-Analytical incident categories from the four 2024 KIMMS surveys.
bm-36-1-010701-f3

Discussion

The findings from the 2024 program are that the incident “No specimen received” appears to have the highest 80th percentile across the patient sources. The cost of a recollection is significant, both in organisational cost and patient time (24). More importantly, it is an error that can be reduced once identified. Suggested mechanisms to reduce this incident would be computer-generated, or manual checklists of collection tubes required for a given set of requested tests.

The commonest site of error is the ED, with an 80th percentile overall. The primary cause is the non-professional collector and the collection process (4,25-27). The solution could include better training of nursing and medical staff in the ED, though this has had poor long-term effects (25). The issue may be costly enough for the organisation to place phlebotomists in ED (28).

The 80th percentiles are consistent for all incidents throughout the year, meaning the occurrence of incidents is also consistent, validating the data collection process. The 80th percentiles are greater overall for incidents that fall under the Collection incident category, such as “No specimen received”, “Insufficient specimen”, “Specimen clotted” and “Specimen hemolysed”. Sites that perform greater than the 80th percentile have the incidents flagged for review. It is then up to the participants to calculate the risk of the incident to their organisation by referring to our KIMMS Matrix Table (https://rcpaqap.com.au/kimms-matrix-table/), which comprises consequence, probability and detectability. Specific incidences, such as WSIT, may have lower rates reported than the actual values, as the detectability of such incidences is low. This is why in the KIMMS risk matrix table, a higher value for detectability can be assigned to increase the overall risk score and severity of these incidents.

The KIMMS program has been operating since 2011 with several improvements (7,21,23). Significant findings and trends have been separately published (21,29). This experience of participants reporting using standardised definitions of incidents has led to consistent data collections as can be seen from year-on-year results. It has taken some years for this to occur, as there is much variation in what laboratories call episodes and what their capability is to accurately collect these data. The QIs used are relevant to Australian laboratory practice and share commonality with the IFCC set. It is difficult to compare results from different publications as many are one-off surveys rather than a system of regular data collection.

The limitations of this program’s data collection process include some manual data collection processes and likely missing data for some QIs, such as WSIT, because the laboratory is unable to identify these errors (30).

In conclusion, the KIMMS data provide valuable, regular and reproducible benchmarking data for the pre- and post-analytical phases of the total testing cycle. The incidents represent areas where the risk to patients can be reduced.

Appendices

Appendix
Table 11 All incidents measured in the KIMMS program and explanation of each incident
CategoriesIncidentExplanation of incident
Test requestClarification of tests requiredCollector or laboratory required to contact requestor to confirm test(s) requested as information not clear or illegible on request
Insufficient requester details or signature missingThis occurs when the information supplied is not sufficient to know which doctor should receive the report, or the request is not signed when a signature is required
Unlabelled specimen and/or requestThis is when sample or request has no patient identifiers
Collection: IdentificationInsufficient patient ID specimen and/or requestMinimum identifiers are full name and DOB. Three identifiers may be required e.g., transfusion samples
ID mismatch between specimen and/or requestID on specimen label does not match accompanying request form or electronic order
Specimen from wrong patient (WSIT)Labelling of specimen and request are concordant, but the actual sample in the container is from a different patient
Essential collection date and/or time not provided or discrepant between specimen and requestSpecimen and form must align especially essential for a Transfusion request
Collection: DocumentationEssential signature missing or discrepant on transfusion sample and/or requestThis refers to collector and patient signature
Essential clinical indication for test not providedCollector fails to record the clinical indication for the test that is required by the laboratory to aid in the interpretation of results
Essential specimen type/site not providedSpecimen type and/or site not provided when information is required to aid in appropriate testing and interpretation of results by the laboratory
Incorrect patient preparationPatient not fasting, medication taken instead of being withheld, preparation procedure for test not followed
Incorrect specimen type or container or acidIncorrect specimen type collected/submitted for test request
No specimen receivedNo sample received to do the test
Insufficient specimenInsufficient specimen to do any or all tests requested
Collection: SpecimenSpecimen incorrect fill leading to incorrect specimen: additive ratioUnder or overfilled Citrate (COAG) tubes
Specimen clotted or other clotting issueMost commonly EDTA tubes, but any tube with anti-coagulant may be clotted
Specimen contaminated - non microbiologicalE.g. diluted from IV drip; or order of draw incorrect
Specimen contaminated - microbiologicalContaminated microbiology samples that require a recollection
Specimen hemolysedHemolysis should be counted if the sample is deemed unsuitable for one or more tests due to the level of hemolysis
Specimen leakingAny leakage should be recorded, even if there is still sufficient specimen to do the test
Collection: Transport, storage and handlingIncorrect transport/storage temperature/handlingStored and/or transported at wrong temperature or handling not in accordance with documented procedure
Transport delay leading to specimen being too old to testAny delay in transport
Incorrect unique specimen identifierAlso known as episode; laboratory or accession number. It is the unique identifier for that specimen from that patient collected/submitted on that date/time
Patient ID wrong patientThis occurs when the wrong patient is chosen at data entry from a list of patients with the same, or similar name. Request associated to incorrect patient demographics
Test registrationError in transcription of patient demographic informationThis occurs when information is incorrectly entered from a request. Registration data manually entered
Incorrect or missed testsTest on request form not added at registration or incorrect test added
Incorrect or missing specimen type, site, collection timeThis is when the information is available on the request form or electronic order but is not entered into the LIS
Incorrect requesting or copy doctorReferring clinician or copy to clinician not registered to receive copy of the report
Internal laboratory process incident - unfixableThis refers to any incident that happens after the sample has arrived at the laboratory that is not directly related to performing the test
AnalyticalWithin laboratory ID errorError transferring patient Identification within the laboratory
Intra or inter-laboratory specimen lost or misplaced - irreplaceableSpecimens lost during intra and inter-laboratory transfer that are irreplaceable
Intra or inter-laboratory specimen lost or misplaced - replaceableSpecimens lost during intra and inter-laboratory transfer that can be recollected
Failure of clinical handover high-risk (critical) resultsFailure to communicate high-risk results to the relevant party within agreed guidelines as documented within your organisation
PostanalyticalFailure of clinical handover non-critical resultsRequests to re-send copy of report due to non-receipt by the requestor, including referring laboratories
Amended report. Significant patient impactThis is amended report due to a diagnostic error or incorrect result being released
DOB - date of birth. LIS - laboratory information system.

Notes

[1] Conflicts of interest Potential conflict of interest

None declared.

Data availability statement

All data generated and analyzed in the presented study are included in this published article.

References

1 

Plebani M. Errors in laboratory medicine and patient safety: The road ahead. Clin Chem Lab Med. 2007;45:700–7. https://doi.org/10.1515/CCLM.2007.170 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17579520

2 

Carraro P, Plebani M. Errors in a stat laboratory: types and frequencies 10 years later. Clin Chem. 2007;53:1338–42. https://doi.org/10.1373/clinchem.2007.088344 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17525103

3 

Plebani M. Exploring the iceberg of errors in laboratory medicine. Clin Chim Acta. 2009;404:16–23. https://doi.org/10.1016/j.cca.2009.03.022 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19302995

4 

McCaughey EJEJ, Vecellio E, Lake R, Li L, Burnett L, Chesher D, et al. Key factors influencing the incidence of hemolysis: A critical appraisal of current evidence. Crit Rev Clin Lab Sci. 2017;54:59–72. https://doi.org/10.1080/10408363.2016.1250247 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28013559

5 

Sciacovelli L, Plebani M. The IFCC Working Group on laboratory errors and patient safety. Clin Chim Acta. 2009;404:79–85. https://doi.org/10.1016/j.cca.2009.03.025 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/19328194

6 

Kristensen GBB, Aakre KM, Kristoffersen AH, Sandberg S. How to conduct External Quality Assessment Schemes for the pre-analytical phase? Biochem Med (Zagreb). 2014;24:114–22. https://doi.org/10.11613/BM.2014.013 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/24627720

7 

Badrick T, Gay S, Mackay M, Sikaris K. The key incident monitoring and management system - History and role in quality improvement. Clin Chem Lab Med. 2018;56:264–72. https://doi.org/10.1515/cclm-2017-0219 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28771429

8 

Nakhleh RE, Souers RJ, Bashleben CP, Talbert ML, Karcher DS, Meier FA, et al. Fifteen years’ experience of a college of American pathologists program for continuous monitoring and improvement. Arch Pathol Lab Med. 2014;138:1150–5. https://doi.org/10.5858/arpa.2014-0148-OA PubMed: http://www.ncbi.nlm.nih.gov/pubmed/25171697

9 

Meier FA, Souers RJ, Howanitz PJ, Tworek JA, Perrotta PL, Nakhleh RE, et al. Seven Q-Tracks monitors of laboratory quality drive general performance improvement: Experience from the College of American Pathologists Q-Tracks program 1999-2011. Arch Pathol Lab Med. 2015;139:762–75. https://doi.org/10.5858/arpa.2014-0090-CP PubMed: http://www.ncbi.nlm.nih.gov/pubmed/26030245

10 

Shcolnik W, De Oliveira CA, De São José AS, De Oliveira Galoro CA, Plebani M, Burnett D. Brazilian laboratory indicators program. Clin Chem Lab Med. 2012;50:1923–34. https://doi.org/10.1515/cclm-2012-0357 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/23096110

11 

Kirchner MJ, Funes VA, Adzet CB, Clar MVD, Escuer MI, Girona JM, et al. Quality indicators and specifications for key processes in clinical laboratories: A preliminary experience. Clin Chem Lab Med. 2007;45:672–7. https://doi.org/10.1515/CCLM.2007.122 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/17484633

12 

Barth JH. Clinical quality indicators in laboratory medicine: A survey of current practice in the UK. Ann Clin Biochem. 2011;48:238–40. https://doi.org/10.1258/acb.2010.010234 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/21367882

13 

Barth JH. Clinical quality indicators in laboratory medicine. Ann Clin Biochem. 2012;49:9–16. https://doi.org/10.1258/acb.2011.011126 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22042979

14 

Simundic AM, Topic E. Quality Indicators. Biochem Med (Zagreb). 2008;18:311–9. https://doi.org/10.11613/BM.2008.027

15 

Sciacovelli L, Padoan A, Aita A, Basso D, Plebani M. Quality indicators in laboratory medicine: State-of-the-art, quality specifications and future strategies. Clin Chem Lab Med. 2023;61:688–95. https://doi.org/10.1515/cclm-2022-1143 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/36660807

16 

Plebani M. Quality indicators to detect pre-analytical errors in laboratory testing. Clin Biochem Rev. 2012;33:85–8. PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22930602

17 

Sciacovelli L, Lippi G, Sumarac Z, West J, Garcia Del Pino Castro I, Furtado Vieira K, et al. Quality Indicators in Laboratory Medicine: The status of the progress of IFCC Working Group “laboratory Errors and Patient Safety” project. Clin Chem Lab Med. 2017;55:348–57. https://doi.org/10.1515/cclm-2016-0929 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/27988505

18 

Aita A, Sciacovelli L, Plebani M. Extra-analytical quality indicators - where to now? Clin Chem Lab Med. 2018;57:127–33. https://doi.org/10.1515/cclm-2017-0964 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/29252191

19 

Plebani M, Sciacovelli L, Aita A, Padoan A, Chiozza ML. Quality indicators to detect pre-analytical errors in laboratory testing. Clin Chim Acta. 2014;432:44–8. https://doi.org/10.1016/j.cca.2013.07.033 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/24012653

20 

Sciacovelli L, Padoan A, Aita A, Basso D, Plebani M. Quality indicators in laboratory medicine: State-of-the-art, quality specifications and future strategies. Clin Chem Lab Med. 2023;61:688–95. https://doi.org/10.1515/cclm-2022-1143 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/36660807

21 

Gay S, Badrick T. Changes in error rates in the Australian key incident monitoring and management system program. Biochem Med (Zagreb). 2020;30:020704. https://doi.org/10.11613/BM.2020.020704 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/32292282

22 

Badrick T, Gay S, McCaughey EJEJ, Georgiou A. External quality assessment beyond the analytical phase: An Australian perspective. Biochem Med (Zagreb). 2017;27:73–80. https://doi.org/10.11613/BM.2017.009 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/28392728

23 

Gay S, Pope B, Badrick T, Whiley M. Review of current incidents and risk calculations used in the Royal College of Australasian Pathologists Key Incident Management and Monitoring Systems - a system that could be used by all Australasian medical laboratories, and easily adapted to worldwide use. Biochem Med (Zagreb). 2022;32:010702. https://doi.org/10.11613/BM.2022.010702 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/34955670

24 

Green SF. The cost of poor blood specimen quality and errors in preanalytical processes. Clin Biochem. 2013;46:1175–9. https://doi.org/10.1016/j.clinbiochem.2013.06.001 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/23769816

25 

Howanitz PJ, Lehman CM, Jones BA, Meier FA, Horowitz GL. Clinical laboratory quality practices when hemolysis occurs. Arch Pathol Lab Med. 2015;139:901–6. https://doi.org/10.5858/arpa.2014-0252-CP PubMed: http://www.ncbi.nlm.nih.gov/pubmed/26125430

26 

Heyer NJ, Derzon JH, Winges L, Shaw C, Mass D, Snyder SR, et al. Effectiveness of practices to reduce blood sample hemolysis in EDs: A laboratory medicine best practices systematic review and meta-analysis. Clin Biochem. 2012;45:1012–32. https://doi.org/10.1016/j.clinbiochem.2012.08.002 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22968086

27 

Lee H, Lee H, Kim C, Shin H, Lee I, Kim Y. Hemolysis Control in the Emergency Department by Interventional Blood Sampling. J Pers Med. 2023;13:651. https://doi.org/10.3390/jpm13040651 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/37109037

28 

Burns ER, Yoshikawa N. Hemolysis in Serum Samples Drawn by Emergency Department Personnel versus Laboratory Phlebotomists. Lab Med. 2002;5:378–80. https://doi.org/10.1309/PGM4-4F8L-2P1M-LKPB

29 

Gay S, Sikaris K. Badrick. Tony. Reporting Critical Pathology Results - what is the current state of play in Australian Laboratories? An opinion based on findings of a KIMMS audit. Australian Journal of Medical Science. 2019;40:126–30.

30 

Dunbar NM, Kaufman RM. Factors associated with wrong blood in tube errors: An international case series – The BEST collaborative study. ransfusion. 2022;62:44–50. https://doi.org/10.1111/trf.16716 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/34726274


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.