Cochlear implant electrode array extrusion – is reimplantation feasible? – case report

Marko Velepič orcid id orcid.org/0000-0002-9826-798X ; Clinic of Otorhinolaryngology Head and Neck Surgery, Clinical Medical Centre, University of Rijeka, Rijeka,Croatia
Dubravko Manestar orcid id orcid.org/0000-0001-8908-1617 ; Clinic of Otorhinolaryngology Head and Neck Surgery, Clinical Medical Centre, University of Rijeka, Rijeka,Croatia
Petra Valković Zujić ; Clinic of Radiology, Clinical Medical Centre, University of Rijeka, Rijeka, Croatia
Goran Malvić ; Clinic of Otorhinolaryngology Head and Neck Surgery, Clinical Medical Centre, University of Rijeka, Rijeka,Croatia
Sanja Zubović Velepič ; Department for Transfusion Medicine, Clinical Medical Centre, University of Rijeka, Rijeka, Croatia


Full text: english pdf 1.291 Kb

page 99-104

downloads: 480

cite


Abstract

Aim: To present two cases of late complications of cochlear implantation. Case report: In both cases extracochlear extrusion of the implant electrode occurred. More than ten years after successful cochlear implantation, the hearing performance gradually worsened to the point of no intelligibility in both patients. Otomicroscopy revealed chronic otitis media (COM) with an electrode in the external ear canal in implanted ears of both patients. Extracochlear electrode extrusion was noticed on CT scans in both cases. In the first case the device was explanted, the electrode was left in the cochlea and tympanoplasty was performed. Regardless the normal status of the middle ear, extracochlear extrusion occurred again. Conclusion: After extracochlear electrode extrusion, the reimplantation is likely to be unsuccessful.

Keywords

chronic otitis media; cochlear implant; extracochlear electrode extrusion

Hrčak ID:

216323

URI

https://hrcak.srce.hr/216323

Publication date:

1.3.2019.

Article data in other languages: croatian

Visits: 1.446 *