What do laboratories know about the pneumatic tube delivery systems they depend on?

Anushka Jayanetti ; Royal College of Pathologists of Australasia Quality Assurance Programs, St Leonards, New South Wales, Australia
Tony Badrick ; Royal College of Pathologists of Australasia Quality Assurance Programs, St Leonards, New South Wales, Australia *

* Dopisni autor.


Puni tekst: engleski pdf 82 Kb

preuzimanja: 0

citiraj

Preuzmi JATS datoteku

Prilozi: 06_Jayanette_Supplementary_data.pdf


Sažetak

Introduction: Pneumatic tube systems (PTS) are a key component of many laboratory sample delivery systems. However, they can be a source of preanalytical error, particularly when samples are lost or damaged. The Key Incident Monitoring and Management (KIMMS) program of the Royal College of Pathologists Australasia Quality Assurance Programs (RCPAQAP) conducted a survey to gather information on the maintenance of these systems.
Materials and methods: The categories of questions included general information, service, maintenance and validation of systems, and incident reporting. The responses were compiled and reviewed using Microsoft Forms.
Results: The survey found a lack of awareness of servicing and validation records, as well as of the PTS’s functionality. Most sites reported incidents with their pneumatic tube systems (37/45, 80%), mainly system failures rather than samples being lost, delayed, or damaged.
Conclusions: Greater oversight of laboratories is needed, and relationships with system vendors must be strengthened to ensure quality service.

Ključne riječi

quality assurance; preanalytic error; specimen quality

Hrčak ID:

351778

URI

https://hrcak.srce.hr/351778

Datum izdavanja:

15.10.2026.

Posjeta: 0 *




Introduction

Pneumatic tube systems (PTS) are extensively used within clinical laboratories to transport patient samples, paperwork, and pharmacy items within or between organisations, and, to a lesser extent, blood products (1). Compared to manual transport, PTS have been shown to improve turnaround times and do not increase hemolysis (2-4). Some measurands may be impacted by the rapid acceleration in the PTS, such as platelet-related tests (5). Lactate dehydrogenase (LD), potassium, D-dimer, and prothrombin time have been shown to be affected by PTS (1,6,7). Also, blood gas samples should not be transported in PTS, and contradicting data exists on the transport of cerebrospinal fluid specimens for the diagnosis of hemorrhage. Adequate centrifugation of samples is essential, or any cells left above the gel plug can hemolyse, dependent on time and speed of the PTS (4). However, all these papers assume that the PTS is operating optimally and regularly serviced.

Methods for validating PTS have been proposed using the relationship between analyte concentrations in plasma, such as LD and potassium, and the PTS forces recorded by 3-axis accelerometers, or by more formal procedures that are not widely used in Australia (6,8). There has been little in the literature on the governance of PTS, which is important when the system may be shared between a laboratory and other hospital units.

Feedback from the 2023 Royal College of Pathologists Australasia Quality Assurance Programs (RCPAQAP) The Key Incident Monitoring and Management (KIMMS) workshop indicated that the maintenance and performance of pneumatic tube systems are not always optimal and may lead to preanalytical errors, such as hemolysed or leaked samples. In the KIMMS program, incidents related to PTS can be reported under the general transport delay category, but there is no specific PTS category. This research sought to identify issues relating to the governance and maintenance of PTS used by clinical laboratories. Therefore, for this project, we aimed to collect information on pneumatic tube systems utilised by, and often shared by, clinical laboratories to assess how they were validated and managed.

Materials and methods

We circulated a Microsoft Forms (Microsoft Corporation, Redmond, USA) audit via email to 89 quality managers across Australia/New Zealand. The survey is included in the Supplementary data. The audit was developed under the guidance of the RCPAQAP KIMMS advisory committee. Responses were anonymous, but an email address was requested. No ethical approval or institutional oversight was required. The categories of questions included general information, service, maintenance and validation of systems, and incident reporting. The responses were compiled and reviewed using Microsoft Forms. There was no statistical analysis – the data and graphs were produced by the Microsoft Forms software.

Results

Forty-five responses (51% response rate) were received from public and private pathology, with some respondents submitting multiple forms if their site had several PTS. Most of the laboratories (78%, 35/45) were part of networks of clinical laboratories, while 16% (7/45) operated as stand-alone laboratories. The remaining 6% (3/45) were laboratories that were part of a larger non-laboratory organisation. Regarding the age of these systems, 30% (14/45) had been in use for 11-20 years, 22% (10/45) for 6-10 years, 15% (7/45) for over 21 years, and 11% (5/45) for under 5 years. Only 4% (1/45) of respondents had systems with sections of varying ages, and the remaining 18% (8/45) were unsure. Most respondents (31/45) had Lamson (Kings Park, NSW, Australia) systems, while others had systems from Benchmark Systems (Willetton, WA, Australia), Aerocom (Thornleigh, NSW, Australia), Airlink Systems (Kilsyth, VIC, Australia), Telecom (Zwijndrecht, The Netherlands) and Swisslog (Redditch, United Kingdom). Hospitals were the primary managers of these systems, overseeing 73% (33/45), with pathology organisations managing 22% (10/45). Regarding transported items, blood samples were the most common at 96% (43/45), followed by paperwork at 93% (42/45), urine samples at 91% (41/45), blood cultures at 89% (40/45), and swabs at 89% (40/45). Although these are the totals, it should be remembered that a container may often contain mixtures of these.

Half of the respondents (23/45) were unsure how often their pneumatic tube systems were serviced (Table 1), and 63% (29/45) didn’t know if the systems were validated before installation (Table 2). Some validation methods included time-to-delivery, reliability, sample damage, and transport of different blood components and products. Some respondents referred to guidelines for validating systems, such as the AABB guidelines for pneumatic tube systems (6).

Table 1 Number and percentage of audit responses on how often the PTS is serviced
How often is the pneumatic tube system serviced?Responses (N = 45)% Response
Weekly00
Never24
Biannually24
Monthly613
Yearly613
Quarterly613
Unsure2351
PTS – pneumatic tube system.
Table 2 Number and percentage of audit responses on whether the system has been validated as part of its installation
Was your organization’s pneumatic tube system been validated as part of installation?Responses (N = 45)% Response
Yes1431
No24
Unsure2964

Most sites reported incidents involving their PTS (37/45, 80%), mainly system failures rather than sample loss, delays, or damage (Table 3). There was no relation between system age and incidents, though some respondents said ageing systems failed more often. There was also no trend between incident frequency and servicing (Table 4).

Table 3 Audit responses on the frequency of incidents involving pneumatic tube systems
Responses (N = 45)
< 55 to 10> 10UnsureOther
How many times has the pneumatic tube system failed in the past 6 months?199853
How many times have specimens been lost, misdirected or delayed when transported by the pneumatic tube system in the past 6 months?2625100
Table 4 Comparison of responses on the frequency of servicing of pneumatic tube systems versus the number of incidents reported
ResponseService interval of PTSFailure of PTS in the past 6 monthsNumber of lost, misdirected or delayed specimens in the PTS in the past 6 monthsFrequency of specimen transport delay through PTS, causing them to age (in the past 6 months)Specimens have been going to the wrong location through PTS in the past 6 monthsNumber of specimens damaged in the PTS in the past 6 months (i.e. hemolyzed, clotted)Frequency of specimen leakage in the system in the past 6 months
13 monthly< 5< 5< 5< 5
26 monthlyUnsureUnsure< 5< 5Unsure< 5
36 monthly< 5< 5
4Approx monthly?42UnsureUnsureUnsureUnsure< 5
5Lamson service monthly, hospital staff do daily maintenanceNever fails, do have issues that are dealt with by the service team< 5
6Monthly5 to 105 to 105 to 10Unsure< 5
7Monthly> 10< 5< 5< 5Unsure5 to 10
8MonthlyJust around 40 times where the shift engineer was called in to fix (April- October> 10
9Monthly
10Never< 5< 5
11Never< 5< 5
12Quarterly5 to 10< 5
13Quarterly5 to 105 to 10< 5< 5< 5< 5
14Quarterly> 10> 10< 5> 10< 5> 10
15Quarterly< 5< 5UnsureUnsure
16Quarterly< 5< 5UnsureUnsure

Discussion

The lack of awareness of servicing and validation records suggests that laboratories may not have sufficient oversight of their operating systems. Instead, others often handle these responsibilities, such as hospital administration. Because of this, laboratories may not realise whether their systems are functioning correctly or can easily investigate issues such as turnaround times, which are crucial for improving their services. While the harm of delayed paperwork is not as severe as that of delayed blood samples or CSF, it may still impact patient care.

Something to note is that many respondents who were unsure whether their systems were validated or serviced also could not identify related incidents, further indicating the need for greater surveillance of laboratories.

Pneumatic tube systems pose a potential patient risk due to preanalytical errors, which can be mitigated through maintenance and vendor servicing adjustments tailored to each site (1). Greater oversight of laboratories is needed, and relationships with system vendors must be strengthened to ensure quality service. The results of the study primarily indicate that many laboratory respondents were unaware of servicing frequency, validation status and incident records, and that governance of PTS may be fragmented. However, another major finding of the study is that validation of the PTS has not been conducted, and that maintenance of these systems is not being undertaken. This is noted as anecdotal, based on the descriptive survey findings as represented in the individual responses in the Supplementary data.

There were limitations to this survey. As this was a survey-based audit, the results were subject to self-reporting bias and differences in respondents’ access to information on PTS servicing, validation and incident records. Also, access to service records, particularly shared PTS records, was not always available to respondents. For some incident categories, the numbers were small, reducing the statistical power of the findings. Finally, we acknowledge potential selection and response bias, as only 45 of 89 invited quality managers responded.

Acknowledgement

The pneumatic tube system audit was developed and reviewed with the guidance of the RCPAQAP KIMMS Advisory Committee.

Notes

[1] Conflicts of interest Potential conflict of interest

None declared.

Data availability statement

All data generated and analyzed in the presented study are included in this published article (and its supplementary files).

References

1 

Plebani M, Zaninotto M. Pneumatic tube delivery systems for patient samples: evidence of quality and quality of evidence. Clin Chem Lab Med. 2011;49:1245–6. https://doi.org/10.1515/CCLM.2011.216 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/21663463

2 

Fernandes CMB, Worster A, Eva K, Hill S, McCallum C. Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality. J Emerg Nurs. 2006;32:139–43. https://doi.org/10.1016/j.jen.2005.11.013 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/16580476

3 

Ausmolo AJT, Pua JFD, Abiog MDH, Alberto BC, Bilo ELM, De Leon CIL, et al. A Systematic Review of the Impact of Pneumatic Tube System Transport on Blood Chemistry and Hematology Parameters. Res Sq [Internet]. 2022;1-35. Available from: https://doi.org/10.21203/rs.3.rs-2274742/v2. Assessed on January 15th 2026.

4 

Evliyaoğlu O, Toprak G, Tekin A, Başarali MK, Kilinç C, Çolpan L. Effect of Pneumatic Tube Delivery System Rate and Distance on Hemolysis of Blood Specimens. J Clin Lab Anal. 2012;26:66–9. https://doi.org/10.1002/jcla.21484 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22467320

5 

Glas M, Mauer D, Kassas H, Volk T, Kreuer S. Sample transport by pneumatic tube system alters results of multiple electrode aggregometry but not rotational thromboelastometry. Platelets. 2013;24:454–61. https://doi.org/10.3109/09537104.2012.718383 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/22931353

6 

Subbarayan D, Choccalingam C, Lakshmi CKA. The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests. Adv Hematol. 2018;2018:6940152. https://doi.org/10.1155/2018/6940152 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/30079089

7 

Farnsworth CW, Webber DM, Krekeler JA, Budelier MM, Bartlett NL, Gronowski AM. Parameters for validating a hospital pneumatic tube system. Clin Chem. 2019;65:694–702. https://doi.org/10.1373/clinchem.2018.301408 PubMed: http://www.ncbi.nlm.nih.gov/pubmed/30808643

8 

Karp JK, Harach M, Mohammed M, Richard R, Uhl L, van der Hoop J. AABB Guide to Pneumatic Tube Delivery Systems: Validation and Use to Transport Blood Components. Association for the Advancement of Blood & Biotherapies; 2020. Available at:https://www.aabb.org/aabb-store/product/aabb-guide-to-pneumatic-tube-delivery-systems-validation-and-use-to-transport-blood-components---digital-15175121. Assessed on January 15th 2026.


This display is generated from NISO JATS XML with jats-html.xsl. The XSLT engine is libxslt.