Skoči na glavni sadržaj

Izvorni znanstveni članak

https://doi.org/10.24138/jcomss.v3i1.267

GTEM Cell Experimental Set up for In Vitro Dosimetry

Giovanna Calò ; Politecnico di Bari, Dipartimento di Elettrotecnica ed Elettronica, Bari, Italy
Francesco Lattarulo ; Politecnico di Bari, Dipartimento di Elettrotecnica ed Elettronica, Bari, Italy
Vincenzo Petruzzelli ; Politecnico di Bari, Dipartimento di Elettrotecnica ed Elettronica, Bari, Italy


Puni tekst: engleski pdf 1.784 Kb

str. 34-43

preuzimanja: 644

citiraj


Sažetak

GTEM cell, suitable for assessing possible biological effects induced on cell samples by electromagnetic fields at the typical frequencies of GSM mobile phones, has been designed and set up. Basic environmental requirements for in-vitro biological experiments, involving a GTEM cell, have been assessed by controlling the electromagnetic field distribution and survival conditions. The GTEM cell has been characterized by Standing Wave Ratio (SWR) and Time Domain Reflectometry (TDR) measurements. The impedance matching at the terminal load section has been optimized by considering different hybridload configurations. Moreover, optimal exposure conditions for the biological sample have been experimentally evaluated by paying special attention to the E-field scenario inside the GTEM cell at 900 MHz and 1800 MHz frequencies. At last, an experimental evaluation of the Specific Absorption Rate (SAR) is reported.

Ključne riječi

Dosmetry; Specific Absorption Rate; GTEM cell

Hrčak ID:

180574

URI

https://hrcak.srce.hr/180574

Datum izdavanja:

26.3.2007.

Posjeta: 1.102 *