hrcak mascot   Srce   HID

Prethodno priopćenje
https://doi.org/10.20471/acc.2019.58.02.24

Probno istraživanje: ortopedska svrdla s protočnim hlađenjem otvorenog tipa - je li standardna sterilizacija dovoljna?

Tomislav Bruketa ; Department of Surgery, Zagreb University Hospital Centre and School of Medicine, University of Zagreb, Zagreb, Croatia
Goran Augustin ; Department of Surgery, Zagreb University Hospital Centre and School of Medicine, University of Zagreb, Zagreb, Croatia
Selma Pintarić ; Department of Microbiology and Infectious Diseases with Clinic, Veterinary Faculty, University of Zagreb, Zagreb, Croatia
Branka Šeol-Martinec ; Department of Microbiology and Infectious Diseases with Clinic, Veterinary Faculty, University of Zagreb, Zagreb, Croatia
Ivan Dobrić ; Department of Surgery, Zagreb University Hospital Centre and School of Medicine, University of Zagreb, Zagreb, Croatia
Bore Bakota ; Trauma and Orthopaedics Department, Brighton and Sussex University Hospitals, NHS Trust, Brighton, United Kingdom

Puni tekst: engleski, pdf (330 KB) str. 379-384 preuzimanja: 151* citiraj
APA 6th Edition
Bruketa, T., Augustin, G., Pintarić, S., Šeol-Martinec, B., Dobrić, I. i Bakota, B. (2019). Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?. Acta clinica Croatica, 58 (2.), 379-384. https://doi.org/10.20471/acc.2019.58.02.24
MLA 8th Edition
Bruketa, Tomislav, et al. "Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?." Acta clinica Croatica, vol. 58, br. 2., 2019, str. 379-384. https://doi.org/10.20471/acc.2019.58.02.24. Citirano 23.10.2020.
Chicago 17th Edition
Bruketa, Tomislav, Goran Augustin, Selma Pintarić, Branka Šeol-Martinec, Ivan Dobrić i Bore Bakota. "Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?." Acta clinica Croatica 58, br. 2. (2019): 379-384. https://doi.org/10.20471/acc.2019.58.02.24
Harvard
Bruketa, T., et al. (2019). 'Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?', Acta clinica Croatica, 58(2.), str. 379-384. https://doi.org/10.20471/acc.2019.58.02.24
Vancouver
Bruketa T, Augustin G, Pintarić S, Šeol-Martinec B, Dobrić I, Bakota B. Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?. Acta clinica Croatica [Internet]. 2019 [pristupljeno 23.10.2020.];58(2.):379-384. https://doi.org/10.20471/acc.2019.58.02.24
IEEE
T. Bruketa, G. Augustin, S. Pintarić, B. Šeol-Martinec, I. Dobrić i B. Bakota, "Pilot Study: Internally Cooled Orthopedic Drills – Standard Sterilization Is Not Enough?", Acta clinica Croatica, vol.58, br. 2., str. 379-384, 2019. [Online]. https://doi.org/10.20471/acc.2019.58.02.24

Sažetak
Bušenje kosti izaziva porast temperature na mjestu kontakta metala i kosti, što može rezultirati termičkom osteonekrozom. Novokonstruirano svrdlo s unutarnjim hlađenjem otvorenog tipa smanjuje porast temperature na mjestu kontakta metala i kosti, ali standardna sterilizacija takvog svrdla može biti nedovoljna zbog zadržavanja bakterija unutar kanala svrdla. Cilj ovog probnog istraživanja bila je procjena učinkovitosti sterilizacije te prijedlog preporuka za sterilizaciju medicinskog svrdla s unutarnjim hlađenjem otvorenog tipa. Testirana su nekorištena medicinska svrdla s unutarnjim hlađenjem otvorenog tipa. Svrdla su kontaminirana sljedećim bakterijama: Pseudomonas aeruginosa, Bacillus sp., beta-hemolitički Streptococcus sp.,
Enterobacter sp. i meticilin-rezistentni Staphylococcus pseudintermedius. Inkubacija je trajala 24 sata na temperaturi od 37 °C. Potom su svrdla sterilizirana u autoklavu 15, 20 i 30 minuta na temperaturi od 132 °C i tlaku od 2,6 bara. Kod sterilizacije u trajanju od 15 minuta jedno od 16 korištenih svrdla bilo je kontaminirano bakterijom Pseudomonas aeruginosa, dok su ostala svrdla bila sterilna. Produženi ciklus sterilizacije u autoklavu u trajanju od 20 odnosno 30 minuta rezultirao je sterilnošću svih svrdla. U zaključku, svrdla s lumenom je potrebno sterilizirati produženim ciklusom sterilizacije. Minimalno preporučeno trajanje sterilizacije je 20 minuta.

Ključne riječi
Kosti – ozljede; Osteonekroza; Ortopedski postupci – štetna djelovanja; Sterilizacija; Kirurška rana, infekcija