Early dysfunction of an implanted pacemaker in a patient with complete atrioventricular block

Ivica Dunđer orcid id orcid.org/0000-0002-3340-7590 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Katica Cvitkušić Lukenda orcid id orcid.org/0000-0001-6188-0708 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Ivan Bitunjac orcid id orcid.org/0000-0002-4396-6628 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Marijana Knežević Praveček orcid id orcid.org/0000-0002-8727-7357 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Krešimir Gabaldo orcid id orcid.org/0000-0002-0116-5929 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Domagoj Mišković orcid id orcid.org/0000-0003-4600-0498 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia
Domagoj Vučić orcid id orcid.org/0000-0003-3169-3658 ; General Hospital ”Dr Josip Benčević”, Slavonski Brod, Croatia


Full text: english pdf 2.515 Kb

page 24-25

downloads: 63

cite

Download JATS file


Abstract

Keywords

atrioventricular block; implantable pacemaker dysfunction

Hrčak ID:

343327

URI

https://hrcak.srce.hr/343327

Publication date:

15.1.2026.

Visits: 180 *



Introduction: Complete atrioventricular (AV) block is a life-threatening condition characterized by a partial or complete interruption of impulse conduction from the atria to the ventricles. The most common causes are fibrosis and sclerosis of the conduction system, ischemic heart disease, and electrolyte imbalance (1).

Case report: 85-year-old patient was admitted via the Emergency Department due to recurrent episodes of loss of consciousness, with ECG showing complete AV block and a ventricular rate of 28 bpm (Figure 1). Multiple comorbidities were present, including type 2 diabetes mellitus with complications, arterial hypertension, and a history of stroke 30 years ago, with resulting right-sided hemiparesis and aphasia. Temporary external pacing electrodes were promptly placed, and upon admission to the Coronary Care Unit, a temporary pacemaker was inserted via the right jugular vein, after which the patient was stabilized. On the following day, a permanent dual-chamber pacemaker was implanted via the cephalic vein, with excellent stimulation parameters (Figures 2 and 3{ label needed for fig[@id='f3'] }). On the third day post-implantation, telemetry recorded an asystolic episode with cardiorespiratory arrest and loss of pacemaker output, prompting resuscitation and re-initiation of external pacing. A temporary pacemaker was reinserted, and the revision of the previously implanted pacemaker was performed. Radiographic imaging revealed no displacement of the electrode, with identical positioning and normal electrode impedance. The ventricular electrode failed to transmit impulses even when connected to the programmer. Therefore, a new ventricular electrode was implanted via the left subclavian vein (Figure 4).

FIGURE 1 12-lead electrocardiogram on admission.
CC202621_1-2_24-5-f1
FIGURE 2 12-lead electrocardiogram after pacemaker implantation.
CC202621_1-2_24-5-f2
{ label needed for fig[@id='f3'] }
FIGURE 3 Chest X-ray at initial implantation.
CC202621_1-2_24-5-f3
FIGURE 4 Chest X-ray after implantation of the new ventricular electrode.
CC202621_1-2_24-5-f4

Conclusion: Dysfunction of the ventricular electrode may have been due to a microfracture of the electrode, which was not visible radiographically, or a microtear into fibrotic tissue (2).