Skoči na glavni sadržaj

Izvorni znanstveni članak

Insulation Verification Using Low Voltage and High Current Sensitivity

Tonko Garma orcid id orcid.org/0000-0001-9712-1084 ; Walter Schottky Institut, Technische Universität München
Željka Milanović ; Tehnički fakultet Sveučilišta u Rijeci
Ivan Marasović ; Fakultet elektrotehnike, strojarstva i brodogradnje Sveučilišta u Splitu


Puni tekst: engleski pdf 283 Kb

str. 86-95

preuzimanja: 1.640

citiraj


Sažetak

The classic approach for determining insulating resistance by using commercial testers is not reliable for testing different types of insulation, especially for cables with one conductor and nanoelectronic oxide-based devices (nanotransistors). In this paper an alternative low-voltage method for measuring insulation resistance based on strongly increased current sensitivity is proposed. The proposed method of measuring insulation resistance at low voltages is realized by using a high sensitivity Keithley SMU 2612 unit. The verification of the proposed method is made by measuring insulation resistance of medical Corkscrew Electrode and gate oxide resistance for nano-scaled MOSFET. Finally, results of measuring insulation resistance obtained by commercial testers and proposed method in the verification process are compared.

Ključne riječi

Commercial insulation testers; Insulation resistance measuring methods; Low-voltage resistance measuring method; Nano-transistor gate-oxide resistance

Hrčak ID:

84222

URI

https://hrcak.srce.hr/84222

Datum izdavanja:

18.6.2012.

Posjeta: 2.433 *